Refer to the complete list of required equipment in the Calibration Executive Help File. For your convenience, summary lists are provided below. For all PXI devices, you will need a PXI chassis and controller. NI recommends using a PXI-1042 or PXI-1042Q chassis with a MXI controller for maximum temperature control and portability.
SCXI (Traditional DAQ)
For SCXI-1100, SCXI-1120/D, SCXI-1121, and SCXI-1124:
- SCXI chassis (e.g., SCXI-1000 or SCXI-1001)
- The PXI-1011 will not work.
- The PXI-1010 will work, but you may not use the internal signal bus. Instead, use external connections.
- SCXI-1349 68-pin adapter
- CB-50 and 50-pin ribbon cable
- 6030E/6031E/6032E/6033E MIO device (6031E/6033E use SH1006868 cable)
- SH6868 or SH1006868 cable to connect to E Series device
- Appropriate terminal block
| SCXI Module | SCXI Terminal Block |
|---|---|
| SCXI-1100 | SCXI-1300 |
| SCXI-1120 Series | SCXI-1320 |
| SCXI-1121 | SCXI-1320 |
| SCXI-1124 | SCXI-1325 |
- Standards
| Instrument | Specifications | Recommended Instrument |
|---|---|---|
| Calibrator | Accurate to 50ppm | Fluke 5700A |
| DMM | 6 1/2 digits | NI 407x |
SCXI (DAQmx)
For SCXI-1102/B/C, SCXI-1104/C, SCXI-1125, SCXI-1520, and SCXI-1600:
- SCXI chassis (e.g., SCXI-1000 or SCXI-1001)
- The PXI-1011 will not work.
- The PXI-1010 will work, but you may not use the internal signal bus. Instead, use external connections.
- SCXI-1349 68-pin adapter
- CB-50 and 50-pin ribbon cable
- 628x MIO device
- SH6868 or SH1006868 cable to connect to E Series device or SHC68-68-EPM to connect to an M Series device.
- Appropriate terminal block
| SCXI Module | SCXI Terminal Block |
|---|---|
| SCXI-1102 Series | SCXI-1300 |
| SCXI-1104 Series | SCXI-1300 |
| SCXI-1125 | SCXI-1320 |
| SCXI-1520 | SCXI-1314 |
- Standards
| Instrument | Specifications | Recommended Instrument |
|---|---|---|
| Calibrator | Accurate to 50ppm | Fluke 5700A |
| DMM | 6 1/2 digits | NI 407x |
SC Series
For NI 4220:
- SMB-to-BNC cable
- 9-pin D-SUB connectors for connecting external signals to the analog ports.
- Standards
| Instrument | Specifications | Recommended Instrument |
|---|---|---|
| Calibrator | Accurate to 50ppm | Fluke 5700A |
| DMM | 6 1/2 digits | NI 407x |
Digital Multimeters
For NI 4050 and NI 4060:
| Instrument | Required Instrument | Specifications |
|---|---|---|
| Calibrator | Fluke 5700A | Accurate to 50 ppm |
For NI 4065, NI 4070, NI 4071, and NI 4072:
- Two sets of Fluke 5440 low thermal electromotive force (EMF) copper cables
- Pomona 5145 insulated double banana plug shorting bar
- Two Pomona B-4 banana-to-banana patch cables or similar banana-to-banana cables with a length less than 4 inches
- 25 Ohm, 125 Ohm, 5 kOhm, and 100 KOhm resistors (for NI 4072 only)
- 270 pF, 1 nF, 100 nF, 10 µF, and 1000 µF standards (for NI 4072 only)
| Instrument | Required Instrument | Specifications |
|---|---|---|
| Calibrator | Fluke 5700A/5720A | Calibrated within the last 90 days with DC current output characterized |
- Optional equipment for verifying frequency performance of the NI 4071
