Table Of Contents

PXI-5142 Specifications

Version:
    Last Modified: December 12, 2017

    Definitions

    Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

    The following characteristic specifications describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

    • Typical specifications describe the performance met by a majority of models.
    • Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.

    Conditions

    Specifications are valid under the following conditions unless otherwise noted.

    • All filter settings
    • All impedance selections
    • Sample clock set to 100 MS/s

    Typical specifications are valid under the following conditions unless otherwise noted:

    • Ambient temperatures of 15 °C to 35 °C

    Vertical

    Analog Input (Channel 0 and Channel 1)

    Number of channels

    Two (simultaneously sampled)

    Connector

    BNC

    Impedance and Coupling

    Input impedance (software-selectable)

    50 Ω ± 2.0%

    1 MΩ ± 0.75% in parallel with a typical capacitance of 27 pF ± 2 pF

    Input coupling (software-selectable)

    AC[1]

    DC

    GND

    Voltage Levels

    Table 1. Full Scale (FS) Input Range and Programmable Vertical Offset
    Range (Vpk-pk) Vertical Offset Range
    50 Ω 1 MΩ
    0.2 V ±0.1 V
    0.4 V ±0.2 V
    1 V ±0.5 V
    2 V ±1 V
    4 V ±2 V
    10 V ±5 V
    20 V
    Maximum input overload

    50 Ω

    7 Vrms with |Peaks| ≤10 V

    1 MΩ

    |Peaks| ≤42 V

    Accuracy

    Resolution

    14 bits

    Table 2. DC Accuracy (Programmable Vertical Offset = 0 V) [2]
    Range (Vpk-pk) 50 Ω 1 MΩ
    0.2 V and 0.4 V ±(0.65% of input + 1.0 mV)
    1 V ±(0.65% of input + 1.2 mV)
    2 V ±(0.65% of input + 1.6 mV)
    4 V ±(0.65% of input + 8.0 mV)
    10 V ±(0.65% of input + 8.0 mV)
    20 V ±(0.65% of input + 13.0 mV)

    Programmable vertical offset accuracy[2]

    ±0.4% of offset setting

    Table 3. DC Drift
    Range (Vpk-pk) 50 Ω 1 MΩ
    0.2 V, 0.4 V, 1 V, and 2 V ±(0.057% of Input + 0.006% of FS + 100 μV) per °C
    4 V, 10 V ±(0.057% of Input + 0.006% of FS + 900 μV) per °C
    20 V ±(0.057% of Input + 0.006% of FS + 900 μV) per °C
    AC amplitude accuracy[2]

    50 Ω

    ±0.06 dB (±0.7%) at 50 kHz, typical

    1 MΩ

    ±0.09 dB (±1.0%) at 50 kHz, typical

    Crosstalk[3]

    -100 dB at 10 MHz, typical

    Bandwidth and Transient Response

    Bandwidth (-3 dB, filters off)

    0.2 Vpk-pk input range

    80 MHz up to 40 °C[4]

    All other input ranges

    100 MHz

    Rise/fall time

    0.2 Vpk-pk input range

    4.2 ns, typical

    All other input ranges

    3.5 ns, typical

    Bandwidth limit filters[5]

    Noise filter

    20 MHz

    2-pole Bessel filter

    Anti-alias Filter

    40 MHz (-6 dB), typical

    35 MHz (-3 dB)

    6-pole Chebyshev filter

    AC coupling cutoff (-3 dB)

    12 Hz[6]

    Table 4. Passband Flatness, Typical [7]
    Filter Settings Input Range (Vpk-pk) 50 Ω and 1 MΩ
    Filters off 0.2 V

    ±0.4 dB (DC to 20 MHz)

    ±1 dB (20 MHz to 40 MHz)

    All other input ranges

    ±0.4 dB (DC to 20 MHz)

    ±1 dB (20 MHz to 50 MHz)

    Anti-alias filter on All input ranges

    ±1.2 dB (DC to 16 MHz)

    ±1.6 dB (16 MHz to 32 MHz)

