Table Of Contents

Fetch Calibration Temperature (G Dataflow)

Version:
    Last Modified: March 3, 2017

    Returns the onboard temperature, in degrees Celsius, of the digitizer at the time of the last self-calibration or external calibration.

    The temperature returned by this function is an onboard temperature read from a sensor on the surface of the digitizer. This temperature should not be confused with the environmental temperature of the digitizer's surroundings. During operation, the onboard temperature is normally higher than the environmental temperature.

    Temperature-sensitive parameters are calibrated during self-calibration. Therefore, the self-calibration temperature is usually the more important temperature to read.

    connector_pane_image
    datatype_icon

    session in

    The instrument handle that you obtain from Start Calibration. The handle identifies a particular instrument session.

    datatype_icon

    error in

    Error conditions that occur before this node runs.

    The node responds to this input according to standard error behavior.

    Standard Error Behavior

    Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

    error in does not contain an error error in contains an error
    If no error occurred before the node runs, the node begins execution normally.

    If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

    If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

    Default: No error

    datatype_icon

    which temperature

    The calibration temperature to return, either the self-calibration or external calibration temperature.

    Name Value Description
    External Calibration 0 Returns the temperature taken during external calibration.
    Self Calibration 1 Returns the temperature taken during self-calibration.
    datatype_icon

    session out

    Passes the instrument handle to the next node in the program. This parameter has the same value as session in.

    datatype_icon

    error out

    Error information.

    The node produces this output according to standard error behavior.

    Standard Error Behavior

    Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

    error in does not contain an error error in contains an error
    If no error occurred before the node runs, the node begins execution normally.

    If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

    If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.
    datatype_icon

    temperature

    The temperature, in degrees Celsius, of the last successful calibration.

    Where This Node Can Run:

    Desktop OS: Windows

    FPGA: Not supported