PXI-4132 Specifications

Version:
    Last Modified: January 23, 2018

    These specifications apply to the PXI-4132.

    Definitions

    Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

    Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

    • Typical specifications describe the expected performance met by a majority of the models.
    • Nominal specifications describe parameters and attributes that may be useful in operation.

    Specifications are Warranted unless otherwise noted.

    Conditions

    Specifications are valid under the following conditions unless otherwise noted.

    • Ambient temperature[1] of 23 °C ± 5 ºC
    • 30 minutes warm-up time
    • niDCPower Auto Zero property or NIDCPOWER_ATTR_AUTO_ZERO attribute set to On
    • niDCPower Aperture Time property or NIDCPOWER_ATTR_APERTURE_TIME attribute set to 1 power-line cycle (PLC)
    • Self-calibration performed within the last 24 hours

    Device Capabilities

    The following table and figure illustrate the voltage and current source and sink ranges of the PXI-4132.
    Table 1. PXI-4132 Current Source and Sink Ranges
    DC Voltage Ranges (CAT I) DC Current Source and Sink Ranges
    • ±10 V
    • ±100 V
    • 10 μA
    • 100 μA
    • 1 mA
    • 10 mA
    • 100 mA
    spd-note-caution
    Caution  

    Shock hazards exist when voltage levels are greater than 30 VRMS, 42.4 V peak, or 60 VDC. Use extreme caution when a shock hazard is present. Always ensure the test system is de-energized before connecting or disconnecting the backshell assembly or cables from the PXI-4132.

    Figure 1. PXI-4132 Quadrant Diagram, Characteristic

    SMU Specifications

    Voltage Programming Accuracy/Resolution

    Range Resolution, Nominal Accuracy ± (% of Output + Offset) Peak to Peak Noise, Typical (0.1 Hz to 10 Hz)
    1 Year 23 °C ± 5 °C
    ±10 V 50 μV 0.025% + 3 mV 70 μV
    ±100 V 500 µV 0.025% + 10 mV 300 μV

    Current Programming Accuracy/Resolution[2]

    Range Resolution, Nominal Accuracy ± (% of Output + Offset) Peak to Peak Noise, Typical (0.1 Hz to 10 Hz)
    1 Year 23 °C ± 5 °C
    10 µA 500 pA 0.034% + 2.0 nA 90 pA
    100 µA 5 nA 0.034% + 20 nA 900 pA
    1 mA 50 nA 0.034% + 0.2 µA 9 nA
    10 mA 500 nA 0.034% + 2.0 µA 90 nA
    100 mA 5 µA 0.034% + 20 µA 900 nA

    Voltage Measurement Accuracy/Resolution

    Range Resolution, Nominal Accuracy ± (% of Reading + Offset)
    1 Year 23 °C ± 5 °C
    ±10 V 10 μV 0.02% + 2.0 mV
    ±100 V 100 µV 0.02% + 5.0 mV

    Current Measurement Accuracy/Resolution

    Range Resolution, Nominal Accuracy ± (% of Reading + Offset)
    1 Year 23 °C ± 5 °C
    10 µA 10 pA 0.028% + 1.0 nA
    100 µA 100 pA 0.028% + 10 nA
    1 mA 1 nA 0.028% + 0.1 µA
    10 mA 10 nA 0.028% + 1.0 µA
    100 mA 100 nA 0.020% + 10 µA

    Additional Specifications

    Temperature Coefficient

    15% of accuracy specification per °C

    Settling time

    <300 µs; Settled to 0.1% of final value (1 V step at 50% load of current range), typical

    Transient response

    Recovers to <0.1% of voltage range within 100 µs after a change in load current from 10% to 90% of current range, typical

    Normal mode noise (source only)

    8 mVp-p into resistive load <1 mV RMS (20 Hz to 20 MHz bandwidth), typical

    Remote sense

    Add 0.5% of HI lead drop to voltage accuracy specification

    (Maximum lead drop) Up to 1 V drop per lead

    Load Regulation

    Voltage

    0.5 mV per mA of output load using local sense

    Current

    0.01% of range per volt of output change

    Guard offset voltage

    <4 mV, typical (Current ≤ 10 mA)

    Isolation voltage (continuous)

    Channel-to-earth ground

    150 VDC, CAT I[3], verified by dielectric withstand test, 5 s, characteristic

    spd-note-caution
    Caution  

    Do not connect to MAINs. Do not connect to signals or use for measurements within CAT II, III, or IV.