    Figure 1. PXI-5142 Frequency Response, Typical

    Spectral Characteristics

    Table 5. Spurious-Free Dynamic Range (SFDR) with Harmonics, Typical [8]
    Input Range (Vpk-pk) 50 Ω 1 MΩ
    0.2 V 75 dBc 70 dBc
    0.4 V 75 dBc 70 dBc
    1 V 75 dBc 70 dBc
    2 V 75 dBc 70 dBc
    4 V 65 dBc 70 dBc
    10 V 65 dBc 60 dBc
    20 V 60 dBc
    Table 6. Total Harmonic Distortion (THD), Typical [9]
    Input Range (Vpk-pk) 50 Ω 1 MΩ
    0.2 V -75 dBc -68 dBc
    0.4 V -75 dBc -68 dBc
    1 V -75 dBc -68 dBc
    2 V -73 dBc -68 dBc
    4 V -63 dBc -68 dBc
    10 V -63 dBc -58 dBc
    20 V -58 dBc

    Intermodulation distortion[10]

    -75 dBc, typical

    Table 7. Signal-to-Noise Ratio (SNR), Typical [11]
    Input Range (Vpk-pk) 50 Ω 1 MΩ
    Filters Off Anti-alias Filter On Filters Off Anti-alias Filter On
    0.2 V 60 dB 60 dB 56 dB 60 dB
    0.4 V 62 dB 62 dB 61 dB 62 dB
    1 V 62 dB 62 dB 62 dB 62 dB
    2 V 62 dB 62 dB 62 dB 62 dB
    4 V 61 dB 62 dB
    Table 8. Signal to Noise and Distortion (SINAD), Typical [12]
    Input Range (Vpk-pk) 50 Ω 1 MΩ
    Filters Off Anti-alias Filter On Filters Off Anti-alias Filter On
    0.2 V 60 dB 60 dB 56 dB 59 dB
    0.4 V 62 dB 62 dB 60 dB 61 dB
    1 V 62 dB 62 dB 61 dB 61 dB
    2 V 62 dB 62 dB 61 dB 61 dB
    4 V 60 dB 61 dB
    Figure 2. PXI-5142 Dynamic Performance, 50 Ω, 1 V Input Range, Typical
    Table 9. RMS Noise (Noise Filter On) [13]
    Input Range (Vpk-pk) 50 Ω 1 MΩ
    0.2 V 46 µVrms (0.023% of FS) 60 µVrms (0.030% of FS)
    0.4 V 92 μVrms (0.023% of FS) 92 μVrms (0.023% of FS)
    1 V 230 μVrms (0.023% of FS) 230 μVrms (0.023% of FS)
    2 V 460 μVrms (0.023% of FS) 460 μVrms (0.023% of FS)
    4 V 920 μVrms (0.023% of FS) 920 μVrms (0.023% of FS)
    10 V 2.3 mVrms (0.023% of FS) 2.3 mVrms (0.023% of FS)
    20 V 4.6 mVrms (0.023% of FS)
    Table 10. RMS Noise (Anti-alias Filter On) [13]
    Input Range (Vpk-pk) 50 Ω 1 MΩ
    0.2 V 66 µVrms (0.033% of FS) 80 µVrms (0.040% of FS)
    0.4 V 100 μVrms (0.025% of FS) 120 μVrms (0.030% of FS)
    1 V 250 μVrms (0.025% of FS) 300 μVrms (0.030% of FS)
    2 V 500 μVrms (0.025% of FS) 600 μVrms (0.030% of FS)
    4 V 1 mVrms (0.025% of FS) 1.2 mVrms (0.030% of FS)
    10 V 2.5 mVrms (0.025% of FS) 3 mVrms (0.030% of FS)
    20 V 6 mVrms (0.030% of FS)
    Table 11. RMS Noise (Filters Off) [13]
    Input Range (Vpk-pk) 50 Ω 1 MΩ
    0.2 V 66 µVrms (0.033% of FS) 110 μVrms (0.055% of FS)
    0.4 V 100 μVrms (0.025% of FS) 160 μVrms (0.040% of FS)
    1 V 250 μVrms (0.025% of FS) 300 μVrms (0.030% of FS)
    2 V 500 μVrms (0.025% of FS) 600 μVrms (0.030% of FS)
    4 V 1 mVrms (0.025% of FS) 1.6 mVrms (0.040% of FS)
    10 V 2.5 mVrms (0.025% of FS) 3 mVrms (0.030% of FS)
    20 V 6 mVrms (0.030% of FS)
    Figure 3. Representation of PXI-5142 Spectral Noise Density on 0.2 V Input Range, Full Bandwidth, 50 Ω Input Impedance
    Figure 4. Representation of PXI-5142 Spectral Noise Density on 0.2 V Input Range, Noise Filter Enabled, 1 MΩ Input Impedance

