PXIe-2542 Specifications

Version:
    Last Modified: March 3, 2017

    This document lists specifications for the PXIe-2542 ( PXIe-2542) multiplexer module. All specifications are subject to change without notice. Visit ni.com/manuals for the most current specifications.

    Topology

    Quad 2 × 1 multiplexers

    About These Specifications

    Specifications characterize the warranted performance of the instrument under the following operating conditions:

    • Chassis is powered on
    • Calibration adjustment cycle maintained
    • 50 Ω termination connected to unused I/O front panel connectors

    Typical Specifications are specifications met by the majority of the instrument under the stated operating conditions and are tested at 23 °C ambient temperature. Typical specifications are not warranted.

    Nominal values (or supplemental information) describe additional information about the product that may be useful, including expected performance that is not covered under Typical Specifications. Nominal values are not covered by warranty.

    All voltages are specified in DC, ACpk, or a combination unless otherwise specified.

    Clean devices and terminal blocks by brushing off light dust with a soft, nonmetallic brush. Remove other contaminants with a soft, lint-free, dampened cloth. Do not use detergent or chemical solvents. The unit must be completely dry and free from contaminants before returning to service.

    Input Characteristics

    Minimum input frequency

    10 MHz

    Characteristic impedance (Z0)

    50 Ω nominal

    Coupling

    AC

    Maximum safe DC input voltage

    ±8V

    Maximum Safe Continuous RF Power

    Chassis power ON

    +30 dBm

    Chassis power OFF

    +20 dBm

    RF Performance Characteristics

    Values in parentheses are typical.

    Insertion loss

    ≤2.4 GHz

    <3.7 dB (<2.5 dB)

    ≤6 GHz

    <4.8 dB (<3.9 dB)

    ≤6.6 GHz

    <6.1 dB (<5.0 dB)

    Insertion loss thermal coefficient

    α = 1,112 ppm/°C

    Use the following equation to calculate the insertion loss at a given temperature:

    I L T = I L T 0 ( 1 + α ( T T 0 ) )
    where IL is insertion loss in dB
      T is the temperature at which the property is being measured in °C
      T0 is the reference temperature in °C
      α represents insertion loss temperature coefficient in ppm/°C
    Table 1. Typical Flatness
    Frequency Flatness / 200 MHz Flatness / 1GHz
    10 MHz to 5 GHz <0.15 dB <0.50 dB
    5 GHz to 6 GHz <0.18 dB <0.62 dB
    6 GHz to 6.6 GHz <0.63 dB <1.25 dB
    Voltage standing wave ratio (VSWR)

    10 MHzx ≤2.4 GHz

    <1.6 (<1.4)

    2.4 GHz < x ≤3.5 GHz

    <1.7 (<1.4)

    3.5 GHz < x ≤6 GHz

    <1.6 (<1.2)

    6 GHz < x ≤6.6 GHz

    <2.2 (<1.8)

    CH-COM isolation

    ≤2.4 GHz

    >78 dB (>94 dB)

    ≤6 GHz

    >60 dB (>86 dB)

    ≤6.6 GHz

    >55 dB (>75 dB)

    CH-CH isolation

    ≤2.4 GHz

    >79 dB (>94 dB)

    ≤6 GHz

    >58 dB (>76 dB)

    ≤6.6 GHz

    >54 dB (>70 dB)

    Bank-to-bank crosstalk

    ≤2.4 GHz

    >75 dB (>94 dB)

    ≤6 GHz

    >60 dB (>85 dB)

    ≤6.6 GHz

    >55 dB (>83 dB)

    Typical channel-to-channel skew

    <3 ps

    Typical propagation delay

    680 ps

    Input 1 dB compression

    Minimum

    >28 dBm

    Typical

    >32 dBm

    Refer to the following figures for typical insertion loss, typical VSWR, and typical isolation, and typical bank-to-bank isolation respectively.

    Figure 1. Typical Insertion Loss
    Figure 2. Typical VSWR
    Figure 3. Typical Isolation
    Figure 4. Typical Bank-to-Bank Isolation

    Linearity

    Second-order harmonic distortion (Input IP2 (IIP2))

    Typical IP2 (input)

    >+91 dBm

    spd-note-note
    Note  

    Measurements are performed with single CW tones, ranging from -26 dBm to +10 dBm at the RF input. This specification is based on both experimental and calculated data.

    Third-order intermodulation distortion (Input IP3 (IIP3))

    Typical IP3 (input)

    >+61 dBm

    spd-note-note
    Note  

    Measurements performed with two 10 dBm input tones = 1 MHz apart. This specification is based on both experimental and calculated data.

