Configuring EMO Measure
- Updated2025-10-31
- 2 minute(s) read
The EMO Measure feature allows you to remotely shut down the rack when the mass interconnect is connected to a test adapter. Refer to the following caution for important details.
The EMO Measure behavior depends on the state of the Adapter Detect interlock (J1, pins A33 and A34). Refer to Interlocks for more information about the Adapter Detect interlock.
Use 12 VDC to 24 VDC with EMO Measure pins. The test system monitors the state of the EMO Measure pins only when the Adapter Detect interlock closes. While the Adapter Detect interlock remains closed, the following statements are true:
- The system does not power on unless DC voltage is present between pins J1, B33 and J1, B34 (EMO Measure + and EMO Measure -). However, providing DC voltage does not turn on the system. You must still push the power button to power on the system.
- The test system shuts down if the EMO Measure pins lose voltage. Shut down occurs unless the Adapter Detect interlock opens simultaneously.
- Putting DC voltage on the EMO Measure pins has no direct effect.
Related Information
- Adapter Detect Interlock
The Adapter Detect interlock is intended to be used to indicate whether an adapter is plugged into the mass interconnect.