PXI-2598 Specifications

Version:
    Last Modified: May 1, 2017

    This document lists specifications for the PXI-2598 (PXI-2598) relay module. All specifications are subject to change without notice. Visit ni.com/manuals for the most current specifications.

    About These Specifications

    Specifications characterize the warranted performance of the instrument under the stated operating conditions. Data in this document are Specifications unless otherwise noted.

    Typical Specifications are specifications met by the majority of the instrument under the stated operating conditions and are tested at 23 °C ambient temperature. Typical specifications are not warranted.

    All voltages are specified in DC, ACpk, or a combination unless otherwise specified.

    Topology

    Dual transfer

    Refer to the NI Switches Help at ni.com/manuals for detailed topology information.

    spd-note-caution
    Caution  

    The protection provided by the PXI-2598 can be impaired if it is used in a manner not described in this document.

    Input Characteristics

    spd-note-caution
    Caution  

    Use precaution when using the PXI-2598 to avoid electrical shock.

    Maximum voltage (cold-switching only)

    65 Vrms, CAT I

    spd-note-caution
    Caution  

    This module is rated for Measurement Category I and intended to carry signal voltages no greater than 100 V. This module can withstand up to 500 V impulse voltage. Do not use this module for connection to signals or for measurements within Categories II, III, or IV. Do not connect to MAINs supply circuits (for example, wall outlets) of 115 or 230 VAC.[1]

    spd-note-caution
    Caution  

    When hazardous voltages (>42.4 Vpk/60 VDC) are present on any channel, safety low-voltage (≤42.4 Vpk/60 VDC) cannot be connected to any other channel.

    spd-note-note
    Note  

    NI recommends against switching active RF signals. As a relay actuates, the channel is momentarily unterminated. Some RF sources can be damaged by reflections if their outputs are not properly terminated. Refer to your RF source documentation for more information.

    spd-note-note
    Note  

    National Instruments recommends against switching RF signals below -35 dBm with this device.

    Maximum carry current (per channel)

    1.25 Arms

    Maximum RF carry power (50 Ω load)

    75 W

    Minimum switch load

    -35 dBm

    RF Performance Characteristics

    Characteristic impedance (Z0)

    50 Ω, nominal

    Insertion loss

    ≤3 GHz

    <0.2 dB

    ≤8 GHz

    <0.3 dB

    ≤12.4 GHz

    <0.4 dB

    ≤18 GHz

    <0.5 dB

    ≤26.5 GHz

    <0.7 dB

    Voltage standing wave ratio (VSWR)

    ≤3 GHz

    <1.2

    ≤8 GHz

    <1.3

    ≤12.4 GHz

    <1.4

    ≤18 GHz

    <1.5

    ≤26.5 GHz

    <1.7

    Open channel isolation

    ≤3 GHz

    >80 dB

    ≤8 GHz

    >70 dB

    ≤12.4 GHz

    >65 dB

    ≤18 GHz

    >60 dB

    ≤26.5 GHz

    >50 dB

    RF carry power

    ≤3 GHz

    75 W

    ≤8 GHz

    50 W

    ≤12.4 GHz

    35 W

    ≤18 GHz

    30 W

    ≤26.5 GHz

    25 W

    Refer to the following figure for the typical insertion loss and isolation of the PXI-2598.

    Figure 1. Insertion Loss and Isolation, Typical

    Refer to the following figure for the typical VSWR of the PXI-2598.

    Figure 2. VSWR, Typical

    Dynamic Characteristics

    Relay operate/release time

    15 ms

    spd-note-note
    Note  

    Certain applications may require additional time for proper settling. Refer to the NI Switches Help at ni.com/manuals for more information about including additional settling time.

    Recommended cycle speed

    5 channels/s

    Expected mechanical relay life

    2.5 × 106 cycles

    Insertion loss repeatability

    <0.03 dB, typical

    Trigger Characteristics

    Input trigger

    Sources

    PXI trigger lines <0...7>

    Minimum pulse width

    150 ns

    spd-note-note
    Note  

    The PXI-2598 can recognize trigger pulse widths less than 150 ns if you disable digital filtering. Refer to the NI Switches Help at ni.com/manuals for information about disabling digital filtering.

