Table Of Contents

PXIe-4162 Specifications

Version:
    Last Modified: December 13, 2018

    Definitions

    Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

    Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

    • Typical specifications describe the performance met by a majority of models.
    • Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.

    Specifications are Nominal unless otherwise noted.

    Conditions

    Specifications are valid under the following conditions unless otherwise noted.

    • Ambient temperature[1] of 23 °C ± 5 ºC
    • Chassis with slot cooling capacity ≥38 W[2]
      • For chassis with slot cooling capacity = 38 W, fan speed set to HIGH
    • Calibration interval of 1 year
    • 30 minutes warm-up time
    • Self-calibration performed within the last 24 hours
    • niDCPower Aperture Time property or NIDCPOWER_ATTR_APERTURE_TIME attribute set to 2 power-line cycles (PLC)

    Instrument Capabilities

    Channels

    0 through 11

    DC voltage range

    ±24 V

    The following table and figure illustrate the voltage and the current source and sink ranges of the PXIe-4162.

    Table 1. PXIe-4162 DC Current Source and Sink Ranges, Warranted
    Chassis Slot Cooling Capacity
    ≥58 W 38 W
    10 μA
    100 μA
    1 mA
    10 mA
    100 mA 60 mA
    Figure 1. PXIe-4162 Quadrant Diagram, Any Channel

    SMU Specifications

    Voltage

    Table 2. Voltage Programming and Measurement Accuracy/Resolution, Warranted
    Range Resolution and Noise (0.1 Hz to 10 Hz) Accuracy (23 °C ± 5 °C) ± (% of Voltage + Offset) Tempco[3] ± (% of Voltage + Offset)/°C, 0 °C to 55 °C
    Tcal ± 5 °C
    24 V 200 μV 0.05% + 5 mV 0.0005% + 1 µV

    Current

    Table 3. Current Programming and Measurement Accuracy/Resolution, Warranted
    Range Resolution and Noise (0.1 Hz to 10 Hz) Accuracy (23 °C ± 5 °C) ± (% of Current + Offset) Tempco[4] ± (% of Current + Offset)/°C, 0 °C to 55 °C
    Tcal ± 5 °C
    10 μA 100 pA 0.10% + 5 nA 0.004% + 10 pA
    100 μA 1 nA 0.10% + 50 nA 0.004% + 100 pA
    1 mA 10 nA 0.10% + 500 nA 0.004% + 1 nA
    10 mA 100 nA 0.10% + 5 μA 0.004% + 10 nA
    60 mA or 100 mA[5] 1 µA 0.10% + 50 μA 0.004% + 100 nA

    Available DC Output Power

    Chassis Slot Cooling Capacity Per Channel Maximum Absolute Maximum
    ≥58 W 2.4 W 28.8 W
    38 W 1.4 W 11.5 W

    Additional Specifications

    Settling time[6]

    <500 μs, typical[7]

    Transient response[8]

    <100 μs, typical[9]

    Wideband source noise[10]

    15 mV RMS, typical

    <100 mVpk-pk, typical

    Cable guard output current limit

    100 µA, typical

    Remote sense

    Voltage

    No additional error due to lead drop

    Current

    No additional error due to lead drop

    Maximum lead drop

    1 V drop/lead

    Load regulation

    Voltage[11]

    50 μV/mA, typical

    Current

    (30 pA + 20 ppm of range)/volt, typical

    Functional isolation voltage, any pin to earth ground

    60 V DC

    Absolute maximum voltage to Output LO
    From Sense HI, Sense LO, or Guard[12]

    When VOutput HI > 0 V

    -0.5 V to (VOutput HI + 0.5 V)

    When VOutput HI0 V

    (VOutput HI - 0.5 V) to 0.5 V

    From all other pins

    ±25 V

    The following figures illustrate noise as a function of measurement aperture for the PXIe-4162.

