Power Entry Panel

Figure 6. Power Entry Panel (PEP) Front and Rear View

1378

Figure Callout Description Use
1 Protective Earthing Ground Stud Provides internal grounding plate connections with the external grounding lug for all rack mount equipment. High levels of leakage current may be present on the measurement rack. Connect the measurement rack to the protective earth terminal before connecting to AC power. Refer to the BTS-16101 Battery Test System Measurement Rack Safety, Environmental, and Regulatory Information document for specifications for protective earth terminal wiring.
Note The facility installation shall provide a means for connection to protective earth, and qualified personnel shall install a protective earthing conductor from the BTS-16101 protective earthing terminal to the protective earth wire in the facility.
2 Power Inlet Connector Provides a low power (IEC 60320 C20) power entry.
3 Main Breaker (CB Outlet) Provides circuit protection as a main breaker by controlling whether AC power will be allowed into the rack. If the Main Breaker switch is off, the rack cannot be powered on.
4 Ethernet Port Facilitates network connectivity to internal equipment. The Gigabit Ethernet port connects to the cRIO-9047 controller.
5 USB 3.0 Ports (x2) Provides access to extension cables within the rack. The internal USB extension cables ship unterminated for user-customizable connection to internal equipment. Use the USB cables that are included in the kit for external connection.
6 DC Output to EPO Panel Provides DC power and inhibit control to/from the EPO panel.
7 Output Connector Provides AC output to the uninterruptible power supply (UPS).
8 Grounding Plate Provides a connection point for the measurement rack ground.

The PEP also includes a line filter for EMI protection within the system.