niRFSG 5652 Adjust Reference Clock Calibration VI
Calculates calibration information pertaining to the 10 MHz OCXO reference of the PXI/PXIe-5650/5651/5652. This calculation is based on user-supplied measurements. The calibration information is stored in the driver session, and you can later write this information to the device.
Supported Devices: PXI/PXIe-5650/5651/5652
![]() |
instrument handle identifies your instrument calibration session. |
![]() |
measured REF OUT frequency specifies the frequency measured at the REF OUT front panel connector. |
![]() |
measured REF OUT power specifies the power measured at the frequency indicated by the measured REF OUT frequency input. |
![]() |
error in describes error conditions that occur before this node runs. This input provides standard error in functionality. |
![]() |
instrument handle out passes a reference to your instrument session to the next VI. |
![]() |
reference clock calibration complete indicates whether the Reference Clock calibration is complete. |
![]() |
error out contains error information. This output provides standard error out functionality. |