Components Assumed Present
- Updated2025-10-31
- 1 minute(s) read
Test System Maintenance Software is not able to verify the details of every component in the ETX-1640x. In the Maintenance Software Report, components that the Test System Maintenance Software cannot positively identify are reported as components assumed present.
Can Continuity Checks Verify Details of the Current Transducer?
Test System Maintenance Software continuity checks cannot verify the presence or correct the connectivity of the current transducer (DS50UB-10V), which measures the output of the power supply (EA-PS 11500-30).
- Verify that only the positive wire of the power supply output passes through the center of the current transducer.
- Verify that the current transducer is the expected model and that the model matches the appropriate settings in the fixture information file.
- Verify the integrity of the control cable between the RM-26999 and the current transducer.