niRFSG 5652 Update ALC Limits Table VI
- Updated2024-05-08
- 3 minute(s) read
niRFSG 5652 Update ALC Limits Table VI
Updates the PXI/PXIe-5650/5651/5652 ALC limits table, which determines the attenuator and path NI-RFSG uses for each frequency and power combination.
This VI computes the ALC limits table based on calibration data saved during the ALC and attenuator adjustment procedures as well as temperature correction data saved when the device was manufactured.
Supported Devices: PXI/PXIe-5650/5651/5652
![]() |
instrument handle identifies your instrument calibration session. instrument handle is obtained from the niRFSG Initialize External Calibration VI. |
![]() |
update max power spec specifies whether to update and overwrite the maximum power specification in the device ALC limits table. |
![]() |
update other table limits specifies whether to update all other ALC limits table values except the maximum power specification value. Set this parameter to TRUE to update all other ALC limits table values. Set this parameter to FALSE to not update all other ALC limits table values. |
![]() |
max power spec override This parameter is for NI internal use only. |
![]() |
error in describes error conditions that occur before this node runs. This input provides standard error in functionality. |
![]() |
instrument handle out passes a reference to your instrument session to the next VI. |
![]() |
max power spec changed? indicates whether the maximum power specification changed. If this parameter returns a value of TRUE, this VI changed the maximum power specification in the device ALC limits table. |
![]() |
error out contains error information. This output provides standard error out functionality. |