Site Requirements
- Updated2025-10-31
- 2 minute(s) read
Ensure that your site can accommodate requirements for the operating environment, physical characteristics, and ventilation clearances.
ETX-1630x Operating Environment
The ETX-1630x is intended for indoor use only. Ensure the location has stable environmental conditions to reduce variability in performance. Ensure the facility has cooling capacity equivalent to the maximum power consumption of the system.
| Operating temperature | 5 °C to 32 °C |
|---|---|
| Operating humidity | 10% RH to 80% RH, noncondensing |
| Pollution degree | 2 |
| Maximum altitude | 2000 m |
ETX-1630x Physical Characteristics
Ensure the location, including passageways and elevators, can accommodate the size and weight of the ETX-1630x.
| Dimensions | 1053 mm × 679 mm × 1971 mm (41.5 in. × 26.7 in. × 77.6 in.) |
|---|---|
| Maximum weight as shipped |
ETX-16301: 473.35 kg (1043 lb) ETX-16309: 468.86 kg (1033 lb) |
| Acoustic emissions | 64 dB, 1 m away |
ETX-1630x Mechanical Stability and Loading
Your rack is configured to be stable and balanced per requirements in IEC 61010-1. Installing additional equipment may increase the risk of your rack tipping over. Make sure to preserve stability when making any changes to your rack configuration.
For IEC 61010-1 compliance, the rack and all its equipment must comply with the following weight constraints, evenly distributed between each caster.
| Model | Maximum Weight |
|---|---|
| ETX-16301, ESD casters | 544.8 kg (1200 lb) |
| ETX-16309, ESD casters | 544.8 kg (1200 lb) |
Refer to Physical Characteristics for more information. Contact NI for more information about the weight of your rack as received.
Improving Rack Stability
To further improve the stability of the rack, you can tether the rack to the floor or a wall. Make sure to keep all cooling clearances open if tethering the rack. Ensure your tether is sufficient for the weight of the rack. Ensure you perform tethering in accordance with local practices. Any securement should be made between the system frame and the surrounding structure by mechanical means able to support four times the load of the system.
ETX-1630x Ventilation Clearances
Ventilation clearances are required to ensure proper airflow to the ETX-1630x. Ensure obstructions are outside the following clearance requirements.
| Left and right sides | 0 mm (0 in.) |
|---|---|
| ITA side | 0 mm (0 in.) |
| HMI side | 305 mm (12 in.) |
| Top | 508 mm (20 in.) |