PXIe-4353 Calibration Procedure
- Updated2025-04-30
- 4 minute(s) read
Calibrate the PXIe-4353 using Calibration Executive.
Calibration Executive Procedure Features
Verify Only |
Verify & Adjust |
Manual Mode |
Optional Adjust |
Approximate Test Time
45 minutes
Test Equipment
The following table lists the test equipment required to calibrate the PXIe-4353 device.
Instrument | Recommended Model | Requirements |
---|---|---|
Voltage calibrator | Fluke 5520A | If this instrument is unavailable for voltage accuracy verification and adjustment, use a high-precision voltage source with an accuracy of at least 70 ppm when sourcing up to 50 µA. |
Resistance calibrator* | Fluke 5520A | If this instrument is unavailable for resistance accuracy verification and adjustment, use a high-precision resistance source with an accuracy of at least 150 ppm 2-wire compensation. |
PXI Express chassis | PXIe-1062Q | — |
Calibration accessory | CAL-4353 | This accessory is used for module verification and adjustment. |
Connecting wires | Banana-jack terminated copper wire | Copper wire is used for validating the voltage accuracy and adjustment of the analog input channels. Copper wire is also used for validating the resistance accuracy and calibration of the CJC resistance input channels. |
Note *When connecting the resistance calibrator, keep the lead wires 1.0 Ω less
than the maximum lead compensation specification of the resistance calibrator. When
using the Fluke 5520A, keep the lead wires <1.1 Ω.
Note Though the Fluke 5500 has specifications matching those named in the PXIe-4353
Calibration Procedure, there are physical interactions of the DUT which cannot
properly read resistance values sourced by this calibrator at all test
points.
Test Conditions
The following setup and environmental conditions are required to ensure the PXIe-4353 meets published specifications.
- Keep connections to the PXIe-4353 device as short as possible. Long cables and wires act as antennas, picking up extra noise that can affect measurements.
- Use shielded copper wire for all cable connections to the device. Use twisted-pair wire to eliminate noise and thermal offsets.
- Maintain an ambient temperature of 23 °C ±5 °C. The module temperature will be greater than the ambient temperature.
- Keep relative humidity below 80%.
- Allow a warm-up time of at 15 minutes to ensure that the measurement circuitry
of the PXIe-4353 and CAL-4353 are at a stable operating temperature. Note If you are using the Fluke 5520A, allow the calibrator to warm up for 30 minutes and maintain an ambient temperature of 23 °C ±3 °C. Refer to the Fluke 5520A user manual for the latest set up requirements and recommendations to achieve the best accuracy.
- Temperature change affects the measurement characteristics of an instrument. To take these changes into account, the tested specifications include the effects of temperature drift. For the PXIe-4353, valid temperature drift is ±10 °C.
- Ensure that the PXI Express chassis fan speed is set to HIGH, that the fan filters are clean, and that the empty slots contain filler panels.
Device Setup
To set up the PXIe-4353 for calibration, complete the following steps:
- Install and configure the PXIe-4353 module with Measurement & Automation Explorer (MAX). Refer to the Measurement & Automation Explorer Help for DAQ for more configuration information.
- Attach the CAL-4353 to the PXIe-4353 so that the connection pane is flush with the chassis and tighten the connection screw. Ensure the connection between the two devices is stable by confirming that MAX shows the CAL-4353 in the PXIe-4353 device drop down list.
- Launch the Calibration Executive procedure. Calibration Executive guides you through the required connections.
Test Limit Equations
The following test limit is derived from published specifications.
Equation Type | Equation |
---|---|
Analog Input Voltage | VoltageAccuracyLimit = TestValue ± ((TestValue * GainPercent for
Timing Mode) + (Offset Value for Timing Mode))
|
Analog Input Offset | OffsetLimit = TestValue ± Offset Value for Timing Mode
|
CJC Resistance Accuracy |
Limits for this test are taken directly from the calibration procedure document. |