Specifies the trigger source. When niDMM Initiate is called, the DMM waits for the trigger specified with this property. After it receives the trigger, the DMM waits the length of time specified with the Trigger Delay property. The DMM then takes a measurement.

To determine which values are supported by each device, refer to the LabVIEW Trigger Routing section in the NI Digital Multimeters Help.

Remarks

The following table lists the characteristics of this property.

Short Name Trigger Source
Data type ci32.png
Permissions Read/Write
High-level VIs niDMM Configure Trigger
Resettable No
Immediate 1

No trigger specified.

External 2

Pin 9 on the AUX Connector

Software Trig 3

Waits until niDMM Send Software Trigger is called.

TTL 0 111

PXI Trigger Line 0

TTL 1 112

PXI Trigger Line 1

TTL 2 113

PXI Trigger Line 2

TTL 3 114

PXI Trigger Line 3

TTL 4 115

PXI Trigger Line 4

TTL 5 116

PXI Trigger Line 5

TTL 6 117

PXI Trigger Line 6

TTL 7 118

PXI Trigger Line 7

PXI Star 131

PXI Star Trigger Line

LBR Trig 1 1004

Local Bus Right Trigger Line 1 of PXI/SCXI combination chassis

AUX_Trig 1 1001

Pin 3 on the AUX connector