Table Of Contents

PXI-2800 Specifications

Version:
    Last Modified: April 19, 2017

    This document lists specifications for the PXI-2800. All specifications are subject to change without notice.

    Definitions

    Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

    Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

    • Typical specifications describe the expected performance met by a majority of the models.
    • Nominal specifications describe parameters and attributes that may be useful in operation.

    Specifications are Typical unless otherwise noted.

    Conditions

    Specifications are valid at 23 °C unless otherwise noted.

    All voltages are specified in DC, ACpk, or a combination unless otherwise specified.

    spd-note-caution
    Caution  

    The protection provided by the PXI-2800 can be impaired if it is used in a manner not described in this document.

    Analog Bus Characteristics

    Maximum voltage (channel-to-ground)

    150 V , CAT I

    spd-note-caution
    Caution  

    This module is rated for Measurement Category I. It is intended to carry signal voltages no greater than 100 V RMS, 150 V PK, or 150 V DC. This module can withstand up to 800 V impulse voltage. Do not use this module for connection to signals or for measurements within Categories II, III, or IV. Do not connect MAINS supply circuits (for example, wall outlets) of 115 V AC or 230 V AC.

    spd-note-caution
    Caution  

    When hazardous voltages (>42.4 V PK/60 V DC) are present on any relay terminal, safety low-voltage (≤42.4 V PK /60 V DC) cannot be connected to any other relay terminal. This includes all cards in the carrier and all cards in other carriers connected via the NI 2806 Expansion Bridge for NI SwitchBlock.

    spd-note-caution
    Caution  

    The maximum voltage is limited to the lowest voltage of any component in the NI SwitchBlock system. Review the specifications of the NI SwitchBlock carrier and cards for this information.

    Maximum current (per analog bus channel)

    2 A

    DC path resistance, analog bus

    <0.3 Ω

    Power dissipation limit

    Depends on installed cards

    spd-note-note
    Note  

    Power dissipation limits depend on the installed cards. Refer to the card specifications and choose the lower number. For more information, visit ni.com/info and enter the Info Code sbpwrlim.

    Physical Characteristics

    Power requirement

    PXI

    5 W at 3.3 V

    20 W at 5 V

    Dimensions (L × W × H)

    3U, four slot, PXI/cPCI module, PXIe compatible, 18.7 cm × 8.1 cm × 12.9 cm (7.4 in. × 3.2 in. × 5.1 in.)

    Weight

    560 g (1 lb 4 oz)

    Environment

    Maximum altitude

    2,000 m (at 25 °C ambient temperature)

    Pollution Degree

    2

    Indoor use only.

    Operating Environment

    Ambient temperature range

    0 °C to 55 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.)

    Relative humidity range

    10% to 90%, noncondensing (Tested in accordance with IEC 60068-2-56.)

    Storage Environment

    Ambient temperature range

    -20 °C to 70 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.)

    Relative humidity range

    5% to 95%, noncondensing (Tested in accordance with IEC 60068-2-56.)

    Shock and Vibration

    Operational shock

    30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.)

    Random vibration

    Operating

    5 Hz to 500 Hz, 0.31 grms (Tested in accordance with IEC 60068-2-64.)

    Nonoperating

    5 Hz to 500 Hz, 2.46 grms (Tested in accordance with IEC 60068-2-64. Test profile exceeds the requirements of MIL-PRF-28800F, Class 3.)

    Compliance and Certifications

    Safety

    This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:

    • IEC 61010-1, EN 61010-1
    • UL 61010-1, CSA C22.2 No. 61010-1
    spd-note-note
    Note  

    For UL and other safety certifications, refer to the product label or the Online Product Certification section.

    Electromagnetic Compatibility

    This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
    • EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
    • EN 55011 (CISPR 11): Group 1, Class A emissions
    • EN 55022 (CISPR 22): Class A emissions
    • EN 55024 (CISPR 24): Immunity
    • AS/NZS CISPR 11: Group 1, Class A emissions
    • AS/NZS CISPR 22: Class A emissions
    • FCC 47 CFR Part 15B: Class A emissions
    • ICES-001: Class A emissions
    spd-note-note
    Note  

    In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended for use only in heavy-industrial locations.

    spd-note-note
    Note  

    Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes.

    spd-note-note
    Note  

    For EMC declarations, certifications, and additional information, refer to the Online Product Certification section.

    CE Compliance

    This product meets the essential requirements of applicable European Directives, as follows:

    • 2014/35/EU; Low-Voltage Directive (safety)
    • 2014/30/EU; Electromagnetic Compatibility Directive (EMC)

    Online Product Certification

    Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.

    Environmental Management

    NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.

    For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.

    Waste Electrical and Electronic Equipment (WEEE)

    spd-note-weee
    EU Customers  

    At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee.

    电子信息产品污染控制管理办法(中国RoHS)

    spd-note-china-rohs
    中国客户  

    National Instruments符合中国电子信息产品中限制使用某些有害物质指令(RoHS)。关于National Instruments中国RoHS合规性信息,请登录 ni.com/environment/rohs_china。(For information about China RoHS compliance, go to ni.com/environment/rohs_china.)


    Recently Viewed Topics