Table Of Contents

Export Signal (G Dataflow)

Last Modified: January 29, 2018

Routes signals (triggers and events) to the specified output terminal.

The route is created when the session is Committed.

connector_pane_image
datatype_icon

session in

Instrument session obtained from the Initialize With Channels node.

datatype_icon

signal identifier

Input reserved for future use.

datatype_icon

signal

Trigger or event to export.

Name Value Description
Source Complete Event 1030 Exports the Source Complete event.
Measure Complete Event 1031 Exports the Measure Complete event.
Sequence Iteration Complete Event 1032 Exports the Sequence Iteration Complete event.
Sequence Engine Done Event 1033 Exports the Sequence Engine Done event.
Pulse Complete Event 1053 Exports the Pulse Complete event.
Ready for Pulse Trigger Event 1052 Exports the Ready for Pulse Trigger event.
Start Trigger 1036 Exports the Start trigger.
Source Trigger 1037 Exports the Source trigger.
Measure Trigger 1038 Exports the Measure trigger.
Sequence Advance Trigger 1039 Exports the Sequence Advance trigger.
Pulse Trigger 1040 Exports the Pulse trigger.
datatype_icon

output terminal

Line to export the selected signal.

"" Do not export signal.
"PXI_Trig0" The signal is exported on PXI or RTSI line 0.
"PXI_Trig1" The signal is exported on PXI or RTSI line 1.
"PXI_Trig2" The signal is exported on PXI or RTSI line 2.
"PXI_Trig3" The signal is exported on PXI or RTSI line 3.
"PXI_Trig4" The signal is exported on PXI or RTSI line 4.
"PXI_Trig5" The signal is exported on PXI or RTSI line 5.
"PXI_Trig6" The signal is exported on PXI or RTSI line 6.
"PXI_Trig7" The signal is exported on PXI or RTSI line 7.
datatype_icon

error in

Error conditions that occur before this node runs.

The node responds to this input according to standard error behavior.

Standard Error Behavior

Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

error in does not contain an error error in contains an error
If no error occurred before the node runs, the node begins execution normally.

If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

Default: No error

datatype_icon

session out

Handle used to identify the session in all subsequent NI-DCPower node calls.

datatype_icon

error out

Error information.

The node produces this output according to standard error behavior.

Standard Error Behavior

Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

error in does not contain an error error in contains an error
If no error occurred before the node runs, the node begins execution normally.

If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

Where This Node Can Run:

Desktop OS: Windows

FPGA: Not supported

Web Server: Not supported in VIs that run in a web application

Hardware Support:

This node does not support the following hardware:

  • PXI-4110
  • PXI-4130


Recently Viewed Topics