Overview
This tutorial is part of the National Instruments Measurement Fundamentals series. Each tutorial in this series will teach you a specific topic of common measurement applications by explaining theoretical concepts and providing practical examples.This tutorial introduces and explains the concepts and techniques of measuring strain with strain gages.
For more in-depth guidance on making strain measurements, visit the how-to guide.To find more information on the Measurement Fundamentals series, return to the NI Measurement Fundamentals Main Page.
Table of Contents
- What Is Strain?
- The Strain Gage
- Strain Gage Measurement
- Signal Conditioning for Strain Gages
- Data Acquisition Systems for Strain Gage Measurements
- Relevant NI Products
What Is Strain?
Strain is the amount of deformation of a body due to an applied force. More specifically, strain (e) is defined as the fractional change in length, as shown in Figure 1.

Figure 1. Definition of Strain
Strain can be positive (tensile) or negative (compressive). Although dimensionless, strain is sometimes expressed in units such as in./in. or mm/mm. In practice, the magnitude of measured strain is very small. Therefore, strain is often expressed as microstrain (me), which is e x 10-6.
When a bar is strained with a uniaxial force, as in Figure 1, a phenomenon known as Poisson Strain causes the girth of the bar, D, to contract in the transverse, or perpendicular, direction. The magnitude of this transverse contraction is a material property indicated by its Poisson's Ratio. The Poisson's Ratio n of a material is defined as the negative ratio of the strain in the transverse direction (perpendicular to the force) to the strain in the axial direction (parallel to the force), or n = eT/e. Poisson's Ratio for steel, for example, ranges from 0.25 to 0.3.
The Strain Gage
While there are several methods of measuring strain, the most common is with a strain gage, a device whose electrical resistance varies in proportion to the amount of strain in the device. The most widely used gage is the bonded metallic strain gage.
The metallic strain gage consists of a very fine wire or, more commonly, metallic foil arranged in a grid pattern. The grid pattern maximizes the amount of metallic wire or foil subject to strain in the parallel direction (Figure 2). The cross-sectional area of the grid is minimized to reduce the effect of shear strain and Poisson Strain. The grid is bonded to a thin backing, called the carrier, which is attached directly to the test specimen. Therefore, the strain experienced by the test specimen is transferred directly to the strain gage, which responds with a linear change in electrical resistance. Strain gages are available commercially with nominal resistance values from 30 to 3,000 Ω, with 120, 350, and 1,000 Ω being the most common values.

Figure 2. Bonded Metallic Strain Gage
It is very important that the strain gage be properly mounted onto the test specimen so that the strain is accurately transferred from the test specimen, through the adhesive and strain gage backing, to the foil itself.
A fundamental parameter of the strain gage is its sensitivity to strain, expressed quantitatively as the gage factor (GF). Gage factor is defined as the ratio of fractional change in electrical resistance to the fractional change in length (strain):
The gage factor for metallic strain gages is typically around 2.
Strain Gage Measurement
In practice, strain measurements rarely involve quantities larger than a few millistrain (e x 10-3). Therefore, to measure the strain requires accurate measurement of very small changes in resistance. For example, suppose a test specimen undergoes a strain of 500 me. A strain gage with a gage factor of 2 will exhibit a change in electrical resistance of only 2 (500 x 10-6) = 0.1%. For a 120 Ω gage, this is a change of only 0.12 Ω.
To measure such small changes in resistance, strain gages are almost always used in a bridge configuration with a voltage excitation source. The general Wheatstone bridge, illustrated in Figure 3, consists of four resistive arms with an excitation voltage, VEX, that is applied across the bridge.

Figure 3. Wheatstone Bridge
The output voltage of the bridge, VO, is equal to:
From this equation, it is apparent that when R1/R2 = R4/R3, the voltage output VO is zero. Under these conditions, the bridge is said to be balanced. Any change in resistance in any arm of the bridge results in a nonzero output voltage.
