Table Of Contents

Digital Multimeter (DMM) Specifications

Version:
    Last Modified: May 14, 2018

    Isolated functions

    DC/AC voltage, DC/AC current, resistance, diode voltage, diode continuity

    Non-isolated functions[1]

    Capacitance, inductance

    Isolation level

    Functional isolation

    Resolution

    4.5 digits

    Input impedance

    10 MΩ

    Input coupling

    DC/AC

    Connectivity

    Banana jacks

    Voltage input protection

    ±60 V

    Current input protection

    2.5 A fuse, 5MF2.5-R

    Measurements
    Voltage measurement

    DC ranges

    50 mV DC, 500 mV DC, 5 V DC, 50 V DC

    AC ranges

    50 mV RMS, 500 mV RMS, 5 V RMS, 30 V RMS

    Input frequency range (AC voltage)

    40 Hz to 1 kHz

    DC voltage measurement accuracy (50 mV DC)

    0.2% of range

    DC voltage measurement accuracy (500 mV DC, 5 V DC, 50 V DC)

    0.1% of range

    AC voltage measurement accuracy at 50 Hz and 60 Hz (50 mV RMS)

    0.2% of range

    AC voltage measurement accuracy at 50 Hz and 60 Hz (500 mV RMS, 5 V RMS, 30 V RMS)

    0.1% of range

    Current measurement

    DC ranges

    2 A DC

    AC ranges

    2 A RMS

    Shunt resistance

    20 mΩ

    Input frequency range (AC current)

    40 Hz to 1 kHz

    DC current measurement accuracy

    0.1% of range

    AC current measurement accuracy at 50 Hz and 60 Hz

    0.1% of range

    Resistance measurement

    Ranges

    50 Ω, 500 Ω, 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ, 50 MΩ

    Resistance measurement accuracy (500 Ω, 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ)

    0.1% of range

    Resistance measurement accuracy (50 Ω, 50 MΩ)

    1% of range

    Table 1. Capacitance Measurement Range
    Range Effective Frequency[2] Effective Test Resistance[2]
    50 pF to 500 pF 10 kHz 100 kΩ
    500 pF to 5 nF 1 kHz 10 kΩ
    5 nF to 50 nF 1 kHz 10 kΩ
    50 nF to 500 nF 1 kHz 1 kΩ
    500 nF to 5 µF 1 kHz 1 kΩ
    5 µF to 50 µF 1 kHz 100 Ω
    50 µF to 500 µF 100 Hz 100 Ω

    Capacitance measurement accuracy

    1% of range

    Table 2. Inductance Measurement Range
    Range Effective Frequency[2] Effective Test Resistance[2]
    10 µH to 100 µH 100 kHz 100 Ω
    100 µH to 1 mH 10 kHz 100 Ω
    1 mH to 10 mH 10 kHz 1 kΩ
    10 mH to 100 mH 1 kHz 1 kΩ

    Inductance measurement accuracy

    1% of range

    • 1 The capacitance and inductance measurements are made through the IV Analyzer connections. Refer to the IV Analyzer specifications for more information.
    • 2 Test configurations used for measurement with single-tone test method.