    Horizontal

    Sample Clock

    Sources

    Internal

    Onboard clock (internal VCXO)[14]

    External

    CLK IN (front panel SMB connector)

    PXI Star Trigger (backplane connector)

    Onboard Clock (Internal VCXO)

    Sample rate range

    Real-time sampling (single shot)

    1.526 kS/s to 100 MS/s[15]

    Random interleaved sampling (RIS)

    200 MS/s to 2 GS/s in multiples of 100 MS/s

    Phase noise density[16]

    100 Hz input frequency

    <-100 dBc/Hz, typical

    1 kHz input frequency

    <-120 dBc/Hz, typical

    10 kHz input frequency

    <-130 dBc/Hz, typical

    Sample clock jitter[17]

    ≤1 psrms (100 Hz to 100 kHz), typical

    ≤2 psrms (100 Hz to 1 MHz), typical

    Timebase frequency

    100 MHz

    Timebase accuracy

    Not phase-locked to Reference clock

    ±25 ppm

    Phase-locked to Reference clock

    Equal to the Reference clock accuracy

    Sample clock delay range

    ±1 Sample clock period

    Sample clock delay resolution

    ≤10 ps

    External Sample Clock

    Sources

    CLK IN (front panel SMB connector)

    PXI Star Trigger (backplane connector)

    requency range[18]

    CLK IN

    30 MHz to 105 MHz

    PXI Star Trigger

    30 MHz to 80 MHz

    Duty cycle tolerance

    45% to 55%

    Sample Clock Exporting

    Table 12. Exported Sample Clock Destinations
    Destination Maximum Frequency
    CLK OUT (front panel SMB connector) 105 MHz
    PXI_Trig <0..6> (backplane connector)[19] 20 MHz
    PFI <0..1> (front panel 9-pin mini-circular DIN connector)[19] 25 MHz
    RTSI <0..6>[19] 20 MHz

    Phase-Locked Loop (PLL) Reference Clock

    Sources

    PXI_CLK10 (backplane connector)

    CLK IN (front panel SMB connector)

    Frequency range

    1 MHz to 20 MHz in 1 MHz increments[20]

    Duty cycle tolerance

    45% to 55%

    Exported Reference clock destinations

    CLK OUT (front panel SMB connector)

    PFI <0..1> (front panel 9-pin mini-circular DIN connector)

    PXI_Trig <0..7> (backplane connector)

    CLK IN (Sample Clock and Reference Clock Input, Front Panel Connector)

    Input voltage range

    Sine wave (Vpk-pk)

    0.65 V to 2.8 V (0 dBm to 13 dBm)

    Square wave (Vpk-pk)

    0.2 V to 2.8 V

    Maximum input overload[21]

    7 Vrms with |Peaks| ≤10 V

    Impedance

    50 Ω

    Coupling

    AC

    CLK OUT (Sample Clock and Reference Clock Output, Front Panel Connector)

    Output impedance

    50 Ω

    Logic type

    3.3 V CMOS

    Maximum drive current

    ±48 mA

    Trigger

    Reference (Stop) Trigger

    Trigger types

    Edge

    Window

    Hysteresis

    Video

    Digital

    Immediate

    Software

    Trigger sources

    CH 0

    CH 1

    TRIG

    PXI_Trig <0..6>

    PFI <0..1>

    PXI Star Trigger

    Software

    RTSI <0..6>

    spd-note-note
    Note  

    Refer to the following sections and the NI High-Speed Digitizers Help for more information about what sources are available for each trigger type.

