Accessing Rack Components at the SRG8 Side Space Location

If you have installed AUX DC 1200 power distribution units (PDUs) or an RMX-10153 3-Phase PDU in the SRG8 side space location in your RMX-10102, you must disconnect this hardware to access rack components at this location. Disconnecting this hardware will give the SRG8 hinge bracket full swinging clearance.

Equipment

  • T20 Torx head screwdriver, 1.30 N · m (11.5 lb. · in.)
  • T30 Torx head screwdriver, 6.00 N · m (53.1 lb. · in.)
  • 4 mm hex wrench, 5.00 N · m (44.3 lb. · in.)
  1. If you have installed AUX DC 1200 PDUs in the SRG8 side space location, unplug the AUX DC 1200 communication cable from the third AUX DC 1200 PDU.


    1. AUX DC 1200 communication cable
  2. If you have installed an RMX-10153 3-Phase PDU in the SRG8 side space location, complete the below procedure to disconnect it.
    1. Unplug the DC contactor cable from the RMX-10153 3-Phase PDU.
    2. Remove the RMX-10153 3-Phase PDU cover using the T20 Torx head screwdriver.
    3. Disconnect the cables connected to the 3-Pole 3-Phase filter using the T30 Torx head screwdriver.
    4. If necessary, disconnect the output cables from the DC contactor using the 4 mm hex wrench.


      1. DC contactor cable
      2. 3-Phase filter input cables