These specifications apply to the NI-9208.
Revision History
Version | Date changed | Description |
---|---|---|
378890B-01 | February 2025 | Updated accuracy table. |
378890A-01 | December 2022 | Initial release. |
Looking For Something Else?
For information not found in the specifications for your product, such as operating instructions, browse Related Information.
Connector Types
The NI-9208 has more than one connector type: NI-9208 with spring terminal and NI-9208 with DSUB. Unless the connector type is specified, NI-9208 refers to both connector types.
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
Specifications are Typical unless otherwise noted.
Conditions
Specifications are valid for the range -40 °C to 70 °C unless otherwise noted. All voltages are relative to COM unless otherwise noted.
NI-9208 with Spring Terminal Pinout
Signal | Description |
---|---|
AI | Analog input signal connection |
COM | Common reference connection to isolated ground |
Vsup | Voltage supply connection |
NI-9208 with DSUB Terminal Pinout
Signal | Description |
---|---|
AI | Analog input signal connection |
COM | Common reference connection to isolated ground |
Vsup | Voltage supply connection |
Input Characteristics
Number of channels | 16 analog input channels |
ADC resolution | 24 bits |
Type of ADC | Delta-Sigma |
Sampling mode | Scanned |
Minimum | ±21.5 mA |
Typical | ±22 mA |
High-resolution mode | 52 ms |
High-speed mode | 2 ms |
Overvoltage protection, channel-to-COM | ±30 V maximum on one channel at a time |
Current | 2 A maximum |
Voltage | 30 V maximum |
Input impedance | 85 Ω |
Calibrated Measurement Conditions | Within 2 Years of Calibration | Within 10 Years of Calibration | ||
---|---|---|---|---|
Percent of Reading (Gain Error) | Percent of Range (Offset Error) | Percent of Reading (Gain Error) | Percent of Range (Offset Error) | |
Maximum (-40 °C to 70 °C) | ±0.76% | ±0.04% | 0.86% | 0.04% |
High-resolution mode | 50 nA RMS |
High-speed mode | 200 nA RMS |
Gain drift | 20 ppm/°C |
Offset drift | 62 nA/°C |
50 Hz | 66 dB |
60 Hz | 68 dB |
NI-9208 with Spring Terminal Safety Voltages
Connect only voltages that are within the following limits:
|
NI-9208 with DSUB Safety Voltages
Connect only voltages that are within the following limits:
|
Measurement Category
Measurement Category I
Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.
Measurement Category II
Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet, for example, 115 V for U.S. or 230 V for Europe.
Environmental Characteristics
| |||||||||||||
| |||||||||||||
Ingress protection | IP40 | ||||||||||||
Pollution Degree | 2 | ||||||||||||
| |||||||||||||
|
To meet these shock and vibration specifications, you must panel mount the system.
Power Requirements
| |||||||
|
Physical Characteristics
| |||||||
Dimensions | Visit ni.com/dimensions and search by module number. |
NI-9208 with Spring Terminal
| |||||||||||||
|
Calibration
Recommended calibration interval | 2 years |