    Table 13. Time Resolution
    Time-to-Digital Conversion Circuit (TDC) Onboard Clock External Clock
    On 100 ps
    Off 10 ns External clock period
    Table 14. Minimum Rearm Time [22]
    TDC Rearm Time
    On 10 μs
    Off 2 μs
    Table 15. Holdoff
    TDC Onboard Clock External Clock
    On 10 μs to 171.79 s
    Off 2 μs to 171.79 s 200 × External clock period to (232 - 1) × External clock period

    Analog Trigger

    Trigger types

    Edge

    Window

    Hysteresis

    Sources

    CH 0 (front panel BNC connector)

    CH 1 (front panel BNC connector)

    TRIG (front panel BNC connector)

    Trigger level range

    CH 0, CH 1

    100% of FS

    TRIG (external trigger)

    ±5 V

    Trigger level resolution

    10 bits (1 in 1,024)

    Edge trigger sensitivity

    CH 0, CH 1

    2.5% of FS up to 50 MHz

    Increases to 5% of FS at 100 MHz

    TRIG (external trigger, Vpk-pk)

    0.25 V up to 100 MHz

    Increases to 1 V at 200 MHz

    Level accuracy

    CH 0, CH 1

    ±3.5% of FS up to 10 MHz, typical

    TRIG (external trigger)

    ±0.35 V (±3.5% of FS) up to 10 MHz, typical

    Jitter

    ≤80 psrms[23]

    Trigger filters

    Low-frequency (LF) reject

    50 kHz

    High-frequency (HF) reject

    50 kHz

    Digital Trigger

    Trigger type

    Digital

    Sources

    PXI_Trig <0..6> (backplane connector)

    PFI <0..1> (front panel 9-pin mini-circular DIN connector)

    PXI Star Trigger (backplane connector)

    Video Trigger

    Trigger type

    Video

    Sources

    CH 0 (front panel BNC connector)

    CH 1 (front panel BNC connector)

    TRIG (front panel BNC connector)

    Video trigger types

    Specific Line

    Any Line

    Specific Field

    Standard

    Negative sync of NTSC, PAL, or SECAM signal

    External Trigger

    Connector

    TRIG (front panel BNC connector)

    Impedance

    1 MΩ in parallel with 22 pF

    Coupling

    AC, DC

    AC coupling cutoff (-3 dB)

    12 Hz

    Input voltage range

    ±5 V

    Maximum input overload

    |Peaks| ≤42 V

    PFI 0 and PFI 1 (Programmable Function Interface)

    Connector

    AUX I/O (9-pin mini-circular DIN)

    Direction

    Bidirectional

    As an Input (Trigger)

    Destinations

    Start trigger (acquisition arm)

    Reference (stop) trigger

    Arm Reference trigger

    Advance trigger

    Input impedance

    150 kΩ

    VIH

    2.0 V

    VIL

    0.8 V

    Maximum input overload

    -0.5 V to 5.5 V

    Maximum frequency

    25 MHz

    As an Output (Event)

    Sources

    Ready for Start

    Start trigger (acquisition arm)

    Ready for Reference

    Reference (stop) trigger

    End of Record

    Ready for Advance

    Advance trigger

    Done (end of acquisition)

    Probe Compensation[24]

    Output impedance

    50 Ω

    Logic type

    3.3 V CMOS

    Maximum drive current

    ±24 mA

    Maximum frequency

    25 MHz

    Waveform Specifications

    Onboard memory size

    64 MB per channel option

    32 MS per channel[25]

    256 MB per channel option

    128 MS per channel[25]

    Minimum record length

    1 sample

    Number of pretrigger samples

    Zero up to full record length[26]

    Number of posttrigger samples

    Zero up to full record length[26]

    Maximum number of records in onboard memory

    64 MB/channel

    100,000[27]

    256 MB/channel

    100,000[27]

    Allocated onboard memory per record

    Real data processing mode

    (Record Length × 2 bytes/S) + 200 bytes, rounded up to next multiple of 128 bytes

    or

    512 bytes, whichever is greater

    Complex data processing mode

    (Record Length × 4 bytes/S) + 200 bytes, rounded up to next multiple of 128 bytes

    or

    512 bytes, whichever is greater

    Onboard Signal Processing (OSP)

    Figure 5. PXI-5142 Onboard Signal Processing Block Diagram
    spd-note-note
    Note  

    To use onboard signal processing (OSP) on the PXI-5142, the DDC Enabled property/attribute must be set to TRUE.