    Step Response

    The following figures illustrate the step response of the PXI-4132 for different loads.
    Figure 2. 100 mA Range Step Response, Typical
    Figure 3. 10 mA Range Step Response, Typical
    Figure 4. 1 mA, 100 µA, and 10 µA Range Step Response, Typical

    Measurement Noise Versus Aperture Time, Typical

    Supplemental Specifications

    Measurement Speed[4]

    Table 2. Maximum Operation Rates Per Second for 60 Hz ( 50 Hz)
    ADC Aperture Time Measure to Host Source-measure to Host[5]
    1/64 PLC 3490 (2900) 1900 (1700)
    1/8 PLC 470 (390) 425 (360)
    1 PLC 59.9 (49.9) 59.0 (49.3)

    Maximum source update rate[5]

    4,200 updates/s

    Trigger in to source delay

    500 ns, nominal

    Triggers

    Input triggers

    Types

    Start, Source, Sequence Advance, Measure

    Sources

    PXI Trigger lines <0...7>[6]

    Polarity, sources

    Configurable

    Minimum pulse width, sources

    100 ns

    Destinations[7]

    PXI Trigger Lines <0...7>[6]

    Polarity, destinations

    Active high (unconfigurable)

    Pulse width, destinations

    150 ns, nominal

    Output Triggers (Events)

    Types

    Source Complete, Sequence Iteration Complete, Sequence Engine Done, Measure Complete

    Destinations

    PXI Trigger Lines <0...7>[6]

    Polarity, destinations

    Configurable

    Pulse width, destinations

    Configurable between 150 ns and 1.6 µs

    Calibration Interval

    Recommended calibration interval

    1 year

    Physical Characteristics

    Dimensions

    3U, one-slot, PXI/cPXI module

    2.0 cm × 13.0 cm × 21.6 cm

    (0.8 in. × 5.1 in. × 8.5 in.), nominal

    Weight

    295 g

    (10.4 oz), typical

    Front panel connectors

    COMBICON, 5.08 mm (8 position), nominal

    spd-note-note
    Note  

    Front panel connectors can accept wire gauges from 12 AWG to 28 AWG.

    Power Requirements

    PXI power requirement

    10 W at 5 V

    1 W at 3.3 V

    2 W at 12 V, typical

    Environment

    Maximum altitude

    2,000 m (at 25 °C ambient temperature)

    Pollution Degree

    2

    Indoor use only.

    Operating Environment

    Ambient temperature range

    0 °C to 55 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.)

    Relative humidity range

    10% to 70%, noncondensing; derate 1.3% per °C above 40 °C (Tested in accordance with IEC 60068-2-56.)

    Storage Environment

    Ambient temperature range

    -40 °C to 70 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.)

    Relative humidity range

    5% to 95%, noncondensing (Tested in accordance with IEC 60068-2-56.)

    Shock and Vibration

    Operational shock

    30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.)

    Random vibration

    Operating

    5 Hz to 500 Hz, 0.3 grms (Tested in accordance with IEC 60068-2-64.)

    Nonoperating

    5 Hz to 500 Hz, 2.4 grms (Tested in accordance with IEC 60068-2-64. Test profile exceeds the requirements of MIL-PRF-28800F, Class 3.)

    Compliance and Certifications

    spd-note-caution
    Caution  

    You can impair the protection provided by the PXI-4132 if you use it in a manner not described in this document.

    Safety

    This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:

    • IEC 61010-1, EN 61010-1
    • UL 61010-1, CSA C22.2 No. 61010-1
    spd-note-note
    Note  

    For UL and other safety certifications, refer to the product label or the Online Product Certification section.

    Electromagnetic Compatibility

    This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
    • EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
    • EN 55011 (CISPR 11): Group 1, Class A emissions
    • EN 55022 (CISPR 22): Class A emissions
    • EN 55024 (CISPR 24): Immunity
    • AS/NZS CISPR 11: Group 1, Class A emissions
    • AS/NZS CISPR 22: Class A emissions
    • FCC 47 CFR Part 15B: Class A emissions
    • ICES-001: Class A emissions
    spd-note-note
    Note  

    In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended for use only in heavy-industrial locations.

    spd-note-note
    Note  

    Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes.

    spd-note-note
    Note  

    For EMC declarations, certifications, and additional information, refer to the Online Product Certification section.

    CE Compliance

    This product meets the essential requirements of applicable European Directives, as follows:

    • 2014/35/EU; Low-Voltage Directive (safety)
    • 2014/30/EU; Electromagnetic Compatibility Directive (EMC)

    Online Product Certification

    Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.

    Environmental Management

    NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.

    For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.

    Waste Electrical and Electronic Equipment (WEEE)

    spd-note-weee
    EU Customers  

    At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee.

    电子信息产品污染控制管理办法(中国RoHS)

    spd-note-china-rohs
    中国客户  

    National Instruments符合中国电子信息产品中限制使用某些有害物质指令(RoHS)。关于National Instruments中国RoHS合规性信息,请登录 ni.com/environment/rohs_china。(For information about China RoHS compliance, go to ni.com/environment/rohs_china.)

    • 1 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake).
    • 2 Minimum programmable current limit/level is 2% of range.
    • 3 Measurement Categories CAT I and CAT O (Other) are equivalent. These test and measurement circuits are not intended for direct connection to the MAINs building installations of Measurement Categories CAT II, III, or CAT IV.
    • 4 Does not include load dependent settling time; niDCPower Auto Zero property or NIDCPOWER_ATTR_AUTO_ZERO attribute set to Off.
    • 5 Source-measure to Host and Maximum Source Update Rate are performed with the source delay set to 200 µs. This is sufficient for the output to settle within 1% of the requested level with a simple resistive load. As you adjust the source delay for your application's requirements, maximum rates vary.
    • 6 Pulse widths and logic levels compliant with PXI Hardware Specification Revision 2.2.
    • 7 Input triggers can be re-exported.