    Output trigger

    Destinations

    PXI trigger lines <0...7>

    Pulse width

    Programmable (1 µs to 62 µs)

    Physical Characteristics

    Relay manufacturer/PN

    Radiall/R577 series

    Relay type

    Electromechanical, latching

    Contact material

    Beryllium copper, gold-plated

    I/O connector

    8 SMA jacks

    SMA torque

    0.8 N · m to 1.1 N · m (7 in. · lbs to 10 in. · lbs)

    PXI power requirement

    2.5 W at 3.3 V

    1 W at 5 V

    4 W at 12 V

    Dimensions (L × W × H)

    3U, two slot, PXI/cPCI module, 21.6 cm × 4.1 cm × 13.0 cm (8.5 in. × 1.6 in. × 5.1 in.)

    Weight

    302 g (10.6 oz)

    spd-note-caution
    Caution  

    Clean the hardware with a soft, nonmetallic brush. Make sure that the hardware is completely dry and free from contaminants before returning it to service.

    Environment

    Maximum altitude

    2,000 m (at 25 °C ambient temperature)

    Pollution Degree

    2

    Indoor use only.

    Operating Environment

    Ambient temperature range

    0 °C to 55 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.)

    Relative humidity range

    10% to 90%, noncondensing (Tested in accordance with IEC 60068-2-56.)

    Storage Environment

    Ambient temperature range

    -40 °C to 70 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.)

    Relative humidity range

    5% to 95%, noncondensing (Tested in accordance with IEC 60068-2-56.)

    Shock and Vibration

    Operational shock

    30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.)

    Random vibration

    Operating

    5 Hz to 500 Hz, 0.31 grms (Tested in accordance with IEC 60068-2-64.)

    Nonoperating

    5 Hz to 500 Hz, 2.46 grms (Tested in accordance with IEC 60068-2-64. Test profile exceeds the requirements of MIL-PRF-28800F, Class 3.)

    Compliance and Certifications

    Safety

    This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:

    • IEC 61010-1, EN 61010-1
    • UL 61010-1, CSA 61010-1
    spd-note-note
    Note  

    For UL and other safety certifications, refer to the product label or the Online Product Certification section.

    Electromagnetic Compatibility

    This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
    • EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
    • EN 55011 (CISPR 11): Group 1, Class A emissions
    • AS/NZS CISPR 11: Group 1, Class A emissions
    • FCC 47 CFR Part 15B: Class A emissions
    • ICES-001: Class A emissions
    spd-note-note
    Note  

    In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended for use only in heavy-industrial locations.

    spd-note-note
    Note  

    Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes.

    spd-note-note
    Note  

    For EMC declarations and certifications, refer to the Online Product Certification section.

    CE Compliance

    This product meets the essential requirements of applicable European Directives, as follows:

    • 2014/35/EU; Low-Voltage Directive (safety)
    • 2014/30/EU; Electromagnetic Compatibility Directive (EMC)

    Online Product Certification

    Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.

    Environmental Management

    NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.

    For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.

    Waste Electrical and Electronic Equipment (WEEE)
    spd-note-weee
    EU Customers  

    At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee.

    电子信息产品污染控制管理办法(中国RoHS)
    spd-note-china-rohs
    中国客户  

    National Instruments符合中国电子信息产品中限制使用某些有害物质指令(RoHS)。关于National Instruments中国RoHS合规性信息,请登录 ni.com/environment/rohs_china。(For information about China RoHS compliance, go to ni.com/environment/rohs_china.)

    • 1 Measurement Categories CAT I and CAT O are equivalent. These test and measurement circuits are not intended for direct connection to the MAINS building installations of Measurement Categories CAT II, CAT III, or CAT IV.

    Recently Viewed Topics