| Instrument | Optional Instrument | Specifications |
|---|---|---|
| Pulse Train Generator | NI PXI-6608 | 1 Hz to 500 kHz (0.01 % accuracy) |
| Terminal Block | NI TB-2715 | |
| Shielded Cable | NI SH68-68D1 | |
| Double Banana Plug with Strain Relief | Pomona MDP 4892 | |
| Coaxial cable | RG178 | |
| Resistor Reference Standard | IET Labs SRL-1G | 1 GΩ |
Power Supplies & SMUs
For NI PXI-4110:
| Instrument | Required Model | Specifications |
|---|---|---|
| DMM | NI 4071 | If this instrument is unavailable, use a DMM that meets the following specifications:
|
| External Load | Clarostat 240C | If this instrument is unavailable, use a power resistor decade box with a range of 4 to 48,000 Ω and an accuracy of ±10%. |
| Auxiliary Power Supply | NI APS-4100 | Use a 5 A power supply with a range of 11 to 15.5 V. |
For NI PXI-4130:
| Instrument | Required Model | Specifications |
|---|---|---|
| DMM | NI 4071 | If this instrument is unavailable, use a DMM that meets the following specifications:
|
| External Calibrator | Fluke 5700a/5720A | - |
| Auxiliary Power Supply | NI APS-4100 | Use a 5 A power supply with a range of 11 to 15.5 V. |
Dataloggers
For NI 4350 and NI 4351
- 4350/4351 calibration fixture - pn 778031-03 (not included with Cal Exec)
- SH6868 cable (except DAQCard 4350)
- For DAQCard 4350, use PSH32-30F cable (part number 183965-01)
- Standards
| Instrument | Specifications | Recommended Instrument |
|---|---|---|
| Calibrator | Accurate to 10ppm | Fluke 5700 |
| DMM | 6 1/2 digits | NI 407x |
Dynamic Signal Acquisition
For NI 4451, NI 4452, NI 4551, NI 4552, NI 4461, NI 4462, NI 4472, NI 4474, NI 4495, NI 4496, and NI 4498:
- BNC-2140 (for NI 445x, NI 455x)
- TBX-68 (for NI 445x, NI 455x)
- Analog cable (for NI 445x, NI 455x)
- Digital cable (for NI 445x, NI 455x)
- For PXI-4472, install in either:
- a PXI chassis controlled through a PXI controller.
- a PXI chassis controlled through a MXI-4 interface.
- Standards
| Instrument | Specifications | Recommended Instrument |
|---|---|---|
| Calibrator | Accurate to 0.1% AC | Fluke 5700 |
| DMM | 6 1/2 digits | NI 407x |
| Counter | Accurate to 6ppm | Agilent 53131A |
| Function Generator | Accurate to 6ppm | Agilent 33250 |
High-Speed Digitizers
For NI 5102, NI 5105, NI 5112, NI 5114, NI 5122, NI 5124, NI 5132, NI 5133, NI 5142, NI 5152, NI 5153, NI 5154, NI 5622, NI 5900, NI 5911 and NI 5922:
- 50-Ohm terminator
- Standards
| Instrument | Specifications | Recommended Instrument |
|---|---|---|
| Calibrator | Accurate to 0.005% DC |
Fluke 5700 (required for NI 5911 and NI 5922) |
| Scope Calibrator | 2% amplitude flatness | Fluke 9500B |
| DMM | 0.25% resistance |
NI 407x (only required for NI 5112) |
| Signal Generator | Frequency Range: 900 kHz to 820 MHz Power Measurement Level Setting Range: –4 dBm to 10 dBm Frequency Accuracy: ±0.1 ppm |
R&S SMA-B103 with SMA-B22 option (only required for NI 5622) |
