PXIe-6674T Calibration Procedure
- Updated2025-04-30
- 3 minute(s) read
Calibrate the PXIe-6674T using Calibration Executive.
Calibration Executive Procedure Features
Verify Only |
Verify & Adjust |
Manual Mode |
Optional Adjust |
Approximate Test Time
Warmup | Verify Only | Verify & Adjust |
---|---|---|
24 hours | 1.5 minutes | 7 minutes |
Test Equipment
The following table lists the test equipment required for calibrating the PXIe-6674T.
Instrument | Recommended Model | Requirements |
---|---|---|
10 MHz clock source | Symmetricom/Datum 8040 | Accurate to within 0.75 parts-per-billion (ppb) |
BNC-SMA cable | — | — |
PXI Express chassis | PXIe-1062Q, PXIe-1075 | Any PXI Express chassis meets requirements. |
Test Conditions
The following setup and environmental conditions are required to ensure the PXIe-6674T meets published specifications.
- Keep connections to the device as short as possible. Long cables and wires act as antennas, picking up extra noise that can affect measurements.
- Verify that all connections to the device, including front panel connections, are secure.
- Maintain an ambient temperature of 25 °C ±5 °C.
- Keep relative humidity below 80%.
- Ensure the PXIe-6674T has been continuously powered on for at least 24 hours.
- Use shielded copper wire for all cable connections to the device. Use twisted-pairs wire to eliminate noise and thermal offsets.
- Ensure that the PXI chassis fan speed is set to HIGH, that the fan filters are clean, and that the empty slots contain filler panels. For more information, refer to the Maintain Forced-Air Cooling Note to Users document available at ni.com/docs.
Device Setup
- Install the PXIe-6674T board into any PXI Express slot of your PXI Express chassis. Refer to the NI PXIe-6674T Installation Guide for more details.
- Power on the PXI Express chassis.
- Configure the hardware using Measurement & Automation Explorer (MAX).
- Launch the Calibration Executive procedure. Calibration Executive guides you through the required connections.
Note The 10MHz reference clock
should be connected directly to the chassis 10MHz Ref In. No connections will be
made directly to the PXIe-6674T. This differs from the manual Calibration Procedure
but is required for calibration to be performed in any slot within the
chassis.
Test Limit Equations
The following test limits are derived from the published specifications.
Equation Type | Equation |
---|---|
Timing Accuracy | 1 year limits are calculated as follows:
|