NI-9157 Specifications

Definitions

Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical specifications describe the performance met by a majority of models.
  • Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.

Specifications are Typical unless otherwise noted.

Conditions

Specifications are valid for the range 0 °C to 55 °C unless otherwise noted.

MXI-Express

Maximum cable length

7 m

Reconfigurable FPGA

FPGA type

Virtex-5 LX85

Number of flip-flops

51,840

Number of 6-input LUTs

51,840

Number of DSP48 slices (25 x 18 multipliers)

48

Embedded block RAM

3,456 kbits

Timebases

40 MHz, 80 MHz, 120 MHz, 160 MHz, or 200 MHz

Accuracy

±100 ppm (maximum)

Frequency dependent on jitter (peak-to-peak)

40 MHz

250 ps

80 MHz

422 ps

120 MHz

422 ps

160 MHz

402 ps

200 MHz

402 ps

Safety Voltages

Connect only voltages that are within the following limits:

V terminal to C terminal

30 VDC maximum, Measurement Category I

Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.

Notice Do not connect the NI-9157 to signals or use for measurements within Measurement Categories II, III, or IV.
Note Measurement Categories CAT I and CAT O are equivalent. These test and measurement circuits are for other circuits not intended for direct connection to the MAINS building installations of Measurement Categories CAT II, CAT III, or CAT IV.

Environmental Characteristics

Temperature

Operating

0 °C to 55 °C

Storage

-40 °C to 85 °C

Humidity

Operating

10% RH to 90% RH, noncondensing

Storage

5% RH to 95% RH, noncondensing

Ingress protection

IP40

Pollution Degree

2

Maximum altitude

2,000 m

Shock and Vibration
Operating vibration

Random

5 g RMS, 10 Hz to 500 Hz

Sinusoidal

5 g, 10 Hz to 500 Hz

Operating shock

30 g, 11 ms half sine; 50 g, 3 ms half sine; 18 shocks at 6 orientations

To meet these shock and vibration specifications, you must panel mount the system.

Power Requirements

Voltage input range

9 V to 30 V

Maximum power input

55 W, 30 VDC maximum

Maximum power consumption

With no I/O modules

14.4 W, maximum

With 14 I/O modules

30.7 W, maximum

Note The power consumption specifications in this document are maximum values for a LabVIEW FPGA application compiled at 80 MHz. Your application power requirements may be different. To calculate the power requirements of the NI-9157, add the power consumption/dissipation for the chassis and the I/O modules you are using. Keep in mind that the resulting total power consumption is a maximum value and that the NI-9157 may require less power in your application. For more information about the I/O module power requirements, refer to the module operating instructions.

Physical Characteristics

Screw-terminal wiring

Gauge

0.2 mm 2 to 2.1 mm2 (24 AWG to 14 AWG) copper conductor wire

Wire strip length

6 mm (0.24 in.) of insulation stripped from the end

Temperature rating

85 °C

Torque for screw terminals

0.20 N · m to 0.25 N · m (1.8 lb · in. to 2.2 lb · in.)

Wires per screw terminal

One wire per screw terminal

Connector securement

Securement type

Screw flanges provided

Torque for screw flanges

0.3 N · m to 0.4 N · m (2.7 lb · in. to 3.5 lb · in.)

Weight

2,231 g (78.7 oz)