Replacing a Connectorized Power Entry Panel
- Updated2024-10-21
- 3 minute(s) read
Replacing a Connectorized Power Entry Panel
- Power off the RMX-10101/10102 by flipping the main disconnect branch breaker to off.
- If your RMX-10101/10102 utilizes a UPS to provide backup power to Phase 1 powered components, press the power button on the UPS to power it down.
- Disconnect the RMX-10101/10102 from facility power.
Shipping Kit Part Number | Description |
---|---|
789480-01 | RMX-10140-16S2 Module Assembly |
789486-01 | RMX-10140-24S2 Module Assembly |
789481-01 | RMX-10140-16D2 Module Assembly |
789482-01 | RMX-10140-16Y2 Module Assembly |
Equipment
- 3 mm hex wrench
- T10 Torx head screwdriver, 0.76 N · m (6.7 lb. · in.)
- #3 Phillips head screwdriver, 1.92 N · m (17.0 lb. · in.)
- T25 Torx head screwdriver, 1.3 N · m (11.5 lb. · in.)
The diagrams in this section illustrate the RMX-10140-24S2 power entry panel (PEP). The same procedure applies for the RMX-10140-16S2 PEP, RMX-10140-16D2 PEP, and RMX-10140-16Y2 PEP.
- Disconnect the input power cable from the interface panel of the connectorized PEP.
- Remove the left side panel of the RMX-10101/10102 using the 3 mm hex wrench.
- Unscrew all retainer brackets securing the PDU power cables to the PEP using the T10 Torx head screwdriver.
- Unscrew the RCU I/O cable ring lug from the PEP using the T10 Torx head screwdriver.
- If necessary, unscrew the 3.3 V I2C cable ring lug from the PEP using the T10 Torx head screwdriver.
- Unplug all cables connected to the PEP.
-
Unscrew the ground cable ring lug from the PEP by using the #3 Phillips head
screwdriver.
- Retainer bracket
- RCU I/O cable
- Ground cable
-
Unscrew the four screws securing the PEP to the rear of the RMX-10101/10102 using the T25 Torx
head screwdriver.
- Remove the PEP from the RMX-10101/10102.
- Install the new connectorized PEP in the RMX-10101/10102 and screw in the four screws with the T25 Torx head screwdriver to secure the PEP to the RMX-10101/10102.
- Screw in the ground cable ring lug using the #3 Phillips head screwdriver.
- Connect all cables to the new PEP.
- If necessary, screw in the 3.3 V I2C cable ring lug to the new PEP using the T10 Torx head screwdriver.
- Screw in the RCU I/O cable ring lug to the new PEP using the T10 Torx head screwdriver.
- Screw in all retainer brackets to secure the PDU power cables to the new PEP using the T10 Torx head screwdriver.
- Reinstall the left side panel of the RMX-10101/10102 using the 3 mm hex wrench.
- Connect the input power cable to the interface panel of the new PEP.
-
If necessary, apply the label with the correct input voltage range on the
interface panel of the new PEP.
- Input voltage range label location