Therefore, if you replace R4 in Figure 3 with an active strain gage, any changes in the strain gage resistance will unbalance the bridge and produce a nonzero output voltage. If the nominal resistance of the strain gage is designated as RG, then the strain-induced change in resistance, DR, can be expressed as DR = RG·GF·e, from the previously defined Gage Factor equation. Assuming that R1 = R2 and R3 = RG, the bridge equation above can be rewritten to express VO/VEX as a function of strain (see Figure 4). Note the presence of the 1/(1+GF·e/2) term that indicates the nonlinearity of the quarter-bridge output with respect to strain.

Figure 4. Quarter-Bridge Circuit
Ideally, you would like the resistance of the strain gage to change only in response to applied strain. However, strain gage material, as well as the specimen material to which the gage is applied, also responds to changes in temperature. Strain gage manufacturers attempt to minimize sensitivity to temperature by processing the gage material to compensate for the thermal expansion of the specimen material for which the gage is intended. While compensated gages reduce the thermal sensitivity, they do not totally remove it.
By using two strain gages in the bridge, you can further minimize the effect of temperature. For example, Figure 5 illustrates a strain gage configuration where one gage is active (RG + DR) and a second gage is placed transverse to the applied strain. Therefore, the strain has little effect on the second gage, called the dummy gage. However, any changes in temperature affect both gages in the same way. Because the temperature changes are identical in the two gages, the ratio of their resistance does not change, the voltage VO does not change, and the effects of the temperature change are minimized. NOTE: In the Wheatstone bridge configuration, the active gage and the dummy gage should be on the same vertical leg of the bridge.

Figure 5. Use of Dummy Gage to Eliminate Temperature Effects
The sensitivity of the bridge to strain can be doubled by making both gages active in a half-bridge configuration. For example, Figure 6 illustrates a bending beam application with one bridge mounted in tension (RG + DR) and the other mounted in compression (RG - DR). This half-bridge configuration, whose circuit diagram is also illustrated in Figure 6, yields an output voltage that is linear and approximately doubles the output of the quarter-bridge circuit.

Figure 6. Half-Bridge Circuit
Finally, you can further increase the sensitivity of the circuit by making all four of the arms of the bridge active strain gages in a full-bridge configuration. The full-bridge circuit is shown in Figure 7.

Figure 7. Full-Bridge Circuit
The equations given here for the Wheatstone bridge circuits assume an initially balanced bridge that generates zero output when no strain is applied. In practice, however, resistance tolerances and strain induced by gage application generate some initial offset voltage. This initial offset voltage is typically handled in two ways. First, you can use a special offset-nulling, or balancing, circuit to adjust the resistance in the bridge to rebalance the bridge to zero output. Alternatively, you can measure the initial unstrained output of the circuit and compensate in software. This topic is discussed in greater detail later.
The equations given above for quarter-, half-, and full-bridge strain gage configurations assume that the lead wire resistance is negligible. While ignoring the lead resistance may be beneficial to understanding the basics of strain gage measurements, doing so in practice can be a major source of error. For example, consider the 2-wire connection of a strain gage shown in Figure 8a. Suppose each lead wire connected to the strain gage is 15 m long with lead resistance RL equal to 1 Ω. Therefore, the lead resistance adds 2 Ω of resistance to that arm of the bridge. Besides adding an offset error, the lead resistance also desensitizes the output of the bridge.
You can compensate for this error by measuring the lead resistance RL and accounting for it in the strain calculations. However, a more difficult problem arises from changes in the lead resistance due to temperature fluctuations. Given typical temperature coefficients for copper wire, a slight change in temperature can generate a measurement error of several microstrain.
Using a 3-wire connection can eliminate the effects of variable lead wire resistance because the lead resistance affects adjacent legs of the bridge. As seen in Figure 8b, changes in lead wire resistance, RL2, do not change the ratio of the bridge legs R3 and RG. Therefore, any changes in resistance due to temperature cancel out each other.