| Power Meter | Frequency Range: 900 kHz to 820 MHz Power Measurement Range: –4 dBm to 3 dBm Maximum VSWR: 1.11 (23 ±5 °C) Absolute Power Measurement Uncertainty: 0.081 dB (23 ±5 °C) Relative Power Measurement Uncertainty: 0.05 dB (23 ±5 °C) |
R&S NRP-Z91 (only required for NI 5622) |
FieldPoint
For FP-AI-100/102/110/111/112, FP-AO-200/210, FP-AIO-600/610, FP-RTD-122/124 and FP-TC-120:
- FP-1000 Network Module
- FP-TB-CAL FieldPoint Calibration Base (part number 777519-90)
- 11-24 VDC power supply
- RS-232 serial cable
- Standards
| Instrument | Specifications | Recommended Instrument |
|---|---|---|
| Calibrator | 10 ppm |
Fluke 5700A (required for AI module) |
| DMM | 6 1/2 digits |
NI 407x (required for AO modules) |
- Standards for FP-RTD-122
| Instrument | Specifications |
|---|---|
| 330 Ω | 0.025% tolerance |
| 100 Ω | 0.025% tolerance |
| 3.3 k Ω | 0.025% tolerance |
| 1.0k Ω | 0.025% tolerance |
- Standards for FP-RTD-124
| Instrument | Specifications |
|---|---|
| 100 Ω | 0.05% Tolerance |
| 330 Ω | 0.05% Tolerance |
Compact FieldPoint
For cFP-AI-100/102/110/111/112/118, cFP-AO-200/210, cFP-AIO-600/610, cFP-RTD-122/124 and cFP-TC-120:
- cFP-CAL-a Compact FieldPoint network interface designed for calibrating I/O modules
- cFP-CB-1 connector block for wiring
- 11-30 VDC power supply
- RS-232 serial cable
- Standards
| Instrument | Specifications | Recommended Instrument |
|---|---|---|
| Calibrator | 10 ppm |
Fluke 5700A (required for AI module) |
| DMM | 6 1/2 digits |
NI 407x (required for AO modules) |
- Standards for cFP-RTD-122
| Instrument | Specifications |
|---|---|
| 330 Ω | 0.025% tolerance |
| 100 Ω | 0.025% tolerance |
| 3.3 k Ω | 0.025% tolerance |
| 1.0k Ω | 0.025% tolerance |
- Standards for cFP-RTD-124
| Instrument | Specifications |
|---|---|
| 100 Ω | 0.05% Tolerance |
| 330 Ω | 0.05% Tolerance |
Multifunction Data Acquisition
For B/E/M/S/X Series:
- E Series calibration fixture - pn778056-03. This fixture is not included with Cal Exec. Only revision B works with the S Series products while both rev A and rev B work with E/M/B Series products.
- Standards
| Calibrator | DMM | Counter | |
|---|---|---|---|
| 16/18-bit Converters | Fluke 5700 - accurate to 10ppm | 6 1/2 digit – NI 407x | Accurate to 0.01% |
| 12-bit Converters | Fluke 5500 - accurate to 50ppm | 6 1/2 digit - NI 407x | Accurate to 0.01% |
- Cables and Accessories
| Device | Connector |
|---|---|
| E/S Series devices with a 68-pin SCSI II connector | SH68-68-EP |
| E/S/M Series devices with 68-pin VHDCI connectors (for example, DAQCard-6036E, NI 6254) | SHC68-68-EP, SHC68-68-EPM, SHC68-68-S |
| E/S Series devices with a 100-pin connector (for example, NI 6025E, NI 6031E, NI 6033E, NI 6071E) | SH1006868.Connect the 68-pin cable labeled MIO-16 to the accessory. The 68-pin cable labeled Extended I/O remains unconnected. |
| M/B Series devices with a 37-pin connector (for example, PCI-6010, NI 6221) | CB-37F-LP, SH37F-37M-1 |
Counter/Timers - NI 660x
For NI 6601, NI 6602 and NI 6608:
- SH68-68-D1 cable
- SCB-68 breakout box
- Highly stable 10 MHz TTL clock source. NI recommends a rubidium frequency standard clock source.
