Vertical:Advanced:Thermal Correction Headroom Range (Degrees C) Property
- Updated2024-05-08
- 1 minute(s) read
Vertical:Advanced:Thermal Correction Headroom Range (Degrees C) Property
Short Name: Thermal Correction Headroom Range
Property of niRFSA
Specifies the expected thermal operating range of the instrument from the self-calibration temperature, in degrees Celsius, returned from the Device Temperature property.
For example, if this property is set to 5.0, and the device is self-calibrated at 35 degrees Celsius, then you can expect to run the device from 30 degrees Celsius to 40 degrees Celsius with corrected accuracy and no overflows. Setting this property with a smaller value can result in improved dynamic range, but you must ensure thermal stability while the instrument is running. Operating the instrument outside of the specified range may cause degraded performance and ADC or DSP overflows.
Default Value:
PXIe-5830/5831/5832/5842: 5
PXIe-5840/5841: 10
Supported Devices: PXIe-5830/5831/5832/5840/5841/5842
Remarks
The following table lists the characteristics of this property.
Datatype | ![]() |
Permissions | Read/Write |
High-level VIs | N/A |
Channel-based | No |
Resettable | Yes |