NI-RFSG VI Reference

Table of Contents

niRFSG 5652 Adjust Reference Clock Calibration VI

  • Updated2025-02-03
  • 2 minute(s) read

niRFSG 5652 Adjust Reference Clock Calibration VI

Calculates calibration information pertaining to the 10 MHz OCXO reference of the PXI/PXIe-5650/5651/5652. This calculation is based on user-supplied measurements. The calibration information is stored in the driver session, and you can later write this information to the device.

Supported Devices: PXI/PXIe-5650/5651/5652

niRFSG 5652 Adjust Reference Clock Calibration

cio.gif

instrument handle identifies your instrument calibration session.

cdbl.gif

measured REF OUT frequency specifies the frequency measured at the REF OUT front panel connector.

cdbl.gif

measured REF OUT power specifies the power measured at the frequency indicated by the measured REF OUT frequency input.

cerrcodeclst.gif

error in describes error conditions that occur before this node runs. This input provides standard error in functionality.

iio.gif

instrument handle out passes a reference to your instrument session to the next VI.

ibool.gif

reference clock calibration complete indicates whether the Reference Clock calibration is complete.

ierrcodeclst.gif

error out contains error information. This output provides standard error out functionality.