Table Of Contents

PXI-4130 Specifications

Version:
    Last Modified: February 17, 2017

    These specifications apply to the PXI-4130 with APS-4100 auxiliary power supply and to the PXI-4130 without auxiliary power supply.

    Definitions

    Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

    Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

    • Typical specifications describe the expected performance met by a majority of the models.
    • Nominal specifications describe parameters and attributes that may be useful in operation.

    Specifications are Warranted unless otherwise noted.

    Conditions

    Specifications are valid under the following conditions unless otherwise noted.

    • Ambient temperature[1] of 23 °C ± 5 ºC
    • 30 minutes warm-up time
    • niDCPower Samples to Average property or NIDCPOWER_ATTR_SAMPLES_TO_AVERAGE attribute set to 300 for optimal 50 Hz and 60 Hz rejection

    Device Capabilities

    Channel DC Voltage Ranges Isolation DC Current Source and Sink Ranges
    SMU Channel (1)
    • -20 V to +20 V
    • -6 V to +6 V
    60 VDC, CAT I
    • 200 μA
    • 2 mA
    • 20 mA
    • 200 mA
    • 2 A[2]
    Utility Channel (0)
    • 0 V to 6 V
    N/A
    • 1 A (6 W)
    Figure 1. Channel 0 Quadrant Diagram
    Figure 2. Channel 1 Quadrant Diagram

    SMU Channel Specifications (Channel 1)

    Voltage Programming Accuracy/Resolution

    Range Resolution Accuracy ± (% of Output + Offset)
    1 Year 23 °C ± 5 °C
    ±20 V 0.33 mV 0.034% + 1.8 mV
    ±6 V 0.1 mV 0.034% + 1.5 mV

    Current Programming Accuracy/Resolution[3]

    Range Resolution Accuracy ± (% of Output + Offset)
    1 Year 23 °C ± 5 °C
    200 μA 10 nA 0.03% + 0.1 μA
    2 mA 100 nA 0.03% + 1 μA
    20 mA 1 μA 0.03% + 10 μA
    200 mA 10 μA 0.03% + 100 μA
    2 A[4] 100 μA 0.12% + 1 mA

    Voltage Measurement Accuracy/Resolution

    Range Resolution Accuracy ± (% of Reading + Offset)
    1 Year 23 °C ± 5 °C
    ±20 V 0.10 mV 0.03% + 1.5 mV
    ±6 V 0.10 mV 0.03% + 1.5 mV

    Current Measurement Accuracy/Resolution

    Range Resolution Accuracy ± (% of Reading + Offset)
    1 Year 23 °C ± 5 °C
    200 μA 1.0 nA 0.03% + 0.02 μA
    2 mA 10 nA 0.03% + 0.2 μA
    20 mA 0.1 μA 0.03% + 2 μA
    200 mA 1 μA 0.03% + 40 μA
    2 A[] 10 μA 0.12% + 200 μA

    Channel 1 Additional Specifications

    Settling time[5]

    500 μs, typical

    Output capacitance

    Low setting

    10 nF, typical

    High setting

    6.8 μF, typical

    Slew rate

    0.08 V/μs, typical

    Transient response

    Recovers to <0.1% of voltage range within 200 μs after a change in load current from 10% to 90% of current range, typical

    Normal Mode noise (Source only)

    15 mVp-p into resistive load <5 mV RMS 20 Hz to 20 MHz bandwidth, typical

    Remote sense

    Up to 1 V drop per lead using internal power or ≥ 12 V auxiliary power supply; Add 120 μV to accuracy specification per volt of lead drop.

    Load regulation

    Voltage

    20 mV per amp of output load using Local Sense

    Current

    0.01% of range per volt of output change

    Line regulation (% of output + offset, per volt of change in auxiliary power input)

    Voltage

    0.01 + 1 mV

    Current

    0.01 + 0.02% of range

    Temperature coefficient (Tempco) is 10% of accuracy specification per °C.

    Utility Channel Specifications (Channel 0)

    Programming Accuracy/Resolution[6]

    Output Function Range Resolution Accuracy ± (% of output + offset)
    1 Year 23 °C ± 10 °C
    Voltage +6 V 0.12 mV 0.05% + 4 mV
    Current 1 A[7] 0.02 mA 0.15% + 4 mA

    Measurement Accuracy/Resolution

    Measurement Type Range Resolution Accuracy ± (% of reading + offset)
    1 Year 23 °C ± 10 °C
    Voltage +6 V 0.06 mV 0.05% + 4 mV
    Current 1 A[] 0.01 mA 0.15% + 4 mA

    Channel 0 Additional Specifications

    Settling time

    <1 ms, 10% to 90% of range, measured with full load, typical

    Output capacitance

    33 µF, typical

    Transient response

    Recovers to <0.1% of voltage range within 50 μs after a change in load current from 50% to 100% of current range, typical

    Normal Mode noise and ripple (source only, voltage)

    <1.5 mV RMS, 20 Hz to 20 MHz bandwidth, typical

    Load regulation

    Voltage

    0.42% of range per amp of output load

    Current

    0.02% of range per volt of output change

    Temperature coefficient (Tempco) is 15% of accuracy specification per °C.