    The following four OSP operations are available:

    • Send one IF signal to CH 0 and perform quadrature downconversion on the signal (complex data is returned).
    • Send I and Q baseband signals to CH 0 and CH 1 and perform alias-protected decimation (complex data is returned).
    • Send a signal to CH 0 and perform alias-protected decimation (real data is returned).
    • Send a signal to CH 0 and perform real downconversion on the signal (real data is returned).

    Number of digital downconverters (DDCs)

    1

    Data processing modes[28]

    Real (I path only)

    Complex (IQ)

    OSP decimation range[29]

    1, 2, 4, 6, 8, 10

    12 to 4,096 (multiples of 4)

    4,096 to 8,192 (multiples of 8)

    8,192 to 16,384 (multiples of 16)

    Sample rate range[30]

    Internal Sample clock timebase

    6.1 kS/s to 100 MS/s (real or complex)

    External Sample clock timebase

    Sample clock timebase/OSP decimation

    Real flat bandwidth

    0.4 × Sample Rate

    Complex flat bandwidth[31]

    0.8 × Sample Rate

    Digital Gain and Offset

    Digital gain and offset resolution

    18 bits

    Digital gain range

    -1.5 to +1.5

    |Values| < 1 attenuate user data

    Digital offset range

    (-0.4 × Vertical Range) to (+0.4 × Vertical Range)[32]

    Output

    (ADC Data × Digital Gain) + Digital Offset.[33]

    Numerically-Controlled Oscillator (NCO)

    Frequency range[34]

    Internal Sample clock timebase

    0 Hz to 50 MHz

    External Sample clock timebase

    0 Hz to (0.5 × Sample Clock Timebase)

    Frequency resolution

    Internal Sample clock timebase

    355 nHz

    External Sample clock timebase

    Sample Clock Timebase/248

    I and Q phase resolution

    0.0055 º

    Tuning time

    1 ms

    Digital Performance

    Maximum NCO spur

    <-100 dBFS

    Decimating filter passband ripple

    <0.1 dB[35]

    Decimating filter out-of-band suppression

    >80 dB[36]

    IF Demodulation Performance

    Table 16. IF Demodulation Performance (Typical)
    Modulation Configuration[37] Measurement Type Value
    GSM Physical Layer[38] Modulation Error Ratio (MER) 62 dB
    Error Vector Magnitude (EVM) <0.2% rms
    W-CDMA Physical Layer[39] MER 52 dB
    EVM <0.4% rms
    DVB Physical Layer[40] MER 48 dB
    EVM <0.4% rms
    20 MSymbols/s, 64 QAM[41] MER 39 dB
    EVM <0.8% rms
    26.09 MSymbols/s, 64 QAM[42] MER 36 dB
    EVM <1.0% rms
    34.78 MSymbols/s, 64 QAM[43] MER 32 dB
    EVM <1.6% rms

    IQ Baseband Demodulation Performance

    Table 17. IQ Baseband Demodulation Performance (Typical)
    Modulation Configuration[44] Measurement Type Value
    GSM physical layer[45] Modulation Error Ratio (MER) 41 dB
    Error Vector Magnitude (EVM) <0.8% rms
    W-CDMA Physical Layer[46] MER 41 dB
    EVM <0.9% rms
    DVB Physical Layer[47] MER 40 dB
    EVM <0.9% rms
    20 MSymbols/s, 64 QAM[48] MER 33 dB
    EVM <1.4% rms

    Waveform Acquisition Times

    Table 18. Maximum Acquisition Time [49]
    Conditions 64 MB 256 MB
    Sample rate = 100 MS/s, OSP disabled 0.336 s 1.34 s
    Sample rate = 1 MS/s, real mode, OSP enabled 33.6 s 2 min 14 s
    Sample rate = 100 kS/s, real mode, OSP enabled 5 min 36 s 22 min 22 s
    Figure 6. Decimation Filter Frequency Response (Real Mode), 10 MS/s Sample Rate
    Figure 7. GSM Physical Layer [50]
    Figure 8. W-CDMA Physical Layer [51]
    Figure 9. DVB Physical Layer [52]
    Figure 10. 20 MSymbols/s 64 QAM [53]

    Calibration

    External Calibration

    External calibration calibrates the VCXO and the voltage reference. All calibration constants are stored in nonvolatile memory.

