Table Of Contents

PXI-2564 Specifications

Version:
    Last Modified: March 2, 2017

    This document lists specifications for the PXI-2564. All specifications are subject to change without notice.

    Definitions

    Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

    Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

    • Typical specifications describe the expected performance met by a majority of the models.
    • Nominal specifications describe parameters and attributes that may be useful in operation.

    Specifications are Typical unless otherwise noted.

    Conditions

    Specifications are valid at 23 °C unless otherwise noted.

    All voltages are specified in DC, ACpk, or a combination unless otherwise specified.

    spd-note-caution
    Caution  

    The protection provided by the PXI-2564 can be impaired if it is used in a manner not described in this document.

    Topology

    Topology

    16 SPST (nonlatching)

    8 DPST

    Input Characteristics

    Maximum switching voltage

    Channel-to-channel

    150 V DC, 150 V AC

    Channel-to-ground

    150 V DC, 150 V AC, CAT II[1]

    spd-note-caution
    Caution  

    This module is rated for Measurement Category II and intended to carry signal voltages no greater than 150 V AC/150 V DC. This module features 1,400 Vrms isolation between input signals and the backplane as verified by a dielectric withstand test, one-minute maximum. Do not use this module for connection to signals or for measurements within Categories III or IV. Do not connect to MAINs supply circuits (for example, wall outlets) greater than 150 V DC; do not connect to MAINs circuits of 208 V AC (US) or 230 V AC (Europe).

    spd-note-caution
    Caution  

    When hazardous voltages (>42.4 Vpk/60 V DC) are present on any relay terminal, safety low-voltage (≤42.4 Vpk/60 V DC) cannot be connected to any other relay terminal.

    spd-note-caution
    Caution  

    The switching power is limited by the maximum switching current and the maximum voltage, and must not exceed 30 W at 150 V DC, 150 W at 30 V DC, or 150 V AC, 5 A.

    Maximum switching power (per channel)

    AC

    750 VA

    DC

    30 V

    150 W

    150 V

    30 W

    Maximum current (switching or carry, per channel)

    5 A

    Simultaneous channels at maximum current (≤55 °C)

    16

    spd-note-caution
    Caution  

    Switching inductive loads (for example, motors and solenoids) can produce high-voltage transients in excess of the rated voltage of the module. Without additional protection, these transients can interfere with module operation and impact relay life. For more information about transient suppression, visit ni.com/info and enter the Info Code induct.

    DC path resistance

    Initial

    <100 mΩ

    End-of-life

    250 mΩ

    DC path resistance typically remains low for the life of the relay. At the end of relay life, the path resistance rises rapidly. Load ratings apply to relays used within the specification before the end of relay life.

    Minimum switch load

    10 mA, 5 V DC

    Dynamic Characteristics

    Relay operate time

    10.4 ms (maximum)

    spd-note-note
    Note  

    Certain applications may require additional time for proper settling. For information about including additional settling time, refer to the NI Switches Help at ni.com/manuals.

    Expected relay life

    Mechanical

    2 × 107 cycles

    Electrical (full load)

    1 × 105 cycles

    spd-note-note
    Note  

    The relays used in the PXI-2564 are field replaceable. Refer to the NI Switches Help at ni.com/manuals for information about replacing a failed relay.

    Trigger Characteristics

    Input trigger

    Sources

    PXI trigger lines <0...7>

    Minimum pulse width

    150 ns

    spd-note-note
    Note  

    The PXI-2564 can recognize trigger pulse widths less than 150 ns if you disable digital filtering. Refer to the NI Switches Help at ni.com/manuals for information about disabling digital filtering.

    Output trigger

    Destinations

    PXI trigger lines <0...7>

    Pulse width

    Programmable (1 µs to 62 µs)

    Physical Characteristics

    Relay type

    Electromechanical, non-latching

    Relay contact material

    Gold-plate silver cadmium oxide

    I/O connector

    37-pin D-SUB, male

    PXI power requirement

    5 W at 5 V, 2.5 W at 3.3 V

    Dimensions (L × W × H)

    3U, one slot, PXI/cPCI module, 21.6 × 2.0 × 13.0 cm (8.5 × 0.8 × 5.1 in.)

    Weight

    267 g (9.5 oz)

    Environment

    Operating temperature

    0 °C to 55 °C

    Storage temperature

    -20 °C to 70 °C

    Relative humidity

    10% to 90%, operational

    5% to 95%, storage

    Pollution Degree

    2

    Maximum altitude

    2,000 m

    Indoor use only.

    Shock and Vibration

    Operational Shock

    30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.)

    Random Vibration

    Operating

    5 Hz to 500 Hz, 0.3 grms

    Nonoperating

    5 Hz to 500 Hz, 2.4 grms (Tested in accordance with IEC 60068-2-64. Nonoperating test profile exceeds the requirements of MIL-PRF-28800F, Class 3.)

    Compliance and Certifications

    Safety

    This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:

    • IEC 61010-1, EN 61010-1
    • UL 61010-1, CSA C22.2 No. 61010-1
    spd-note-note
    Note  

    For UL and other safety certifications, refer to the product label or the Online Product Certification section.

    Electromagnetic Compatibility

    This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
    • EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
    • EN 55011 (CISPR 11): Group 1, Class A emissions
    • EN 55022 (CISPR 22): Class A emissions
    • EN 55024 (CISPR 24): Immunity
    • AS/NZS CISPR 11: Group 1, Class A emissions
    • AS/NZS CISPR 22: Class A emissions
    • FCC 47 CFR Part 15B: Class A emissions
    • ICES-001: Class A emissions
    spd-note-note
    Note  

    In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended for use only in heavy-industrial locations.

    spd-note-note
    Note  

    Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes.

    spd-note-note
    Note  

    For EMC declarations, certifications, and additional information, refer to the Online Product Certification section.

    CE Compliance

    This product meets the essential requirements of applicable European Directives, as follows:

    • 2014/35/EU; Low-Voltage Directive (safety)
    • 2014/30/EU; Electromagnetic Compatibility Directive (EMC)

    Online Product Certification

    Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.

    Environmental Management

    NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.

    For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.

    Waste Electrical and Electronic Equipment (WEEE)

    spd-note-weee
    EU Customers  

    At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee.

    电子信息产品污染控制管理办法(中国RoHS)

    spd-note-china-rohs
    中国客户  

    National Instruments符合中国电子信息产品中限制使用某些有害物质指令(RoHS)。关于National Instruments中国RoHS合规性信息,请登录 ni.com/environment/rohs_china。(For information about China RoHS compliance, go to ni.com/environment/rohs_china.)

    • 1 Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet, for example, 115 V for U.S. or 230 V for Europe.

    Recently Viewed Topics