    Figure 2. Voltage RMS Noise Versus Aperture Time, [13] Typical
    Figure 3. Current RMS Noise Versus Aperture Time, [14] Typical

    Supplemental Specifications

    Measurement and Update Timing

    Available sample rates[15]

    (600 kS/s)/N

    where

    • N = 6, 7, 8, … 220
    • S is samples

    Sample rate accuracy

    ±50 ppm

    Maximum measure rate to host[16]

    100,000 S/s per channel, continuous

    Maximum source update rate[17]

    Single channel

    100,000 updates/s

    All channels simultaneously

    40,000 updates/s per channel

    Input trigger to

    Source event delay

    8.5 µs

    Source event jitter

    1.7 μs

    Measure event jitter

    1.7 μs

    Triggers

    Input triggers

    Types

    Start

    Source

    Sequence Advance

    Measure

    Sources (PXI trigger lines 0 to 7)[18]

    Polarity

    Active high (not configurable)

    Minimum pulse width

    100 ns

    Destinations[19] (PXI trigger lines 0 to 7)[18]

    Polarity

    Active high (not configurable)

    Minimum pulse width

    >200 ns

    Output triggers (events)

    Types

    Source Complete

    Sequence Iteration Complete

    Sequence Engine Done

    Measure Complete

    Destinations (PXI trigger lines 0 to 7)[18]

    Polarity

    Active high (not configurable)

    Pulse width

    230 ns

    Calibration Interval

    Recommended calibration interval

    1 year

    Physical

    Dimensions

    3U, one-slot, PXI Express/CompactPCI Express module

    2.0 cm × 13.0 cm × 21.6 cm (0.8 in. × 5.1 in. × 8.5 in.)

    Weight

    394 g (13.9 oz)

    Front panel connectors

    Custom 62-position D-SUB, female

    Power Requirements

    Chassis Slot Cooling Capacity +3.3 V Current Draw, Typical +12 V Current Draw, Typical
    Idle Full Output Load Idle Full Output Load
    38 W 1 A 1 A 1.5 A 3 A
    ≥58 W 1 A 4.5 A

    Environmental Characteristics

    Temperature and Humidity

    Temperature
    Operating

    Chassis with slot cooling capacity ≥58 W[20]

    0 °C to 55 °C

    All other compatible chassis

    0 °C to 40 °C

    Storage

    -40 °C to 71 °C

    Humidity

    Operating

    10% to 90%, noncondensing

    Storage

    5% to 95%, noncondensing

    Pollution Degree

    2

    Maximum altitude

    2,000 m (800 mbar) (at 25 °C ambient temperature)

    Shock and Vibration

    Random vibration

    Operating

    5 Hz to 500 Hz, 0.3 g RMS

    Non-operating

    5 Hz to 500 Hz, 2.4 g RMS

    Operating shock

    30 g, half-sine, 11 ms pulse

    • 1 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake).
    • 2 For increased capability, NI recommends installing the PXIe-4162 in a chassis with slot cooling capacity ≥58 W.
    • 3 Temperature coefficient applies beyond 23 °C ± 5 °C within 5 °C of Tcal.
    • 4 Temperature coefficient applies beyond 23 °C ± 5 °C within 5 °C of Tcal.
    • 5 100 mA range available only when installed in chassis with slot cooling capacity ≥58 W. 60 mA range available in all other compatible chassis.
    • 6 Current limit set to ≥1 mA and ≥10% of the selected current limit range. PXIe-4162 configured for fast transient response.
    • 7 To settle to 0.1% of voltage step.
    • 8 PXIe-4162 configured for fast transient response.
    • 9 To recover within ±20 mV after a load current change from 10% to 90% of range.
    • 10 20 Hz to 20 MHz bandwidth. PXIe-4162 configured for normal transient response. Measured at the end of the 1 m SHDB62M-DB62M-LL cable.
    • 11 At connector pins when using local sense.
    • 12 Where VOutput HI is the voltage at the Output HI pin in the same channel as a Sense HI, Sense LO, or Guard pin.
    • 13 All channels averaged. Channel 11 has degraded performance.
    • 14 All channels averaged. In the 100 mA range, channel 4 has degraded performance.
    • 15 When source-measuring, both the NI-DCPower Source Delay and Aperture Time properties affect the sampling rate. When taking a measure record, only the Aperture Time property affects the sampling rate.
    • 16 Load dependent settling time is not included. Normal DC noise rejection is used.
    • 17 As the source delay is adjusted or if advanced sequencing is used, maximum source update rates may vary.
    • 18 Pulse widths and logic levels are compliant with PXI Express Hardware Specification Revision 1.0 ECN 1.
    • 19 Input triggers can come from any source (PXI trigger or software trigger) and be exported to any PXI trigger line. This allows for easier multi-board synchronization regardless of the trigger source.
    • 20 Not all chassis with slot cooling capacity ≥58 W can achieve this ambient temperature range. Refer to PXI chassis specifications to determine the ambient temperature ranges your chassis can achieve.