Figure 8. 2-Wire and 3-Wire Connections of Quarter-Bridge Circuit
Signal Conditioning for Strain Gages
Strain gage measurement involves sensing extremely small changes in resistance. Therefore, proper selection and use of the bridge, signal conditioning, wiring, and data acquisition components are required for reliable measurements. To ensure accurate strain measurements, it is important to consider the following:
- Bridge completion
- Excitation
- Remote sensing
- Amplification
- Filtering
- Offset
- Shunt calibration
Bridge Completion – Unless you are using a full-bridge strain gage sensor with four active gages, you need to complete the bridge with reference resistors. Therefore, strain gage signal conditioners typically provide half-bridge completion networks consisting of high-precision reference resistors. Figure 9a shows the wiring of a half-bridge strain gage circuit to a conditioner with completion resistors R1 and R2.

Figure 9a. Connection of Half-Bridge Strain Gage Circuit
Excitation – Strain gage signal conditioners typically provide a constant voltage source to power the bridge. While there is no standard voltage level that is recognized industry wide, excitation voltage levels of around 3 and 10 V are common. While a higher excitation voltage generates a proportionately higher output voltage, the higher voltage can also cause larger errors because of self-heating.
Remote Sensing – If the strain gage circuit is located a distance away from the signal conditioner and excitation source, a possible source of error is voltage drop caused by resistance in the wires connecting the excitation voltage to the bridge. Therefore, some signal conditioners include a feature called remote sensing to compensate for this error. Remote sense wires are connected to the point where the excitation voltage wires connect to the bridge circuit, as seen in Figure 9b. The extra sense wires serve to regulate the excitation supply through negative feedback amplifiers to compensate for lead losses and deliver the needed voltage at the bridge.

Figure 9b. Remote Sensor Error Compensation
Amplification – The output of strain gages and bridges is relatively small. In practice, most strain gage bridges and strain-based transducers output less than 10 mV/V (10 mV of output per volt of excitation voltage). With 10 V excitation, the output signal is 100 mV. Therefore, strain gage signal conditioners usually include amplifiers to boost the signal level to increase measurement resolution and improve signal-to-noise ratios.
Filtering – Strain gages are often located in electrically noisy environments. It is therefore essential to be able to eliminate noise that can couple to strain gages. Lowpass filters, when used with strain gages, can remove the high-frequency noise prevalent in most environmental settings.
Offset Nulling – When a bridge is installed, it is very unlikely that the bridge will output exactly zero volts when no strain is applied. Slight variations in resistance among the bridge arms and lead resistance will generate some nonzero initial offset voltage. Offset nulling can be performed by either hardware or software:
1. Software Compensation – With this method, you take an initial measurement before strain input is applied, and use this offset to compensate subsequent measurements. This method is simple, fast, and requires no manual adjustments. The disadvantage of the software compensation method is that the offset of the bridge is not removed. If the offset is large enough, it limits the amplifier gain you can apply to the output voltage, thus limiting the dynamic range of the measurement.
2. Offset-Nulling Circuit – The second balancing method uses an adjustable resistance, a potentiometer, to physically adjust the output of the bridge to zero. By varying the resistance of potentiometer, you can control the level of the bridge output and set the initial output to zero volts.
Shunt Calibration – The normal procedure to verify the output of a strain gage measurement system relative to some predetermined mechanical input or strain is called shunt calibration. Shunt calibration involves simulating the input of strain by changing the resistance of an arm in the bridge by some known amount. This is accomplished by shunting, or connecting, a large resistor of known value (Rs) across one arm of the bridge, creating a known DR as seen in Figure 9c. The output of the bridge can then be measured and compared to the expected voltage value. The results are used to correct span errors in the entire measurement path, or to simply verify general operation to gain confidence in the setup.