    Dynamic Characteristics

    Relay Operate Time

    Typical

    1 ms

    Maximum

    3.3 ms

    Simultaneous drive limit

    90 relays

    Expected mechanical relay life

    1 × 108 cycles

    Expected electrical relay life

    ≤30 mV, ≤10 mA resistive

    2.5 × 106 cycles

    30 V, 1 A

    5 × 105 cycles

    30 V, 2 A

    1 × 105 cycles

    60 VDC, 1 ADC resistive

    1 × 105 cycles

    spd-note-note
    Note  

    Relays are field replaceable. Refer to the NI Switches Help at ni.com/manuals for more information about replacing a failed relay.

    spd-note-note
    Note  

    Certain applications may require additional time for proper settling. Refer to the NI Switches Help at ni.com/manuals for more information about including additional settling time.

    Trigger Characteristics

    Input trigger

    Sources

    PXI trigger lines <0...7>

    Minimum pulse width

    150 ns

    spd-note-note
    Note  

    The PXIe-2542 can recognize trigger pulse widths less than 150 ns if you disable digital filtering. Refer to the NI Switches Help at ni.com/manuals for information about disabling digital filtering.

    Output trigger

    Destinations

    PXI trigger lines <0...7>

    Pulse width

    Programmable (1 µs to 62 µs)

    Physical Characteristics

    Switch type

    FET

    Front panel I/O connectors

    12 SMA jacks, female

    Power requirement
    PXI

    5 V

    0.1 W

    3.3 V

    0.7 W

    PXI Express

    12 V

    0.6 W

    3.3 V

    0.9 W

    Dimensions (L × W × H)

    3U, one slot, PXI/cPCI module, PXI Express compatible 21.6 cm × 2.0 cm × 13.0 cm (8.5 in. × 0.8 in. × 5.1 in.)

    Weight

    553 g (19.5 oz)

    Environment

    Maximum altitude

    2,000 m (800 mbar) (at 25 °C ambient temperature)

    Pollution Degree

    2

    Indoor use only.

    Operating Environment

    Ambient temperature range

    0 °C to 55 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2. Meets MIL-PRF-28800F Class 3 low temperature limit and MIL-PRF-28800F Class 2 high temperature limit.)

    Relative humidity range

    10% to 90%, noncondensing (Tested in accordance with IEC 60068-2-56.)

    Storage Environment

    Ambient temperature range

    -40 °C to 71 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2. Meets MIL-PRF-28800F Class 3 limits.)

    Relative humidity range

    5% to 95%, noncondensing (Tested in accordance with IEC 60068-2-56.)

    Shock and Vibration

    Operating shock

    30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Meets MIL-PRF-28800F Class 2 limits.)

    Random vibration

    Operating

    5 Hz to 500 Hz, 0.3 grms (Tested in accordance with IEC 60068-2-64.)

    Nonoperating

    5 Hz to 500 Hz, 2.4 grms (Tested in accordance with IEC 60068-2-64. Test profile exceeds the requirements of MIL-PRF-28800F, Class 3.)

    Compliance and Certifications

    Safety

    This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:

    • IEC 61010-1, EN 61010-1
    • UL 61010-1, CSA 61010-1
    spd-note-note
    Note  

    For UL and other safety certifications, refer to the product label or the Online Product Certification section.

    Electromagnetic Compatibility

    This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
    • EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
    • EN 55011 (CISPR 11): Group 1, Class A emissions
    • EN 55022 (CISPR 22): Class A emissions
    • EN 55024 (CISPR 24): Immunity
    • AS/NZS CISPR 11: Group 1, Class A emissions
    • AS/NZS CISPR 22: Class A emissions
    • FCC 47 CFR Part 15B: Class A emissions
    • ICES-001: Class A emissions
    spd-note-note
    Note  

    In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended for use only in heavy-industrial locations.

    spd-note-note
    Note  

    Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes.

    spd-note-note
    Note  

    For EMC declarations, certifications, and additional information, refer to the Online Product Certification section.

    CE Compliance

    This product meets the essential requirements of applicable European Directives, as follows:

    • 2014/35/EU; Low-Voltage Directive (safety)
    • 2014/30/EU; Electromagnetic Compatibility Directive (EMC)

    Online Product Certification

    Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.

    Environmental Management

    NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.

    For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.

    Waste Electrical and Electronic Equipment (WEEE)

    spd-note-weee
    EU Customers  

    At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee.

    电子信息产品污染控制管理办法(中国RoHS)

    spd-note-china-rohs
    中国客户  

    National Instruments符合中国电子信息产品中限制使用某些有害物质指令(RoHS)。关于National Instruments中国RoHS合规性信息,请登录 ni.com/environment/rohs_china。(For information about China RoHS compliance, go to ni.com/environment/rohs_china.)