Triggers Summary
- Updated2025-10-10
- 1 minute(s) read
The following table describes the triggers supported by signal generators. The Supported Types column denotes which trigger types are valid for a given trigger.
| Trigger Name | Supported Types | Description |
|---|---|---|
| Start | Digital Edge, Software | The Start trigger transitions a device from an idle state to a generation state where the device can respond to Sample clocks. |
| Script | Digital Edge, Digital Level, Software | The Script trigger is a general-purpose trigger with a role that is entirely determined by the context of the generation script. A script allows you to create sophisticated generation operations. For example, the script could configure the device to generate waveform A, then wait for the Script trigger, then generate waveform B. You can create multiple Script triggers for use in your application. Once a digital edge Script trigger has been received, that trigger remains true for all subsequent instructions until the clear instruction is called or the trigger is reset after being used in the wait, repeat/end repeat, or if instructions. |
Related Information
- Edge Trigger
- Software Trigger
- Level Trigger
- Script Mode
Script mode allows you to use scripting to link and loop multiple waveforms in complex combinations using a script.