- Standards
| Device | Specifications |
|---|---|
| NI 6608 | accurate to 0.75 ppb |
| NI 6601/6602 | accurate to 1 ppm |
Analog Output
For NI 670x, NI 671x, NI 672x and NI 673x:
- CB-68 connector block
- SH68-68-EP cable for NI 671x/673x (except for the NI 6715)
- SHC68-68-EP cable for NI 6715
- SH68-68-S cable for NI 672x
- Standards
| Calibrator | DMM | Counter | |
|---|---|---|---|
| NI 673x | Fluke 5700 - accurate to 10ppm | 6 1/2 digit - NI 407x | Accurate to 0.01% |
| NI 671x | Fluke 5500 - accurate to 50ppm | 6 1/2 digit - NI 407x | Accurate to 0.01% |
A calibrator is not needed to calibrate the NI 670x
Signal Generators
For NI 5402, NI 5406, NI 5412, NI 5421, NI 5422, NI 5441 and NI 5450:
- Standards
| Instrument | Recommended Model | Specifications |
|---|---|---|
| Digital multimeter (DMM) | NI PXI-4070 | DC accuracy =±50 ppm Resolution =1 µV |
| Counter | Agilent/HP 53131A | Ability to measure 10 MHz or higher sine waves Frequency accuracy to ±500 ppb |
| Digital Oscilloscope | Tektronix DPO70404 |
Analog Bandwidth: ≥ 4 GHz (–3 dB) (only required for NI 5450) |
| Power Meters (Two) | R&S NRP-Z91 |
VSWR (50 kHz to 120 MHz): ≤ 1.11 (only required for NI 5450) |
| Spectrum Analyzer | R&S FSU26 #Serial Number 20xxxx and above with improved phase noise: • FSU-B23 20 dB Preamplifier • FSU-B25 Electronic Attenuator |
Frequency Accuracy: ≤ 100 Hz Requirements for the following parameters must be better than or equal to the equipment recommended for f ≤ 200 MHz: (only required for NI 5450) |
- Connectors
- Male banana to female BNC adapter
- Male BNC to female SMB cable-50 O, RG-22
RF Products
For NI 5600:
| Instrument | Required Model | Specifications |
|---|---|---|
| RF Signal Generator | Agilent ESG-4422B (with option UNB, high-output power, lockable to external frequency reference), Agilent ESG-4433B (with option UNB, high-output power, lockable to external frequency reference), Agilent 8340A | Frequency range: 20 MHz to 2.8 GHz Amplitude range: –30 to +16 dBm Phase noise/Spectral purity: <–95 dBc/Hz at 1 kHz |
| Power Meter | Anritsu ML2438A, Agilent E4419B | Frequency range: 10 kHz to 110 GHz, sensor-dependent Power measurement range: –70 to +47 dBm (0.1 nW to 50 W), sensor-dependent Voltage measurement range: 0.00 to 20.00 V, nominal 1.00 mW power reference: 50 MHz nominal, ±1.2% within 1 year, VSWR <1.04 |
| (2) Power Meter Sensors | Anritsu MA2421A Thermal Sensor, Agilent E9304A Power Sensor | Range of IF output power sensor on CH-A: –30 to +20 dBm Range of frequency of power sensor on CH-A: 3 to 27 MHz Standard uncertainty (including mismatch)*: 0.07 dB |
| Anritsu MA2473A Power Sensor, Agilent E4412A Power Sensor | Range of RF input power sensor on CH-B: –70 to +20 dBm Range of frequency of power sensor on CH-B: 20 MHz to 2.8 GHz Standard uncertainty (including mismatch)†: 0.11 dB | |
| Frequency Counter | Agilent 53131A (lockable to external frequency reference) | Frequency range: 10 MHz Frequency resolution: >.5 Hz at 10 MHz |
| Rubidium Frequency Standard | Symmetricom 8040 | Frequency: 10 MHz Frequency accuracy: 1 ppb (typically ±6E-10 within 1 year) Amplitude range: >5 dBm |
| Broadband Precision Power Splitter | Weinschel 1507R | Frequency range: DC to 3.0 GHz Output maximum VSWR: 1.15 Input maximum VSWR: 1.25 Amplitude tracking: <0.1 dB** Phase tracking: <4 Degrees Insertion loss: 6 dB, nominal |
| 10 MHz Distribution Unit | Symmetricom 6502B | RF input: One, 10 MHz RF outputs: Three or more, 10 MHz Nominal output: 1 Vrms |
| PXI Chassis | NI PXI 1042, PXI-1042Q | - |
| Chassis Controllers | NI MXI-4 kit | - |
| SMA Torque Wrench | NI RF Torque Wrench | 1 N · m |
| SMA Cable | - | Use a male-male, instrument grade, 50 Ω coaxial cable that is ≤1 m in length |
| N-type to SMA Cable | - | Use a male-male, instrument grade, 50 Ω coaxial cable that is ≤1 m in length |
| SMA Adapter | - | Use a male-male, instrument grade, 50 Ω adapter with a VSWR ≤1.1. |
| N-type to SMA Adapter | - | Use a N-type female-SMA male, 50 Ω adapter with a VSWR ≤1.12. |
*With adapter standing wave ratio (SWR) ≤ 1.12 and NI 5600 IF output SWR ≤ 1.2.