    Self-Calibration

    Self-calibration is done on software command. The calibration corrects for gain, offset, frequency response, triggering, and timing adjustment errors for all input ranges.

    Calibration Specifications

    Interval for external calibration

    2 years

    Warm-up time

    15 minutes

    Software

    Driver Software

    Driver support for this device was first available in NI-SCOPE 3.0.

    NI-SCOPE is an IVI-compliant driver that allows you to configure, control, and calibrate the PXI-5142. NI-SCOPE provides application programming interfaces for many development environments.

    Application Software

    NI-SCOPE provides programming interfaces, documentation, and examples for the following application development environments:

    • LabVIEW
    • LabWindows™/CVI™
    • Measurement Studio
    • Microsoft Visual C/C++
    • .NET (C# and VB.NET)

    Interactive Soft Front Panel and Configuration

    The NI-SCOPE Soft Front Panel (SFP) allows interactive control of the PXI-5142.

    Interactive control of the PXI-5142 was first available in NI-SCOPE SFP version 2.4. The NI-SCOPE SFP is included on the NI-SCOPE media.

    NI Measurement Automation Explorer (MAX) also provides interactive configuration and test tools for the PXI-5142. MAX is included on the NI-SCOPE media.

    TClk Specifications

    You can use the NI TClk synchronization method and the NI-TClk driver to align the Sample clocks on any number of supported devices, in one or more chassis. For more information about TClk synchronization, refer to the NI-TClk Synchronization Help, which is located within the NI High-Speed Digitizers Help. For other configurations, including multichassis systems, contact NI Technical Support at ni.com/support.

    Intermodule SMC Synchronization Using NI-TClk for Identical Modules

    Synchronization specifications are valid under the following conditions:
    • All modules are installed in one NI PXI-1042 chassis.
    • The NI-TClk driver is used to align the Sample clocks of each module
    • All parameters set to identical values for each module.
    • Sample clock set to 100 MS/s
    • All filters are disabled.
    spd-note-note
    Note  

    Although you can use NI-TClk to synchronize non-identical SMC-based modules, these specifications apply only to synchronizing identical modules.

    Skew, typical[54]

    500 ps

    Average skew after manual adjustment, typical

    ≤5 ps

    Sample clock delay/adjustment resolution, typical

    ≤5 ps

    Power

    Current draw

    +3.3 VDC

    1.0 A, typical

    +5 VDC

    1.7 A, typical

    +12 VDC

    800 mA, typical

    -12 VDC

    270 mA, typical

    Total power

    24.7 W, typical

    Physical

    Dimensions

    3U, one-slot, PXI/cPCI module

    21.6 cm × 2.0 cm × 13.0 cm

    (8.5 in × 0.8 in × 5.1 in)

    Weight

    459 g (16.2 oz)

    Environment

    Maximum altitude

    2,000 m (at 25 °C ambient temperature)

    Pollution Degree

    2

    Indoor use only.

    Operating Environment

    Ambient temperature range

    0 °C to 45 °C when installed in an NI PXI-1000/B or PXI-101x chassis.

    0 °C to 55 °C when installed in any other NI PXI chassis.

    (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.)

    Relative humidity range

    10% to 90%, noncondensing (Tested in accordance with IEC 60068-2-56.)

    Storage Environment

    Ambient temperature range

    -40 °C to 71 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.)

    Relative humidity range

    5% to 95%, noncondensing (Tested in accordance with IEC 60068-2-56.)

    Shock and Vibration

    Operational shock

    30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.)

    Storage shock

    50 g peak, half-sine, 11 ms pulse (Meets IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.)

    Random vibration

    Operating

    5 Hz to 500 Hz, 0.31 grms (Tested in accordance with IEC 60068-2-64.)

    Nonoperating

    5 Hz to 500 Hz, 2.46 grms (Tested in accordance with IEC 60068-2-64. Test profile exceeds the requirements of MIL-PRF-28800F, Class 3.)

