Table Of Contents

Collector Voltage Sweep Settings (IV Analyzer)

Version:
    Last Modified: May 14, 2018

    This appears for the Transistor Analyzer mode only.

    The following Voltage sweep settings are available.

    Setting Description Operational Limits
    Vc start Initial collector voltage.

    Range: -10 V to 10 V

    Resolution: 10 mV

    Vc stop Final collector voltage.

    Range: -10 V to 10 V

    Resolution: 10 mV

    Vc step Size of the collector voltage steps during the sweep.

    Range: 10 mV to (Vcstop - Vcstart)

    Resolution: 10 mV

    Ic limit Maximum collector current during the sweep.

    Range: 0 µA to 30 mA

    Resolution: 1 µA