Adjust Residual Current Offset Calibration

Calculates the calibration constants for the residual current offsets for the specified output channel.

This node can be called only in an external calibration session. The node requires that the output be open prior to it being invoked.

Inputs/Outputs

datatype_icon

session in

Instrument calibration session obtained from the Initialize External Calibration node.

datatype_icon

channel name

Output channel to which these calibration settings apply.

Only one channel at a time can be calibrated.

datatype_icon

error in

Error conditions that occur before this node runs.

The node responds to this input according to standard error behavior.

Standard Error Behavior

Default value: No error

datatype_icon

session out

Handle used to identify the session in all subsequent NI-DCPower node calls.

datatype_icon

error out

Error information.

The node produces this output according to standard error behavior.

Standard Error Behavior

Hardware Support

This node does not support the following hardware:

  • PXI-4110
  • PXIe-4112/4113
  • PXI-4130
  • PXI-4132
  • PXIe-4154

Residual Current Offsets

Calibration Procedure

Refer to the calibration procedure for the device you are calibrating for detailed instructions on the appropriate use of this node.