    Compliance and Certifications

    Safety

    This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:

    • IEC 61010-1, EN 61010-1
    • UL 61010-1, CSA C22.2 No. 61010-1
    spd-note-note
    Note  

    For UL and other safety certifications, refer to the product label or the Online Product Certification section.

    Electromagnetic Compatibility

    This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
    • EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
    • EN 55011 (CISPR 11): Group 1, Class A emissions
    • EN 55022 (CISPR 22): Class A emissions
    • EN 55024 (CISPR 24): Immunity
    • AS/NZS CISPR 11: Group 1, Class A emissions
    • AS/NZS CISPR 22: Class A emissions
    • FCC 47 CFR Part 15B: Class A emissions
    • ICES-001: Class A emissions
    spd-note-note
    Note  

    In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended for use only in heavy-industrial locations.

    spd-note-note
    Note  

    Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes.

    spd-note-note
    Note  

    For EMC declarations, certifications, and additional information, refer to the Online Product Certification section.

    CE Compliance

    This product meets the essential requirements of applicable European Directives, as follows:

    • 2014/35/EU; Low-Voltage Directive (safety)
    • 2014/30/EU; Electromagnetic Compatibility Directive (EMC)

    Online Product Certification

    Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.

    Environmental Management

    NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.

    For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.

    Waste Electrical and Electronic Equipment (WEEE)

    spd-note-weee
    EU Customers  

    At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee.

    电子信息产品污染控制管理办法(中国RoHS)

    spd-note-china-rohs
    中国客户  

    National Instruments符合中国电子信息产品中限制使用某些有害物质指令(RoHS)。关于National Instruments中国RoHS合规性信息,请登录 ni.com/environment/rohs_china。(For information about China RoHS compliance, go to ni.com/environment/rohs_china.)