    Programming and Measurement Timing[8]

    Maximum output update rate

    3000 Updates/s, nominal

    Maximum measurement rate (samples to average = 1)

    3 kS/s, nominal

    Single point update latency

    600 µs, typical

    Protection

    Auxiliary power input protection

    Overvoltage

    >15.5 VDC shut-off; >20 VDC crowbar (fused)

    Overcurrent or reverse voltage

    Fused

    Calibration Interval

    Recommended calibration interval

    1 year

    Accuracy Specification Derating versus Load Current

    Maximum Sinking Power versus Ambient Temperature for Channel 1

    Physical Characteristics

    Dimensions

    3U, one-slot PXI/cPCI module

    2.0 cm × 13.0 cm × 21.6 cm

    (0.8 in. × 5.1 in. × 8.5 in.), nominal

    Weight

    312 g (11 oz), typical

    User-replaceable fuses

    Channel 0 (internally-socketed)

    1, Littelfuse 045301.5

    (F 1.5 A 125 V), characteristic

    Auxiliary power input (front panel mount)

    1, 5 × 20 mm glass fuse

    (T 6.3 A L 250 V), characteristic

    spd-note-note
    Note  

    NI recommends Littelfuse 21806.3 for Auxiliary Power Input fuse.

    Front panel connectors

    Output channels

    MINI-COMBICON, 3.81 mm (6 position), nominal

    Auxiliary power input

    MINI-COMBICON, 3.5 mm (2 position), nominal

    spd-note-note
    Note  

    Front panel connectors can accept wire gauges from 16 AWG to 28 AWG.

    Power Requirements

    PXI power requirement

    10 W at 5 V

    1 W at 3.3 V

    6 W at 12 V

    2.5 W at -12 V, typical

    Auxiliary power source[9] input requirements

    11 VDC to 15.5 VDC

    5 A max

    Environment

    Maximum altitude

    2,000 m (at 25 °C ambient temperature)

    Pollution Degree

    2

    Indoor use only.

    Operating Environment

    Ambient temperature range

    0 °C to 55 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.)

    Relative humidity range

    10% to 90%, noncondensing (Tested in accordance with IEC 60068-2-56.)

    Storage Environment

    Ambient temperature range

    -40 °C to 70 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.)

    Relative humidity range

    5% to 95%, noncondensing (Tested in accordance with IEC 60068-2-56.)

    Shock and Vibration

    Operational shock

    30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.)

    Random vibration

    Operating

    5 Hz to 500 Hz, 0.3 grms (Tested in accordance with IEC 60068-2-64.)

    Nonoperating

    5 Hz to 500 Hz, 2.4 grms (Tested in accordance with IEC 60068-2-64. Test profile exceeds the requirements of MIL-PRF-28800F, Class 3.)

    Compliance and Certifications

    spd-note-caution
    Caution  

    You can impair the protection provided by the PXI-4130 if you use it in a manner not described in this document.

    spd-note-caution
    Caution  

    To ensure the specified EMC performance, operate this product only with shielded cables and accessories.

    Safety

    spd-note-caution
    Caution  

    Do not connect to signals or use for measurements within CAT II, III, or IV.

    Isolation Voltage

    Channel-to-earth ground, continuous

    60 VDC, CAT I[10], verified by dielectric withstand test, 5 s

    This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:

    • IEC 61010-1, EN 61010-1
    • UL 61010-1, CSA C22.2 No. 61010-1
    spd-note-note
    Note  

    For UL and other safety certifications, refer to the product label or the Online Product Certification section.

    Electromagnetic Compatibility

    This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
    • EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
    • EN 55011 (CISPR 11): Group 1, Class A emissions
    • EN 55022 (CISPR 22): Class A emissions
    • EN 55024 (CISPR 24): Immunity
    • AS/NZS CISPR 11: Group 1, Class A emissions
    • AS/NZS CISPR 22: Class A emissions
    • FCC 47 CFR Part 15B: Class A emissions
    • ICES-001: Class A emissions
    spd-note-note
    Note  

    In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended for use only in heavy-industrial locations.

    spd-note-note
    Note  

    Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes.

    spd-note-note
    Note  

    For EMC declarations, certifications, and additional information, refer to the Online Product Certification section.

    CE Compliance

    This product meets the essential requirements of applicable European Directives, as follows:

    • 2014/35/EU; Low-Voltage Directive (safety)
    • 2014/30/EU; Electromagnetic Compatibility Directive (EMC)

    Online Product Certification

    Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.

    Environmental Management

    NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.

    For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.

    Waste Electrical and Electronic Equipment (WEEE)

    spd-note-weee
    EU Customers  

    At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee.

    电子信息产品污染控制管理办法(中国RoHS)

    spd-note-china-rohs
    中国客户  

    National Instruments符合中国电子信息产品中限制使用某些有害物质指令(RoHS)。关于National Instruments中国RoHS合规性信息,请登录 ni.com/environment/rohs_china。(For information about China RoHS compliance, go to ni.com/environment/rohs_china.)

    • 1 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake).
    • 2 Current input/output for channel 1 is limited to 2 W or 300 mA when operating under internal power. Continuous sinking power for channel 1 is limited to 10 W, subject to derating above 30 °C. Refer to the figure Maximum Sinking Power versus Ambient Temperature for Channel 1.
    • 3 Minimum programmable current limit/level is 2% of range.
    • 4 For currents ≥ 500 mA, refer to the additional derating information in the figure Accuracy Specification Derating versus Load Current.
    • 5 Settled to 1%, 1 V step, 50% of current range at final value, output capacitance set to low, using auxiliary power supply.
    • 6 Minimum programmable current limit/level is 2% of range.
    • 7 For currents ≥500 mA, refer to the additional derating information in the figure Accuracy Specification Derating versus Load Current.
    • 8 Does not include load dependent settling time.
    • 9 Optional; Channel 1 only.
    • 10 Measurement Categories CAT I and CAT O (Other) are equivalent. These test and measurement circuits are not intended for direct connection to the MAINs building installations of Measurement Categories CAT II, III, or CAT IV.