niRFSG 5611 Adjust Impairment Calibration VI
Calculates calibration information pertaining to the impairments of the device based on user-supplied measurements. The calibration information is stored in the driver session and can later be written to the device.
Supported Devices: PXIe-5611, PXIe-5673/5673E
Related Topics
![]() |
instrument handle identifies your instrument calibration session. instrument handle is obtained from the niRFSG Initialize External Calibration VI. |
![]() |
measured RF OUT power specifies the power measured at the frequency indicated by the frequency to measure parameter in the niRFSG 5611 Configure Impairment Calibration VI. |
![]() |
error in describes error conditions that occur before this node runs. This input provides standard error in functionality. |
![]() |
instrument handle out passes a reference to your instrument session to the next VI. |
![]() |
impairment calibration complete indicates whether or not the impairment calibration is successful. If impairment calibration complete returns a value of TRUE, impairment calibration is successfully complete. |
![]() |
error out contains error information. This output provides standard error out functionality. |