Figure 9c. Shunt Resistor Connected Across R3
Data Acquisition Systems for Strain Gage Measurements
Using NI CompactDAQ with Strain Gages
NI CompactDAQ hardware provides the plug-and-play simplicity of USB to sensor and electrical measurements. The NI CompactDAQ system consists of an NI cDAQ-9174 4-slot or NI cDAQ-9178 8-slot USB 2.0-compliant chassis that can hold up to eight C Series I/O modules and connect to a PC using a 1.8 m USB cable. NI CompactDAQ delivers fast and accurate measurements with more than 45 self-contained measurement modules available. And with three timing engines, the cDAQ-9174/9178 chassis can run analog modules at up to three different rates. Because all circuitry required for the specific measurement is contained in the C Series I/O module itself, you can connect many different types of sensors, including strain gages, directly to the modules.

Figure 10. NI cDAQ-9174/9178 Chassis with C Series I/O Modules
The NI 9219 is a 4-channel universal C Series module designed for multipurpose testing in any NI CompactDAQ or CompactRIO chassis. With the NI 9219, you can measure several signals from sensors such as strain gages, RTDs, thermocouples, load cells, and other powered sensors. The channels are individually selectable, so you can perform a different measurement type on each of the four channels. The NI 9219 uses six-position spring terminal connectors in each channel for direct signal connectivity and contains built-in quarter-, half-, and full-bridge support.
For C Series I/O modules specifically designed for the measurement of strain gages, National Instruments offers the NI 9235, NI 9236, and the NI 9237. These bridge modules contain all the signal conditioning required to power and measure bridge-based sensors simultaneously. The NI 9235 and NI 9236 have a higher channel count and include completion for quarter-bridge sensors. The NI 9237 supports up to four full- and half bridge sensors and can measure from quarter bridge strain gages using a completion accessory.
The NI 9237 can perform offset/null as well as shunt calibration and remote sense, making the module the best choice for low- to medium-channel-count strain and bridge measurements.
Recommended Starter Kit for Strain Gage NI CompactDAQ System:
1. cDAQ-9174 or cDAQ-9178 chassis
2. NI 9237 with an RJ50 cable and an NI 9949 (full and half bridge) or NI 9944/NI 9945 (quarter bridge)
3. Refer to ni.com/sensors for recommended sensor vendors
Using PXI with Strain Gages
PXI is a rugged PC-based platform that offers a high-performance, low-cost deployment solution for measurement and automation systems. PXI combines the PCI bus with the rugged, modular Eurocard mechanical packaging of CompactPCI and adds specialized synchronization buses and key software features. PXI can integrate a controller and provides up to 18 slots in a single chassis. There are timing and triggering lines on the backplane of the PXI chassis for tight synchronization of the various I/O modules. PXI Express delivers PCI Express data transfer technology to the PXI platform, increasing backplane bandwidth from 132 MB/s to 6 GB/s.

Figure 11. PXI Express Chassis with SC Express Sensor Measurement Modules
The NI SC Express family features PXI Express data acquisition modules with integrated signal conditioning for sensor measurements, such as strain gages and other Wheatstone bridge-based transducers. The NI PXIe-4330 8-channel simultaneous bridge input module offers 24-bit resolution, 0.02% accuracy and 25 kS/s per channel sample rate for high-performance strain measurements. The NI PXIe-4330 can perform quarter-, half-, and full bridge-based measurements with automatic synchronization features; the included driver software ensures tight synchronization across multiple modules and chassis with inter-channel skews as low as 5 parts per billion (PPB). This device offers per channel excitation from 0.625 to 10 V with remote sensing to compensate for error caused by resistance in the wires connecting the voltage source to the bridge. For added flexibility, built-in bridge completion (120 Ω, 350 Ω, and 1 kΩ) and shunt calibration (50 kΩ, 100 kΩ) are software-selectable on a per channel basis. The NI PXIe-4330 should be used with the NI TB-4330 front-mounting terminal block for screw terminal connectivity.
