Configure Frequency Voltage Range (G Dataflow)

Last Modified: September 16, 2017

Specifies the expected maximum amplitude of the input signal for frequency and period measurements.

The minimum peak-to-peak signal amplitude that a DMM can detect is 10% of the specified range.

connector_pane_image
datatype_icon

session in

Instrument identifier from previous NI-DMM nodes.

Default: None

datatype_icon

frequency voltage range

Expected maximum amplitude of the input signal.

NI-DMM sets the Frequency Voltage Range property to this value. For a list of valid values, refer to the help for your DMM.

You can use this parameter to turn Auto Range on or off. The following table describes valid values for configuring Auto Range.

Value Setting Description
-1.0 Auto Range ON Configures the DMM to take an Auto Range measurement to calculate the voltage range before each frequency or period measurement.
-2.0 Auto Range OFF Disables Auto Range. NI-DMM sets the voltage range to the last calculated voltage range.
datatype_icon

error in

Error conditions that occur before this node runs.

The node responds to this input according to standard error behavior.

Standard Error Behavior

Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

error in does not contain an error error in contains an error
If no error occurred before the node runs, the node begins execution normally.

If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

Default: No error

datatype_icon

session out

Instrument identifier for subsequent NI-DMM nodes.

datatype_icon

error out

Error information.

The node produces this output according to standard error behavior.

Standard Error Behavior

Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

error in does not contain an error error in contains an error
If no error occurred before the node runs, the node begins execution normally.

If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

Where This Node Can Run:

Desktop OS: Windows

FPGA: Not supported

Web Server: Not supported in VIs that run in a web application

Hardware Support:

The following DMMs support this node:

  • PXI-4070
  • PXI-4071
  • PXI-4072
  • PXIe-4080
  • PXIe-4081
  • PXIe-4082