niDCPower Cal Adjust Output Resistance VI
- Updated2025-04-18
- 3 minute(s) read
Compares the array in requested outputs to the array in measured outputs and calculates the calibration constants for the output resistance of the specified channel. Refer to the calibration procedure for the device you are calibrating for detailed instructions on the appropriate use of this VI. This VI can only be called from an external calibration session.

Inputs/Outputs
![]() instrument handle identifies a particular instrument calibration session. instrument handle is obtained from the niDCPower Initialize External Calibration VI. ![]() channel name specifies the channel to which these calibration settings apply. Only one channel at a time can be calibrated. ![]() number of measurements specifies the number of elements in requested outputs and measured outputs. ![]() error in describes error conditions that occur before this node runs. This input provides standard error in functionality. ![]() requested outputs specifies an array of the output values requested in the niDCPower Configure Current Limit VI. ![]() measured outputs specifies an array of the output values measured by an external precision digital multimeter. ![]() instrument handle out passes the handle used to identify the session in all subsequent NI-DCPower VI calls. ![]() error out contains error information. This output provides standard error out functionality. |