Recommended Starter Kit for Strain Gage SC Express System:
1. NI PXIe-1073 chassis
2. NI PXIe-4330 universal bridge input module with TB-4330 for connectivity
3. Refer to ni.com/sensors for recommended sensor vendors
Using SCXI with Strain Gages
National Instruments SCXI is a signal conditioning system for PC-based instrumentation applications. An SCXI system consists of a shielded chassis that houses a combination of signal conditioning input and output modules, which perform a variety of signal conditioning functions. You can connect many different types of sensors, including strain gages, directly to SCXI modules. The SCXI system operates as a front-end signal conditioning system for PC plug-in data acquisition devices (USB, PCI, and PCMCIA) or PXI data acquisition modules.

Figure 12. SCXI Signal Conditioning System
The NI SCXI-1520 is an 8-channel universal strain gage input module that offers a variety of features for strain measurements. With this single module, signals from strain, force, torque, and pressure sensors can be easily read. The SCXI-1520 also offers a programmable amplifier and programmable four-pole Butterworth filter on each channel, and simultaneous sampling with track-and-hold circuitry. In addition, the SCXI-1520 system offers a half-bridge completion resistor network in the module and a socketed 350 W quarter-bridge completion resistor. Table 1 summarizes some additional features of the SCXI-1520 that relate to strain gage measurements.
Table 1. SCXI-1520 Features for Strain Gages
| Number of channels | 8 |
| Multiplexer scan rate | Up to 333 kS/s1 |
| Amplifier gain | 1 to 1,000 |
| Excitation voltage source | 0.0 to 10.0 V in 0.635 V increments |
| Excitation current drive | 29 mA throughout excitation voltage range |
| Half-bridge completion | Yes |
| Offset nulling | Yes |
| Shunt calibration | Yes |
| Remote excitation sensing | Yes |
Recommended Starter Kit for Strain Gage SCXI Data Acquisition System:
1. USB-1600 USB Data Acquisition and Control Module for SCXI
2. NI SCXI-1000 chassis
3. SCXI-1520 with NI SCXI-1314 terminal block
4. Refer to ni.com/sensors for recommended sensor vendors
Relevant NI Products
Customers interested in this topic were also interested in the following NI products:
For more tutorials, return to the NI Measurement Fundamentals Main Page.
Download the Complete Guide to Building a Measurement System for a checklist of questions to consider for your application.
Reader Comments | Submit a comment »
Excellent - Very well written and very Helpful.
Excellent. Thank you. I'm currently preparing
for a possible interview for an industrial job
using strain gauges. The information here
will help significantly in my preparation.
Thanks once again.
-
Mike Giles,United Kingdom.
mikecgiles@gmail.com
- Aug 27, 2012
A very clear and precise explanation!
I just would like to add that (as far as I
know) the strain gages only work within
the elastic limit of the metal.
Thanx!
- Apr 19, 2012
Good Job !!
Helpful as preliminary document for
professional also helpful for r&d job but at
prel. level. Thanks
-
R.K.ROY.CHOUDHURY,INTEGRATED ELECTROLIFE SYSTEM.
iesinstruments@gmail.com
- Mar 21, 2012
excellent tutorial, my question is, does the
analysis done for the bridge assuming R1 =
R2 and R3 = R4 above, delivers the same if
we assume that R1 = R4 and R2 = R3?
-
JOHN F CAIPA,UPTC.
johncaipa2@yahoo.es
- Jan 25, 2012
Good Job!
-
mike cappua,Dallas Painters.
mikecappua@gmail.com
- Jul 21, 2011
no AC excitation
NI should provide a c-module that
provide ac excitation. This would allow
the use of torque sensors.
-
Bobby Caraway,Eaton.
robertacaraway@eaton.com
- Jun 25, 2011
marvelouus
wow thank a lot Ni. your products help me so
much!
