Electromagnetic Compatibility Guidelines

This product was tested and complies with the regulatory requirements and limits for electromagnetic compatibility (EMC) as stated in the product specifications. These requirements and limits are designed to provide reasonable protection against harmful interference when the product is operated in its intended operational electromagnetic environment.

This product is intended for use in industrial locations. There is no guarantee that harmful interference will not occur in a particular installation, when the product is connected to a test object, or if the product is used in residential areas. To minimize the potential for the product to cause interference to radio and television reception or to experience unacceptable performance degradation, install and use this product in strict accordance with instructions in the product documentation.

Furthermore, any changes or modifications to the product not expressly approved by NI could void your authority to operate it under your local regulatory rules.

Caution To ensure the specified EMC performance, you must install PXI EMC Filler Panels (NI part number 778700-01) in adjacent chassis slots. For more information about installing PXI EMC filler panels in your system, refer to the Installing EMI Controls section of this document.
Caution To ensure the specified EMC performance, operate this product only with shielded cables and accessories.
Caution To ensure the specified EMC performance, install the included snap-on ferrite bead (NI part number 711856-01) on any cable attached to the AUX I/O connector in accordance with the instructions listed in the Installing EMI Controls section.
Caution This product is sensitive to electrostatic discharge (ESD). To ensure the specified EMC performance, follow the programming instructions listed at the end of the Using Your Device with a LabVIEW FPGA Example VI and Creating a LabVIEW Project and Running a VI on an FPGA Target sections of this document.