Table Of Contents

Fetch Multimode Function (Single Waveform » Complex WDT) (G Dataflow)

Version:
    Last Modified: March 3, 2017

    Retrieves complex waveforms the digitizer has acquired for the specified channel. This function returns a one-dimensional array of data that includes timing information.

    connector_pane_image
    datatype_icon

    session out

    Identifies a particular instrument session.

    datatype_icon

    error in

    Error conditions that occur before this node runs.

    The node responds to this input according to standard error behavior.

    Standard Error Behavior

    Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

    error in does not contain an error error in contains an error
    If no error occurred before the node runs, the node begins execution normally.

    If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

    If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

    Default: No error

    datatype_icon

    channels

    The channel from which to acquire data.

    Channel String Syntax

    You can input channels with the following options:
    • A single channel, such as 0
    • A list of channels, such as 0,1 or 3,2,1,0.
    • A range of channels, such as 0-7 or 0:7.
    spd-note-note
    Note  

    The order of channels in a list is important. The data for each channel is returned in the order the channels occur in the list.

    datatype_icon

    numsamples

    The maximum number of samples to fetch for each waveform.

    This function reads the actual record length and attempts to acquire all available samples. If the function fails to complete within the timeout period, the function returns an error. If the acquisition finishes with fewer points than requested, some devices return partial data if the acquisition finished, was aborted, or a timeout of 0 was used.

    Fetching All Available Samples

    Use -1 for this parameter if you want to fetch all available samples.

    Default: -1

    datatype_icon

    timeout

    The time, in seconds, to wait for the data to be acquired.

    Alternate Uses

    Use 0 for this parameter to fetch whatever is currently available. Use -1 for this parameter to imply an infinite timeout.

    Default: 5

    datatype_icon

    timestamp type

    The time basis for the timestamp on the waveform data type data.

    Name Value Description
    absolute 1 Uses an absolute time basis for the time stamp.
    relative 0 Uses a relative time basis for the time stamp.
    datatype_icon

    session out

    Passes the instrument handle to the next node in the program. This parameter has the same value as session in.

    datatype_icon

    error out

    Error information.

    The node produces this output according to standard error behavior.

    Standard Error Behavior

    Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

    error in does not contain an error error in contains an error
    If no error occurred before the node runs, the node begins execution normally.

    If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

    If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.
    datatype_icon

    wfm info

    The timing and scaling information about the waveform.

    datatype_icon

    actualSamples

    The actual number of samples in the acquired waveform.

    spd-note-note
    Note  

    The number returned in actualSamples may be less than numSamples if the number you request is not available.

    datatype_icon

    absoluteInitialX

    The timestamp, in seconds, of the first fetched sample. This timestamp is comparable between records and acquisitions; devices that do not support this parameter use 0 for this output.

    datatype_icon

    relativeInitialX

    The time, in seconds, from the trigger to the first sample in the acquired waveform.

    datatype_icon

    xIncrement

    The time, in seconds, between two samples in the acquired waveform.

    datatype_icon

    offset

    The offset factor of the given channel. Use for scaling binary data with the following formula:voltage = binary data x gain factor + offset

    datatype_icon

    gain

    The gain factor of the given channel.Use for scaling binary data with the following formula: voltage = binary data × gain factor + offset.

    datatype_icon

    reserved1

    A reserved value. Do not use.

    datatype_icon

    reserved2

    A reserved value. Do not use.

    datatype_icon

    wfm

    An array of waveform data containing one waveform without any timing information.

    Where This Node Can Run:

    Desktop OS: Windows

    FPGA: Not supported