-
handoko,university of indonesia.
dhwongtegal@gmail.com
- Feb 20, 2011
Gage vs Gauge
To the commenter about UK vs US spelling,
for some reason in the US, "Gage" is the
spelling associated with strain gages but you
see that spelling "Gauge" about everywhere
else. For instance: American Wire Gauge.
-
Nic Linley,Wendell Hull & Associates.
nic@wendellhull.com
- Feb 14, 2011
Tension vs Compression, sign convention
It appears that NI reads tension and
compression opposite the typical sign
convention of tension being positive. Also,
the cantilever beam example shows it both
ways. . .
-
jrichardson@cel-tci.com
- Feb 02, 2011
Gage vs Gauge
Gage, I would expect, is the USA english
spelling of the English word gauge. It is
not unusual for the USA spelling to leave
out letters in English words they
consider superfluous (in a phonetic
sense), e.g. UK colour = USA color etc.
Not worth insisting on either, as long as
you know they mean the same thing. Of
course if I was writing for a UK audience
I would use gauge.
- Jan 31, 2011
thanks this was helpful
-
,.
- Nov 13, 2010
in response to comment on - Jan 14, 2010
Apparently you have not read a lot of
literature on strain gage's recently. If you
had, you would have realized that all
manufacturers and text books that discuss
strain gages spell it as "gage" not "gauge".
This is an industry standard that I have seen
since I started learning about strain gages.
- Nov 07, 2010
Measuring Strain is excellent item
discussed in here, full comprehensive
and easy to understand. Thanks NI !
-
Mohammad Kaiful Wara Khan,Construction Technical Services Arabia Limited, KSA.
mohammad.warakhan@aramco.com
- Jul 22, 2010
Good refresher
Very good refresher.
Shame you spelt gauge wrong.
A "gage" is a challenge or pledge, not a
measure or measuring device.
- Jan 14, 2010
Thanks a lot , its very good summary .... help
me a lot ....
-
Ali,Umea.
anon12707596
- Dec 05, 2009
its informative and self explanatory
article on strain gaug measurement
Cheers up!
-
Parmeshwar Karad,Sinhgad College of Engineering.
karadpa@gmail.com
- Sep 11, 2008
very informative article
very helpful article on priliminary
learning about strain guages. cheers!!
great work
-
harminder saini,Student, brunel university.
bs07hss@brunel.ac.uk
- Jul 08, 2008
Awesome!
- Apr 16, 2008
An extremely helpful and self explanatory article
-
ahsan dar,PN Engineering College,NUST.
ahsandar@hotmail.com
- Dec 18, 2007
most useful information
This is most useful information for me
and my student because Iam teaching
manufacturing automation subject
-
SONAWANE V. Y. ,AISSMS COLLEGE OF ENGG NEAR RTO KENEDY ROAD PUNE-1.
sonawanevy@yahoo.co.in
- Oct 26, 2007
i have a 3 hour exam in the morning. superb
refresher. Cheers
- May 13, 2007
very useful
Thank you for providing great
information about strain guage sensor
which helped me a lot in designing this
sensor practically.
-
vamshidhar.smt@gmail.com
- Apr 05, 2007
Thank You Much. I can now start my
tech paper. Just what I needed to get
started
-
Starving Student,.
- Apr 03, 2007
Need numerical results
Information provided is very good and
useful.Come up with practical results
eg:what is the range of usability of the
instrument,sensitivity etc
-
vamshidhar s,IIT Kharagpur.
vamshidhar.smt@gmail.com
- Mar 18, 2007
this information about strain gauge is
not enough . this is just copy from book.
i thing u have done it practically.
-
jassi,infotech.
jaspreetd@rediffmail.com
- Mar 13, 2007
Comprehensive article
Thank you for this elaborative content. It
gives me a kick start for an indepth
strain measurement activities.
