NIDCPOWER_ATTR_SEQUENCE_ENGINE_DONE_EVENT_PULSE_WIDTH
- Updated2023-04-15
- 1 minute(s) read
Specific Attribute
Numeric Value | Data type |
Access | Applies to | Coercion | High-Level Functions |
---|---|---|---|---|---|
1150049 | ViReal64 | R/W | Channel | Yes | N/A |
Description
Specifies the width of the Sequence Engine Done event, in seconds.
The minimum event pulse width value for PXI devices is 150 ns, and the minimum event pulse width value for PXI Express devices is 250 ns.
The maximum event pulse width value for all devices is 1.6 microseconds.
![]() |
Note This attribute is not supported by all devices. Refer to Supported Attributes by Device for information about supported devices. |
![]() |
Note To specify a channel name when accessing this attribute, you must first initialize the session using the niDCPower_InitializeWithIndependentChannels function. |
Valid Values: 1.5e-7 to 1.5e-6 seconds
Default Value: The default value for PXI Express devices is 250 ns.