†Including splitter and adapter uncertainty: 0.11 dB standard uncertainty when device under test (DUT) on other side of the splitter has SWR ≤ 2.5.
**Screened for <0.1 dB tracking (Weinschel 1507R specification <0.2 dB).
For NI 5601:
| Instrument | Recommended Model | Specification | Value | |
|---|---|---|---|---|
| Power Meter | Anritsu ML2438A | Display Resolution | <0.01 dB | |
| Settling | <0.1% | |||
| Instrument Accuracy | <0.5% | |||
| Noise | <0.5% full-scale (lowest range) | |||
| Zero Set and Drift | <0.5% full-scale (lowest range) | |||
| Reference Power Uncertainty | <0.9% | |||
| Reference Output VSWR | <1.04:1 | |||
| (2) Power Sensors | Anritsu MA2473D | Input VSWR | 10 MHz to 50 MHz | <1.9:1 |
| 50 MHz to 150 MHz | <1.17:1 | |||
| 150 MHz to 2 GHz | <1.12:1 | |||
| 2 GHz to 12.4 GHz | <1.22:1 | |||
| Linearity | –70 dBm to 20 dBm | <1.8% | ||
| Rise Time | <4 µs | |||
| Calibration Factor Uncertainty | 50 MHz | <1.48% | ||
| 100 MHz | <1.37% | |||
| 300 MHz | <1.42% | |||
| 500 MHz | <0.96% | |||
| 1 GHz | <0.99% | |||
| 2 GHz | <1.04% | |||
| 3 GHz | <1.05% | |||
| 4 GHz | <1.64% | |||
| 5 GHz | <1.39% | |||
| 6 GHz | <1.45% | |||
| 7 GHz | <1.26% | |||
| Signal Generator | NI PXI-5652 | - | - | |
| (LO Source) | ||||
| Signal Generator(RF Source) | Anritsu MG3692B | Frequency Range | 10 MHz to 6.6 GHz | |
| Power Level | –60 dBm to 20 dBm | |||
| Spectrum Analyzer | Rohde & Schwarz FSU26 | Frequency Range | 10 MHz to 6.6 GHz | |
| Power Level | –60 dBm to 20 dBm | |||
| Resolution Bandwidth | 10 Hz to 1 MHz | |||
| Phase Noise | 100 Hz | –100 dBc/Hz | ||
| 1 kHz | –120 dBc/Hz | |||
| 10 kHz | –130 dBc/Hz | |||
| 100 kHz | –130 dBc/Hz | |||
| 1 MHz | –142 dBc/Hz | |||
| Chassis | NI PXIe-1065, | - | - | |
| NI PXIe-1075 | ||||
| Chassis Controller | NI MXI-Express Kit | - | - | |
| SMA (m) to SMA (f) Right Angle Adapter | Huber+Suhner 53_SMA-50-0-2/111_N | - | - | |
| 3.5 mm (m) to 3.5 mm (m) Adapter | Huber+Suhner 32_PC35-50-0-2/199_NE | - | - | |
| (2) 50 Ω Terminators | NI SMA 50 Ω Termination Plug | - | - | |
| 3.5 mm (f) Power Splitter (2 Resistor Type) | Aeroflex 1593 | - | - | |
| (2) 36 in SMA (m) to SMA (m) Cables | Huber+Suhner ST-18/SMAm/36 | - | - | |
| 3.5 mm (m) to 3.5 mm (f) 30 dB Attenuator | Huber+Suhner 6630_SMA-50-1/199_NE | - | - | |
| SMA Torque Wrench | Agilent 8710-1582 | - | 5 in/lb | |
| 3.5 mm Torque Wrench | Agilent 8710-1765 | - | 8 in/lb | |
For NI 5610:
| Instrument | Required Model | Specifications |
|---|---|---|
| RF Signal Generator | Agilent ESG-4422B (with option UNB, high-output power, lockable to external frequency reference), Agilent ESG-4433B (with option UNB, high-output power, lockable to external frequency reference), Agilent 8340A |
Frequency range: 20 MHz to 2.8 GHz Amplitude range: –50 to +16 dBm Phase noise/Spectral purity: <–95 dBc/Hz at 1 kHz |
| Power Meter | Anritsu ML2438A, Agilent E4419B |
Frequency range: 10 kHz to 110 GHz, sensor-dependent Power measurement range: –70 to +47 dBm (0.1 nW to 50 W), sensor-dependent Voltage measurement range: 0.00 to 20.00 V, nominal 1.00 mW power reference: 50 MHz nominal, ±1.2% within 1 year, VSWR <1.04 |