-
Demissie Solomon,Ethiopian Airlines.
solomond@ethiopianairlines.com
- Dec 20, 2006
Self explainatory
From basic to higher end , it is very
technical.
-
awachar@yahoo.com
- Dec 05, 2006
Diverse, robust, and thorough. Easy to
read and practical.
- Nov 02, 2006
Brilliant!
Very basic, yet a comprehensive article.
5/5
-
Usman Y. Bhatti,All Sensors.
usmanyasin@gmail.com
- Oct 10, 2006
What a great work!!
Thanks for your information!!it's very
useful for me.thanks!!
-
Ajmal Khan,International Islamic University Malaysia,Kuala Lumpur.
ir_mal001@yahoo.com
- Sep 30, 2006
thaaaaaaaaanx
-
Charanpreet Singh,.
rahelcharanpreet@raediffmail.com
- Sep 24, 2006
excellent
wonderfull explanation.covers basics
very well
-
ritula thakur,Faculty in NITTTR,chandigarh.
ritula_212yahoo.com
- Jun 05, 2006
Keep it up!
Great job, helpful with school research
project
-
Kaycee,Coventry University, UK.
masterc34@yahoo.co.uk
- May 04, 2006
helpful but needs more examples on how to calculate resistance values.
-
Louis,McGill U.
- Apr 20, 2006
Perfect theory about concepts of electrical strain gauge
It is a very nice effort covering the basic
concepts of electrical strain gauge. Your
effort became very useful for me. Its a
perfect tutorial.
-
Haris Aslam,University of Engineering and Technology Lahore Pakistan.
haris_uetian@yahoo.com
- Apr 14, 2006
i think ...it is the best document on
gauges available on the internet
-
brij,iit kharagpur.
- Apr 10, 2006
Informative on an introductory level
-
Damien Lea,Aerospace Engineering Ltd.
damienlea@yahoo.com
- Mar 25, 2006
very helpful...
-
gilbert,Copenhagen University.
- Mar 17, 2006
strain measurement fundas
The article gives requisite optimum and
information on the measurement system that
is practical in application without
flambuayant add ons.
-
v.k.khanna,Intertech Sales Corporation.
intertec@bol.net.in
- Mar 07, 2006
-
NICE & GOOD,.
- Mar 05, 2006
Man That was gr8.. we need this stuff to do
our Main Project.. a teleremote controlled
robotic hand.. THANX!!!
-
Anoop Kumar,NSS College of engineering.
- Feb 21, 2006
great
-
Ubaier,Coventry University, UK.
- Jan 13, 2006
Clear and useful
-
Kate,.
- Dec 16, 2005
Thankyou for this resourse.
- Oct 31, 2005
very good!! thanks for the valuable
information!! =)
-
teo,.
- Aug 31, 2005
really informative.good work
-
sanju thomas,vellore institute of technology.
sanjuthom@gmail.com
- Jul 13, 2005
excellent
This document is very much helpful for me to
do my project work, thanks to NI
-
anbalagan,PSG College of Technology,tamilnadu,india.
anbuxyz@yahoo.co.in
- May 12, 2005
basics covered well
basics covred with visulaization help.Good
for preliminary info.
-
Ravi,IIT,Kharagpur,.
thebroadmind@rediffmal.com
- May 03, 2005
Really Good!
Very Well Written. Keep It Up
-
Kiran,.
- May 02, 2005
doing M.tech from IIT kharagpur
very good site giving useful information...
thank you
-
rajani kanth gandla,student.
grajanikanth@yahoo.com
- Apr 21, 2005
Great help with my primary school project
Great help with my primary school project -
thanks!
-
The child prodigy,.
- Mar 08, 2005
How very spiffing.
I really enjoyed this "web page" on the
Intranet. I am looking into how 'phones work
and this page was really useful to me. Thank-
you so much for creating such a useful site.
-
Jim,.
