NI 5652 Calibration

Owning Palette: External Calibration

Use the NI 5652 Calibration VIs to calibrate your PXI/PXIe-5650/5651/5652 device. Refer to the calibration procedure for your device for more information about device calibration.

Palette Object Description
niRFSG 5652 Initialize ALC Calibration

Generates external calibration constants for calibrating the automatic leveling control (ALC).

Supported Devices: PXI/PXIe-5650/5651/5652

niRFSG 5652 Configure ALC Calibration

Configures the PXI/PXIe-5650/5651/5652 for the next automatic leveling control (ALC) calibration point and waits for the output signal to settle.

Supported Devices: PXI/PXIe-5650/5651/5652

niRFSG 5652 Adjust ALC Calibration

Calculates calibration information pertaining to the ALC of the PXI/PXIe-5650/5651/5652. This calculation is based on user-supplied measurements. The calibration information is stored in the driver session, and you can later write this information to the device.

Supported Devices: PXI/PXIe-5650/5651/5652

niRFSG 5652 Initialize Attenuator Calibration

Generates external calibration constants for calibrating the attenuator.

Supported Devices: PXI/PXIe-5650/5651/5652

niRFSG 5652 Configure Attenuator Calibration

Configures the PXI/PXIe-5650/5651/5652 for the next attenuator calibration point and waits for the output signal to settle.

Supported Devices: PXI/PXIe-5650/5651/5652

niRFSG 5652 Adjust Attenuator Calibration

Calculates calibration information pertaining to the selected attenuator on the PXI/PXIe-5650/5651/5652. This calculation is based on user-supplied measurements. The calibration information is stored in the driver session, and you can later write this information to the device.

Supported Devices: PXI/PXIe-5650/5651/5652

niRFSG 5652 Update ALC Limits Table

Updates the PXI/PXIe-5650/5651/5652 ALC limits table, which determines the attenuator and path NI-RFSG uses for each frequency and power combination.

This VI computes the ALC limits table based on calibration data saved during the ALC and attenuator adjustment procedures as well as temperature correction data saved when the device was manufactured.

Supported Devices: PXI/PXIe-5650/5651/5652

niRFSG 5652 Validate ALC Limits Table

Validates the PXI/PXIe-5650/5651/5652 ALC limits table, which determines the attenuator and path NI-RFSG uses for each frequency and power combination.

Supported Devices: PXI/PXIe-5650/5651/5652

niRFSG 5652 Initialize Reference Clock Calibration

Generates external calibration constants for calibrating the 10 MHz reference.

Supported Devices: PXI/PXIe-5650/5651/5652

niRFSG 5652 Configure Reference Clock Calibration

Configures the PXI/PXIe-5650/5651/5652 with specific settings for calibrating the 10 MHz OCXO reference.

Supported Devices: PXI/PXIe-5650/5651/5652

niRFSG 5652 Adjust Reference Clock Calibration

Calculates calibration information pertaining to the 10 MHz OCXO reference of the PXI/PXIe-5650/5651/5652. This calculation is based on user-supplied measurements. The calibration information is stored in the driver session, and you can later write this information to the device.

Supported Devices: PXI/PXIe-5650/5651/5652