| 2 Power Meter Sensors | Anritsu MA2421A Thermal Sensor, Agilent E9304A Power Sensor |
Range of IF output power sensor on CH-A: –30 to +20 dBm Range of frequency of power sensor on CH-A: 10 kHz to 27 MHz Standard uncertainty (including mismatch)*: 0.07 dB |
| Anritsu MA2473A Power Sensor, Agilent E4412A Power Sensor |
Range of RF input power sensor on CH-B: –70 to +20 dBm Range of frequency of power sensor on CH-B: 20 MHz to 2.8 GHz Standard uncertainty (including mismatch)†: 0.11 dB |
|
| High-Speed Digitizer | NI PXI-5122 | - |
| Arbitrary Waveform Generator | NI PXI-5421 | - |
| Frequency Counter | Agilent 53131A (lockable to external frequency reference) |
Frequency range: 10 MHz Frequency resolution: >.5 Hz at 10 MHz |
| Rubidium Frequency Standard | Symmetricom 8040 | Frequency: 10 MHz Frequency accuracy: 1 ppb (typically ±6E-10 within 1 year) Amplitude range: >5 dBm (typical: 7 dBm) |
| Broadband Precision Power Splitter | Weinschel 1507R | Frequency range: DC to 3.0 GHz Output maximum VSWR: 1.15 Input maximum VSWR: 1.25 Amplitude tracking: <0.1 dB** Phase tracking: <4 Degrees Insertion loss: 6 dB, nominal |
| 10 MHz Distribution Unit | Symmetricom 6502B | RF input: One, 10 MHz RF outputs: Three or more, 10 MHz Nominal output: 1 Vrms |
| *With adapter SWR ≤ 1.12, and NI 5600 IF output SWR ≤ 1.2. †Including splitter and adapter uncertainty: 0.11 dB standard uncertainty when device under test (DUT) on other side of the splitter has SWR ≤ 2.5. **Screened for <0.1 dB tracking (Weinschel 1507R specification <0.2 dB). |
||
For NI 5611:
| Instrument | Recommended Model | Requirements |
|---|---|---|
| Spectrum Analyzer | Rohde & Schwarz FSU Spectrum | Frequency Range: 50 MHz to 6.6 GHz |
| Analyzer with high-frequency preamplifier option (B23) | Noise Floor: <–152 dBm/Hz to 6.6 GHz | |
| Power Meter | Anritsu ML2438A | Range: –50 dBm to +10 dBm |
| with a MA2472 diode sensor | Frequency range: 50 MHz to 6.6 GHz | |
| Accuracy: 0.5% | ||
| NI 5450 | NI 5450 | This device must be fully calibrated before calibrating the NI 5611 |
| NI 5652 | NI 5652 | This device must be fully calibrated before calibrating the NI 5611 |
| Power Splitter | Aeroflex/Weinschel 1593 | SWR: 1.25 |
| Amplitude tracking: <0.25 dB | ||
| 6 dB Attenuator | Anritsu 41KB-6 | Frequency Range: DC to 12 GHz |
| SWR: 1.1 | ||
| 50 Ω Terminator | NI SMA 50 Ω Termination Plug | - |
| Type N(f) to SMA(m) Adapter | S.M. Electronics SM4241 | VSWR: 1.15: 1 |
| SMA Adapter Plug/Plug | Huber+Suhner 32_N-SMA-50-1/11-_NE | VSWR: 1.05: 1 |
| SMA Torque Wrench | NI RF Torque Wrench | 1 N ∙ m |
| Connection Accessory | NI PXIe-5673 Cable Accessory Kit | (4) Matched-Length I/Q Seni-Rigid SMA Cable |
| (1) LO Semi-Rigid SMA Cable | ||
| (1) RF Flexible SMA Cable | ||
| BNC(m) to BNC(m) Cable | - | Use a cable that is 36 inches in length. |
| Chassis | NI PXIe-1065, | - |
| NI PXIe-1075 | ||
| Chassis Controller | NI MXI-Express Kit | - |
For NI 5620:
| Instrument | Required Model | Specifications |
|---|---|---|
| RF Signal Generator | Agilent ESG-4422B (with option UNB, high-output power, lockable to external frequency reference), Agilent 8340A, Agilent ESG-4433B (with option UNB, high-output power, lockable to external frequency reference) | Maximum power output: +10 dBm Frequency range: 100 kHz to 32 MHz Timebase accuracy: ≤±2 ppm under calibration conditions Level accuracy: ≤±1 dB |
| Power Meter with Sensor | Anritsu ML2438A with Anritsu MA2421A Thermal Sensor, Agilent E4419B with Agilent E9304A Power Sensor | Frequency range: 100 kHz to 32 MHz Standard uncertainty (including mismatch)*: ≤±0.1 dB at 0 dBm, source voltage standing wave ratio (VSWR) ≤1.4:1 |
| Broadband Precision Power Splitter | Weinschel 1507R | Frequency range: DC to 32 MHz Output maximum VSWR: 1.15 Amplitude tracking: 0.1 dB† |
*With adapter SWR ≤ 1.12, and NI 5620 input SWR ≤ 1.2; source VSWR ≤1.4:1
† Screened for <0.1 dB tracking (Weinschel 1507R specification <0.2 dB)
For NI 5650, NI 5651 and NI 5652:
| Instrument | Required Model | Specifications |
|---|---|---|
| Spectrum Analyzer | Rohde & Schwarz (R&S) FSU Spectrum Analyzer | Frequency range: 500 kHz to 9 GHz Noise floor: <–150 dBm/Hz |
| Power Meter | Anritsu ML2438A with a MA2472 diode sensor | Range: –70 dBm to +20 dBm Frequency range: 10 MHz to 6.6 GHz Accuracy: 0.5% |
| Frequency Reference | Datum 8040 Rubidium Frequency Standard | Frequency: 10 MHz Frequency Accuracy: 1 ppb (typically ±5E-11) |
For NI 5663:
| Instrument | Required Model | Specifications |
|---|---|---|
| Power Meter | Anritsu ML2438A | Display Resolution: <0.01 dB Settling: <0.1% Instrument Accuracy: <0.5% Noise: <0.5% full-scale (lowest range) Zero Set and Drift:<0.5% full-scale (lowest range) Reference Power Uncertainty: <0.9% Reference Output VSWR: <1.04:1 |
| (2) Power Sensors | Anritsu MA2472D | Input VSWR 10 MHz to 50 MHz: <1.9:1 Input VSWR 50 MHz to 150 MHz: <1.17:1 Input VSWR 150 MHz to 2 GHz: <1.12:1 Input VSWR 2 GHz to 12.4 GHz: <1.22:1 Linearity: –70 dBm to 20 dBm<1.8% Rise Time: <4 µs Calibration Factor Uncertainty 50 MHz: <1.48% Calibration Factor Uncertainty 100 MHz: <1.37% Calibration Factor Uncertainty 300 MHz: <1.42% Calibration Factor Uncertainty 500 MHz: <0.96% Calibration Factor Uncertainty 1 GHz: <0.99% Calibration Factor Uncertainty 2 GHz: <1.04% Calibration Factor Uncertainty 3 GHz: <1.05% Calibration Factor Uncertainty 4 GHz: <1.64% Calibration Factor Uncertainty 5 GHz: <1.39% Calibration Factor Uncertainty 6 GHz: <1.45% Calibration Factor Uncertainty 7 GHz: <1.26% |
High Speed Digital I/O
For NI 6551 and NI 6552:
- Standards
| Instrument | Required Model | Specifications |
|---|---|---|
| Precision voltmeter or digital multimeter (DMM) | NI PXI-4070 or HP 34401 | DC ±1 mV or better at –2 V to 5.5 V |
| Precision voltage source | Fluke 5700A | DC ±1 mV or better at –2 V to 5.5 V |
| Digital oscilloscope | Tektronix TDS 3052, TDS 3052A, TDS 3054 or TDS 3054A | - |
- Additional requirements
- The following equipment is also required for calibration.