- Jan 20, 2005
Seems brilliant. Just what I need.
- Jan 18, 2005
Good work on theory. However, more
description on how to configure with your
system would be helpful i.e SCXI 1314
wiring diagram.
-
,.
- Aug 18, 2004
Clear and concise. Could benefit from coverage of practical considerations - how to measure the right strain
- Aug 18, 2004
A strain free coverage of the subject.
-
Moses Zulfikar,Domino Printing UK.
mzulfikar@waitrose.com
- Aug 05, 2004
Great Help for my project!!!
-
Edwin,.
- Jun 07, 2004
Great summary; please add cFP DAQ!
This great summary on signal conditioning
for strain gauges will be complete when You
add FP and cFP to the already covered SCXI
and SCC DAQ options!
-
Richard Larker,Indexator AB.
Richard.Larker@indexator.se
- May 24, 2004
Thanks
Great help with the 2004 AS Physics
Practical Assesment 19th May
-
Xylomn,.
xylomn@hotmail.com
- May 18, 2004
great help with physics AS planning exercise
supplied a lot of the information I needed
to find for my AS Physics planning exercise.
Thanks a lot
-
Sarah Osborne-White,.
miniarmygirl@hotmail.com
- May 16, 2004
great help with school project
not too wordy. told me everything I needed
to know. thank you
-
Iain Stenson,.
iaianstenson22@hotmail.com
- May 10, 2004
Thank you
Just done a HNC project on this and i'd
like to say this cleared up several issue
and gave more background to my write up
-
Anon,.
- May 09, 2004
Very clear and helpful
-
Graham,.
- May 08, 2004
Very informative, but...
Very informative of the electrical and
measurement aspects, although being number 1
google result, I think there'll be a lot of
quarter bridges used.
This being for A-Level physicss coursework.
The main downside is it doesn't advise me
how to fix a strain guage to a surface. With
Super Glue?
-
Chiral,member.lycos.co.uk/chiralutd/phpBB2/index.php.
chiralutd@yahoo.co.uk
- Mar 23, 2004
A thorough approch
A thorough scientific and emperical approch.
Keep it up !!!
Kind regards
Nicholas Lekatsas-Lucas
Senior electronic R/D Eng.
-
Nicholas Lekatsas-Lucas,ICE/HT-FORTH.
nicklek@iceht.forth.gr
- Nov 15, 2002
excellent coverage of the electronic theory behind strain gauge measurement
Question:
I'm searching for a practical, unbiased
performance comparison between tension
measurement devices that use silicon-type
strain gauges versus those that use foil-type
strain gauges. What are the advantages of one
technology over the other?
Could you point me toward any research that's
been done?
-
Mark Breen,Dover Flexo Electronics.
mbreen@dfe.com
- Oct 08, 2002
good but improvements possible
This is good but theres room for info about
its applications and advantages/comparisons
to other methods
- Sep 22, 2002
Very clear and concise - good job
- Jul 09, 2002
excellent presentation helping to understand
the process of selecting the
right meter!
-
mstraka@scientech.com
- Apr 17, 2002
hi, im an electromechanical engineer student, the information given here was
great, i just think that you have to translate it, at 2 or 3 languajes more.. i
did in spanish =).
tanx
-
,.
crdh@pa.inter.net
- Aug 11, 2001
Legal
This tutorial (this "tutorial") was developed by National Instruments ("NI"). Although technical support of this tutorial may be made available by National Instruments, the content in this tutorial may not be completely tested and verified, and NI does not guarantee its quality in any way or that NI will continue to support this content with each new revision of related products and drivers. THIS TUTORIAL IS PROVIDED "AS IS" WITHOUT WARRANTY OF ANY KIND AND SUBJECT TO CERTAIN RESTRICTIONS AS MORE SPECIFICALLY SET FORTH IN NI.COM'S TERMS OF USE (http://ni.com/legal/termsofuse/unitedstates/us/).