    • 1 AC coupling available on 1 MΩ input only.
    • 2 Within ±5 °C of self-calibration temperature.
    • 3 CH 0 to/from CH 1, and external trigger to CH 0 or CH 1.
    • 4 78 MHz above 40 °C.
    • 5 Only one filter can be enabled at any given time. The anti-alias filter is enabled by default.
    • 6 AC coupling available on 1 MΩ input impedance only.
    • 7 Referenced to 50 kHz.
    • 8 10 MHz, -1 dBFS input signal. Includes the 2nd through the 5th harmonics. Measured from DC to 50 MHz.
    • 9 10 MHz, -1 dBFS input signal. Includes the 2nd through the 5th harmonics.
    • 10 0.2 V to 2.0 V input ranges with 50 Ω input impedance. Two tones at 10.2 MHz and 11.2 MHz. Each tone is -7 dBFS.
    • 11 10 MHz, -1 dBFS input signal. Excludes harmonics. Measured from DC to 50 MHz.
    • 12 10 MHz, -1 dBFS input signal. Includes harmonics. Measured from DC to 50 MHz.
    • 13 50 Ω terminator connected to input.
    • 14 Internal Sample clock is locked to the Reference clock or derived from the onboard VCXO.
    • 15 In normal operation mode (non-OSP mode), divide by n decimation is used for all rates less than 100 MS/s. Non-OSP decimation does not protect the acquired data from undersampling aliasing. Non-OSP decimation and OSP decimation are mutually exclusive.
    • 16 10 MHz input signal.
    • 17 Includes the effects of the converter aperture uncertainty and the clock circuitry jitter. Excludes trigger jitter.
    • 18 In normal operation mode (non-OSP mode), divide by n decimation is available, where 1 ≤ n ≤ 65,535. Non-OSP decimation does not protect the acquired data from undersampling aliasing. Non-OSP decimation and OSP decimation are mutually exclusive.
    • 19 Decimated Sample clock only.
    • 20 10 MHz default. The PLL Reference clock frequency must be accurate to ±50 ppm.
    • 21 Overvoltage and reverse polarity protected.
    • 22 Holdoff set to 0. Onboard Sample clock at maximum rate.
    • 23 Within ±5 °C of self-calibration temperature.
    • 24 1 kHz, 50% duty cycle square wave. PFI 1 front panel connector only.
    • 25 Assumes 2-byte samples. In Complex data processing mode (only available when using onboard signal processing), each sample is 4 bytes, so this number is halved.
    • 26 Single-record mode and multiple-record mode.
    • 27 It is possible to exceed these numbers if you fetch records while acquiring data.
    • 28 Complex mode is used for both IQ baseband decimation and quadrature downconversion.
    • 29 OSP decimation protects acquired data from high-frequency aliasing within the ADC Nyquist zone, whereas non-OSP decimation does not. Non-OSP decimation and OSP decimation are mutually exclusive.
    • 30 For sample rates less than 6.1 kS/s, use an external Sample clock or perform additional software decimation.
    • 31 For example, complex bandwidth is 40 MHz with a complex sample rate of 50 MS/s.
    • 32 Applied after digital gain.
    • 33 (-0.5 × Vertical Range) ≤ Output ≤ (+0.5 × Vertical Range)
    • 34 Undersampling can be used for carrier frequencies >50 MHz.
    • 35 Passband is from 0 to (0.4 × IQ Rate).
    • 36 Stopband suppression from (0.6 × IQ Rate).
    • 37 1 V vertical range, 50 Ω input impedance, no analog filter, 25 MHz carrier. Demodulation, including resampling (sample rate conversion) and pulse shaping, was done with the NI Modulation Toolkit in the host PC/ controller.
    • 38 1.25 MS/s sample rate, MSK modulation, 270.833 kSymbols/s, Gaussian, BT = 0.3.
    • 39 6.25 MS/s sample rate, QPSK modulation, 3.84 MSymbols/s, root raised cosine, alpha = 0.22.
    • 40 10 MS/s sample rate, 32 QAM modulation, 6.92 MSymbols/s, root raised cosine, alpha = 0.15.
    • 41 50 MS/s sample rate, 64 QAM modulation, 20 MSymbols/s, root raised cosine, alpha = 0.15.
    • 42 50 MS/s sample rate, 64 QAM modulation, 26.09 MSymbols/s, root raised cosine, alpha = 0.15, 30 MHz bandwidth.
    • 43 50 MS/s sample rate, 64 QAM modulation, 34.78 MSymbols/s, root raised cosine, alpha = 0.15, 40 MHz bandwidth.
    • 44 1 V input range, 50 Ω input impedance, no analog filter. Demodulation, including resampling (sample rate conversion) and pulse shaping, done with the NI Modulation Toolkit on the host PC/controller. This is a measurement of system performance. The IQ Baseband generation was implemented with two TClk-synchronized NI PXI-5421 arbitrary waveform generators.
    • 45 1.25 MS/s sample rate, MSK modulation, 270.833 kSymbols/s, Gaussian, BT = 0.3
    • 46 6.25 MS/s sample rate, QPSK modulation, 3.84 MSymbols/s, root raised cosine, alpha = 0.22
    • 47 10 MS/s sample rate, 32 QAM modulation, 6.92 MSymbols/s, root raised cosine, alpha = 0.15
    • 48 50 MS/s sample rate, 64 QAM modulation, 20 MSymbols/s, root raised cosine, alpha = 0.15
    • 49 For complex (IQ) mode, the acquisition time is halved.
    • 50 1 V input range, 50 Ω input impedance, no analog filter, 25 MHz carrier, 4.17 MS/s sample rate, MSK modulation, 270.833 kSymbols/s, Gaussian, BT = 0.3.
    • 51 1 V input range, 50 Ω input impedance, no analog filter, 25 MHz carrier, 25 MS/s sample rate, QPSK modulation, 3.84 MSymbols/s, root raised cosine, alpha = 0.22.
    • 52 1 V input range, 50 Ω input impedance, no analog filter, 25 MHz carrier, 25 MS/s sample rate, 32 QAM modulation, 6.92 MSymbols/s, root raised cosine, alpha = 0.15.
    • 53 1 V input range, 50 Ω input impedance, no analog filter, 25 MHz carrier, 50 MS/s sample rate, 64 QAM modulation, 20 MSymbols/s, root raised cosine, alpha = 0.15.
    • 54 Caused by clock and analog path delay differences. No manual adjustment performed.

    Recently Viewed Topics