Move the ETX-1630x by lifting the rack from the top or by using a forklift.

Moving the Rack by Lifting from the Top

Complete the following tasks before lifting the rack from the top:

  • Power off the system.
  • Disconnect any cabling that might be disconnected or damaged from moving the rack.
  • Disconnect the AC power cable and switch the main breaker to the off position.

Equipment

  • Eyebolt (x4), NI shipping kit part number 789509-01
  1. Insert the eyebolts to the mounting holes on the top of the rack and torque the eyebolts to 6.0 N · m (53.1 in. · lb).
    Figure 19. Eyebolt


  2. Connect the four eyebolts to a crane or lifting apparatus.
    Note Ensure that the weight of the rack is balanced evenly across all four eye bolts.
  3. Using a crane or lifting apparatus, lift the rack and move it to the new location.
    Figure 20. Lifting the Rack from the Top (Generic, Empty Rack Shown for Reference)

    This image shows an empty rack being lifted from the top.

Moving the Rack with a Forklift

Complete the following tasks before moving the rack with a forklift:

  • Power off the system.
  • Disconnect any cabling that may be disconnected or damaged when moving the rack.
  • Disconnect the AC power cable and switch the main breaker to the off position.

Equipment

  • Forklift
Figure 21. Forklift Dimensions


Table 27. Forklift Dimensions
Lifting Position E h L
When lifting from the left or right side <590 mm (23.2 in.) <90 mm (3.5 in.) 670 mm (26.3 in.)
When lifting from the HMI or ITA side <450 mm (17.7 in) <90 mm (3.5 in) >670 mm (26.3 in)
  1. Place the forks under the correct side of the rack for the forklift you are using.
  2. Lift the rack and move it to the new location.
    Note Ensure that the rack is secure and stable during the move.