| Quantity | Description |
|---|---|
| 1 | PC with Calibration Executive 3.0 or greater installed |
| 1 | PXI Chassis1 and SCXI chassis or PXI-1010 combo chassis |
| 1 | NI 6030/6031/6032/6033 DAQ device |
| 2 | NI SCXI-1193 switch modules |
| 1 | NI SCXI-11xx module to control SCXI chassis2 |
| 1 | NI SCXI-1349 cable assembly2 |
| 1 | NI SMB-2163 accessory (with RESERVED0 access labeled) |
| 26 | MCX-SMB cables (1 m) |
| 4 | MCX-MCX cables (0.3 m) |
| 1 | BNC-SMB cables (0.3 m) |
| 1 | BNC-MCX cable (1 m) |
| 1 | NI SHC68-C68-D2 cable |
| 1 | MCX-SMB cable (2.23 m) |
| 1 | NI MXI-4 kit1 |
| 1 | Male BNC to male dual banana adapter |
1 Not required if you are calibrating PCI devices installed in a PC.
2 Not required if you are using a PXI-1010 combo chassis.
C Series
For NI 9201, NI 9203, NI 9205, NI 9206, NI 9211, NI 9213, NI 9215, NI 9217, NI 9221, NI 9225, NI 9229, NI 9234, NI 9235, NI 9236, NI 9237, NI 9239, NI 9263, NI 9264, NI 9265:
- cDAQ Chassis - NI cDAQ-9178
- Standards for NI 9201, NI 9203, NI 9205, NI 9206, NI 9211, NI 9215, NI 9221, NI 9229, NI 9234, NI 9239
| Instrument | Required Model | Specifications |
|---|---|---|
| Calibrator | Fluke 5700A | 10 ppm |
- Standards for NI 9213
| Instrument | Required Model | Specifications |
|---|---|---|
| Calibrator | Fluke 5520A | 100 ppm |
- Standards for NI 9217, NI 9263, NI 9264, NI 9265
| Instrument | Required Model | Specifications |
|---|---|---|
| DMM | NI 407x | 6 1/2 digits with 40 ppm |
- Standards for NI 9225
| Instrument | Required Model | Specifications |
|---|---|---|
| Calibrator | Fluke 5500A | 10 ppm |
- Standards for NI 9235, NI 9236, NI 9237
| Instrument | Required Model | Specifications |
|---|---|---|
| DMM | NI 4070 | 6 1/2 digits with 40 ppm |
| Calibrator | Fluke 5500A | 10 ppm |
PXI Timing and Synchronization
For NI PXI-6652, NI PXI-6653 and NI PXIe-6672:
- BNC-SMB Cable
- SMB-SMB Cable
- Standards
| Instrument | Required Model | Specifications |
|---|---|---|
| 10 MHz Clock Source | Symmetricom/Datum 8040 | a clock source that is accurate to within 75 ppb for NI PXI-6653, and 150 ppb for NI PXI-6652 and NI PXIe-6672. |
R Series
For NI 7830R, NI 7831R, NI 7833R, NI 7841R, NI 7842R, NI 7851R and NI 7852R:
- Low thermal copper EMF plug-in cables
- Shielded MIO cable - NI SHC68-68-RMIO
- DAQ Accessory - NI SCB-68, NI CB-68LP, NI TBX-68, NI CB-68LPR
- Standards for NI 7830R, NI 7831R, NI 7833R, NI 7841R, NI 7842R, NI 7851R, NI 7852R
| Instrument | Required Model | Specifications |
|---|---|---|
| Calibrator | Fluke 5700A | 10 ppm |
| DMM | NI 4070 | 6 1/2 digits with 40 ppm |
| Counter | HP 53131A | accurate to 0.01% |
If you have any questions or comments, please e-mail us at support@ni.com.
Reader Comments | Submit a comment »
Legal
This tutorial (this "tutorial") was developed by National Instruments ("NI"). Although technical support of this tutorial may be made available by National Instruments, the content in this tutorial may not be completely tested and verified, and NI does not guarantee its quality in any way or that NI will continue to support this content with each new revision of related products and drivers. THIS TUTORIAL IS PROVIDED "AS IS" WITHOUT WARRANTY OF ANY KIND AND SUBJECT TO CERTAIN RESTRICTIONS AS MORE SPECIFICALLY SET FORTH IN NI.COM'S TERMS OF USE (http://ni.com/legal/termsofuse/unitedstates/us/).