NI ELVIS III Toolkit

Table of Contents

FGEN Configure Frequency Modulation Waveform VI

  • Updated2023-02-21
  • 2 minute(s) read

FGEN Configure Frequency Modulation Waveform VI

Owning Palette: Configuration VIs

Requires: ELVIS III Toolkit

Configures the frequency modulation waveform.

Instrument Handle In specifies the session created by the FGEN Initialize VI.
Channel specifies the name of the channel. You can specify only one channel.
Waveform specifies the modulation waveform that modulates the generated signal. The instantaneous amplitude of Waveform decides by which factor the frequency of the generated signal is increased. The factor equals the instantaneous amplitude of Waveform plus one. The amplitude of the waveform should be between -1 to 1.
Run Once? specifies if the instrument generates the waveform only once or continuously.
TRUEGenerates the waveform only once.
FALSEGenerates the waveform continuously.
Error In describes error conditions that occur before this node runs. This input provides standard error in functionality.
Instrument Handle Out returns a reference to the instrument.
Error Out contains error information. This output provides standard error out functionality.