Disconnect Trigger Terminals (G Dataflow)

Version:
Last Modified: March 15, 2017

Closes a route between a source trigger terminal and a destination trigger terminal.

After you invoke this function, you can use the associated terminal for other operations. You must use this node any time you connect trigger terminals using Connect Trigger Terminals.

connector_pane_image
datatype_icon

instrument handle in

The instrument handle that you obtain from Initialize.

datatype_icon

error in

Error conditions that occur before this node runs.

The node responds to this input according to standard error behavior.

Standard Error Behavior

Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

error in does not contain an error error in contains an error
If no error occurred before the node runs, the node begins execution normally.

If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

Default: No error

datatype_icon

source terminal

The source trigger terminal to connect to the destination terminal.

Name Description
PXI_Trig<n> The basic trigger lines of a PXI or PXIe chassis.
PXI_Star<n> The star trigger lines of a PXI or PXIe chassis.
spd-note-note
Note  

Each PXI_Star trigger is mapped to a single slot. This mapping is vendor specific. Refer to the hardware documentation to determine the orientation of PXI_Star lines in the chassis.

PFI<n> The PFI connectors on the front panel of the module. You can use PFI connectors to route triggers between multiple chassis or devices.
PFI_LVDS<n> The PFI low voltage differential signaling (LVDS) input/output connectors on the front panel of your device. Signals on PFI LVDS lines use the standard PFI synchronization clock.
Ground The Ground connector continuously outputs a logic low signal, unless it is inverted.
Full Speed Clock The full speed synchronization clock signal of the destination terminal zone. Use this source to send a full-speed clock signal along a trigger line (for example, to route a PXI_Clk10 clock signal to a PFI line).
spd-note-caution
Caution  

Routing a clock signal through a PXI_Trig line is not recommended, due to poor clock signal integrity caused by the topology of the PXI_Trig lines. Use PXI_Star, PXIe_DStar, or PFI lines instead.

Divided Clock 1 The first divided clock signal of the destination terminal zone. This source divides the synchronization clock of the destination terminal by the value you specify in the Clock Divisor 1 parameter of niSync Properties and uses the result as the trigger source.
Divided Clock 2 The second divided clock signal of the destination terminal zone. This source divides the synchronization clock of the destination terminal by the value you specify in the Clock Divisor 2 property of niSync Properties and uses the result as the trigger source.
ClkIn The ClkIn connector on the front panel of the device. Use this terminal to route triggers from an external device.
PXIe_DStarC<n> The differential star trigger lines of the PXIe chassis. Use PXIe_DStarC lines to send trigger and clock signals from a peripheral slot to the system timing slot of the chassis.
spd-note-note
Note  

Each PXIe_DStarC trigger is mapped to a single slot. This mapping is vendor specific. Refer to your chassis documentation to determine the orientation of differential star trigger lines.

PXIe_DStarB The differential star trigger lines of the PXIe chassis. Use PXIe_DStarB lines to send trigger signals from the system timing slot to a peripheral slot of the chassis.
datatype_icon

destination terminal

The destination trigger terminal to which the source terminal connects.

Name Description
PXI_Trig<n> The basic trigger lines of the PXI or PXIe chassis.
PXI_Star<n> The star trigger lines of the PXI or PXIe chassis. Each star trigger line is a dedicated connection between the system timing slot and one other slot.
spd-note-note
Note  

Each PXI_Star trigger is mapped to a single slot. This mapping is vendor specific. Refer to the hardware documentation to determine the orientation of PXI_Star lines in the chassis.

PFI<n> The PFI connectors on the front panel of the module.
PFI_LVDS<n> The PFI low voltage differential signaling (LVDS) input/output connectors on the front panel of the module.
PXIe_DStarB<n> The differential star trigger lines of the PXIe chassis. Use PXIe_DStarB lines to send trigger signals from the system timing slot to a peripheral slot of the chassis.
spd-note-note
Note  

Each PXIe_DStarB trigger is mapped to a single slot. This mapping is vendor specific. Refer to the chassis documentation to determine the orientation of differential star trigger lines.

PXIe_DStarC The differential star trigger lines of the PXIe chassis. Use PXIe_DStarC lines to send trigger and clock signals from a peripheral slot to the system timing slot of the chassis.
datatype_icon

instrument handle out

The session handle that you obtain from Initialize.

datatype_icon

error out

Error information.

The node produces this output according to standard error behavior.

Standard Error Behavior

Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

error in does not contain an error error in contains an error
If no error occurred before the node runs, the node begins execution normally.

If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

Where This Node Can Run:

Desktop OS: Windows

FPGA: Not supported