ETX-1630x Power Requirements
- Updated2025-10-31
- 3 minute(s) read
ETX-1630x Power Requirements
Provide AC power connection to facility power source and connection to protective earth ground.
Power Entry Panel
The power entry panel (PEP) is the RMX-10140-50Y2P.
| Overvoltage category | II |
|---|---|
| Branch circuit rating | 63 A |
| Voltage rating |
ETX-16301: 100 V/173 V to 240 V/415 V, 3P+N+PE, 3Φ ETX-16309: 230 V/400 V, 3P+N+PE, 3Φ |
| Frequency | 50 Hz or 60 Hz |
| Current rating | 50 A |
The overall power consumption is 30 kW for the RMX-10140-50Y2P.
| Component | RMX-10140-50Y2P |
|---|---|
| Hardwired input protective enclosure screws |
Tool: 3 mm hex driver Torque: 1.3 N m (11.5 lb · in.) |
| Earth termination mechanical lug |
Tool: 3/16 hex driver Torque: 5.1 N · m (45 lb · in.) Wire gauge: 14 AWG to 2/0 AWG Wire strip length: 15.7 mm (0.6 in.) |
| M6 screw thread input terminal block |
Tool: #2 Phillips driver Torque: 2.5 N m to 3.0 N m (22.1 in. lb. to 26.6 in lb.) Wire gauge: 14 AWG to 2 AWG Wire strip length: 14.0 mm (0.6 in.) |
| Cable grommet |
Part number: Heyco M4860 Cable bundle diameter: 22.0 mm to 32.0 mm (0.9 in. to 1.2 in.) |
For the input power cable, NI recommends copper wire at least 10 mm2 (8 AWG) minimum; installation of the ETX-1630x to the facilities must be made per the local electrical code requirement, and a qualified electrical contractor should make the installation.