Table Of Contents

Initialize Child Calibration Session (G Dataflow)

Version:
    Last Modified: September 27, 2017

    Creates a new acquisition or generation session within an existing external calibration session, based on the specified session type.

    This node returns a handle to the new child session.

    spd-note-note
    Note  

    An error occurs if another child session of the same type is already open.

    connector_pane_image
    datatype_icon

    session in

    Instrument handle that identifies your calibration session.

    This parameter is obtained from the niHSDIO Initialize External Calibration node

    datatype_icon

    session type

    The type of session, either acquisition or generation.

    datatype_icon

    error in

    Error conditions that occur before this node runs.

    The node responds to this input according to standard error behavior.

    Standard Error Behavior

    Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

    error in does not contain an error error in contains an error
    If no error occurred before the node runs, the node begins execution normally.

    If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

    If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

    Default: No error

    datatype_icon

    session out

    Session handle that passes a reference to your calibration session to the next node.

    datatype_icon

    child session out

    The niHSDIO acquisition or generation session that is created by the calibration session for the purpose of measurement.

    You can create one acquisition and/or one generation session at a time within the calibration session.

    datatype_icon

    error out

    Error information.

    The node produces this output according to standard error behavior.

    Standard Error Behavior

    Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

    error in does not contain an error error in contains an error
    If no error occurred before the node runs, the node begins execution normally.

    If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

    If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

    Where This Node Can Run:

    Desktop OS: Windows

    FPGA: Not supported

    Web Server: Not supported in VIs that run in a web application