Table Of Contents

Self-Test (G Dataflow)

Version:
    Last Modified: September 16, 2017

    Performs a self test on the DMM to ensure that it is functioning properly.

    This node does not calibrate the DMM. This node calls Reset, and any configurations previous to the call will be lost. All properties will be set to their default values after this node runs.

    If self-test fails, self-test message returns NIDMM_ERROR_SELF_TEST_FAILURE. This message indicates that the DMM should be repaired.

    connector_pane_image
    datatype_icon

    session in

    Instrument identifier from previous NI-DMM nodes.

    Default: None

    datatype_icon

    error in

    Error conditions that occur before this node runs.

    The node responds to this input according to standard error behavior.

    Standard Error Behavior

    Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

    error in does not contain an error error in contains an error
    If no error occurred before the node runs, the node begins execution normally.

    If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

    If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

    Default: No error

    datatype_icon

    session out

    Instrument identifier for subsequent NI-DMM nodes.

    datatype_icon

    self-test result

    Value returned from the instrument self test. A zero value indicates success.

    datatype_icon

    self-test message

    String returned from the instrument self-test.

    If you specify a string length for this parameter, it must be at least 256 elements long.

    datatype_icon

    error out

    Error information.

    The node produces this output according to standard error behavior.

    Standard Error Behavior

    Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

    error in does not contain an error error in contains an error
    If no error occurred before the node runs, the node begins execution normally.

    If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

    If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

    Where This Node Can Run:

    Desktop OS: Windows

    FPGA: Not supported

    Web Server: Not supported in VIs that run in a web application

    Hardware Support:

    For the PXI/PCI-4070, PXI-4072, PXIe-4080, and PXIe-4082, error code 1013 in self test result indicates that you should check the fuse and replace it if necessary.

    For the PXI/PCI/PCIe/USB-4065, PXI-4071, and PXIe-4081, self-test does not check the fuse. For these DMMs, self-test does not return error code 1013 even if the fuse is blown.