Performs external calibration adjustment on a DSA device with the specified gain and coupling configuration.


icon

Inputs/Outputs

  • cu32.png calhandle in

    calhandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration VI.

  • cdbl.png reference voltage

    reference voltage specifies in volts the known voltage to use as a reference for calibration. This unit is Vrms for AC coupling or VDC for DC coupling.

  • cdbl.png gain (dB)

    gain (dB) specifies the gain setting to calibrate, in decibels.

  • ci32.png coupling

    coupling specifies the coupling setting to calibrate.

    AC (10045)

    AC coupling.

    DC (10050)

    DC coupling.

  • cerrcodeclst.png error in

    error in describes error conditions that occur before this VI or function runs.

    The default is no error. If an error occurred before this VI or function runs, the VI or function passes the error in value to error out. If an error occurs while this VI or function runs, the VI or function runs normally and sets its own error status in error out. Use the Simple Error Handler or General Error Handler VIs to display the description of the error code. Use error in and error out to check errors and to specify execution order by wiring error out from one node to error in of the next node.

  • iu32.png calhandle out

    calhandle out is a reference to the calibration session. Wire this output to other external calibration VIs.

  • ierrcodeclst.png error out

    error out contains error information. If error in indicates that an error occurred before this VI or function ran, error out contains the same error information. Otherwise, error out describes the error status that this VI or function produces.

    Right-click the error out indicator on the front panel and select Explain Error from the shortcut menu for more information about the error.