Common Issues and Solutions
- Updated2022-11-09
- 2 minute(s) read
Common Issues and Solutions
Refer to the following table to debug common issues found in an audio and acoustics test system.
| Issue | Possible Cause | Solution |
|---|---|---|
| The resistance or the total loss of connection increases. | Some parts, such as the electronic contacts that connect a DUT to the test system, are worn. | Check the parts and replace them if worn. |
| The acquired signals are abnormal. | The following problems related to hardware components occur:
|
Provide protection for the system. For example, set the lower and upper limits for input ranges of instruments. |
| Measurement results are unstable. | Temperature significantly affects the electronic parameters of the speaker and microphones. | For speaker tests:
Before a microphone test, keep a microphone with a preamplifier on for a period specified by the manufacturer in accordance with IEC 60268-4:2018. This allows the microphone to reach the stationary temperature under rated conditions. |
| The delay is longer than expected. | Cables are too long. | Use short cables. |
| The measured noise value is high. |
|
|
| Test results are inconsistent, or repeatability issues occur. | There is much acoustic background noise or electric noise. |
|
| The phase frequency response curve is ramping up or down. |
|
|
| The distortion measured by the reference microphone is higher than expected. | The speaker generates a distorted signal. |
|
| The fluctuation of magnitude frequency response is greater than expected, especially at high frequencies. | The distance between the DUT and reference microphone is too large. | Adjust the position of the reference microphone. |
| The position of the speaker is inappropriate. | Adjust the position of the speaker. | |
| A hard surface is too close to the microphones. | Reduce the acoustic impacts of fixtures. |