Configure Scan Trigger (G Dataflow)

Version:
Last Modified: January 17, 2018

Configures the scan triggers for the scan list established with the niSwitch Configure Scan List node.

This node sets the location where the switch expects to receive an input trigger to advance through the scan list. This node also sets the location where it generates a scan advanced signal after it completes an entry in the scan list.

connector_pane_image
datatype_icon

scan delay

Minimum length of time the switch device waits after it creates a path until it asserts a trigger on the Scan Advanced Output channel. NI-SWITCH uses this value to set the Scan Delay property. The scan delay is in addition to the settling time.

Default: 0.0 (seconds)

datatype_icon

session in

Control that identifies a particular NI-SWITCH session established with the niSwitch Initialize With Topology node, the niSwitch Initialize With Options node, or the niSwitch Initialize node.

datatype_icon

error in

Error conditions that occur before this node runs.

The node responds to this input according to standard error behavior.

Standard Error Behavior

Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

error in does not contain an error error in contains an error
If no error occurred before the node runs, the node begins execution normally.

If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

Default: No error

datatype_icon

trigger input

Trigger source used by the switch module during scanning. NI-SWITCH uses this value to set the Trigger Input property.

The switch module waits for a trigger at the specified location when it encounters a semicolon in the scan list. When the trigger occurs, the switch device advances to the next entry in the scan list and waits for a trigger from the location specified in trigger input.

Default: External

datatype_icon

scan advanced output (External)

Output terminal for the Scan Advanced Output trigger signal. NI-SWITCH uses this value to set the Scan Advanced Output property.

After the switch processes each entry in the scan list, it waits the length of time specified in scan delay, then asserts a trigger on the line specified in scan advanced output.

Name Description
None

No Scan Advanced Output trigger is produced.

External

Sends the Scan Advanced Output trigger to the front connector of the switch module. Same as Front Connector.

TTLx

Sends the Scan Advanced Output trigger to PXI or SCXI trigger line x.

PXI Star

The switch waits until it receives a trigger on the PXI star trigger bus before processing the next entry in the scan list.

Rear Connector

Valid only for SCXI. Sends the Scan Advanced Output trigger to the rear connector of the switch module.

Front Connector

Sends the Scan Advanced Output trigger to the front connector of the switch module. Same as External.

Rear Connector <1..12>

Valid only for SCXI. Sends a Scan Advanced Output trigger to the rear connector of Slot <1..12>. Can be used as an alternative to calling the niSwitch Route Scan Advanced Output node.

Front Connector <1..12>

Valid only for SCXI. Sends a Scan Advanced Output trigger to the front connector of Slot <1..12>. Can be used as an alternative to calling the niSwitch Route Scan Advanced Output node.

Default: External

datatype_icon

session out

Indicator that identifies a particular NI-SWITCH session established with the niSwitch Initialize With Topology node, the niSwitch Initialize With Options node, or the niSwitch Initialize node and used for all subsequent NI-SWITCH calls.

datatype_icon

error out

Error information.

The node produces this output according to standard error behavior.

Standard Error Behavior

Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

error in does not contain an error error in contains an error
If no error occurred before the node runs, the node begins execution normally.

If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

Synchronous Scanning

For synchronous scanning, set scan advanced output to None and define trigger input. For handshaking, trigger input and scan advanced output must be set.

Triggers generally originate from the SCXI or PXI trigger bus, the terminals on the front connector of the switch module, or the AUX connector on the rear of the switch module (referred to as the rear connector and supported by SCXI only).

When scanning a module that is not directly cabled to the trigger from its own front or rear connector, use the niSwitch Route Trigger Input node and/or the niSwitch Route Scan Advanced Output node to route the triggers on the cabled module to one of the backplane lines. When using these nodes to route triggers to the backplane, specify the backplane trigger location when calling the niSwitch Configure Scan Trigger node (even if it is the module cabled to the trigger source).

Where This Node Can Run:

Desktop OS: Windows

FPGA: Not supported

Web Server: Not supported in VIs that run in a web application