Table Of Contents

Adjust Output Resistance Calibration (G Dataflow)

Version:
    Last Modified: March 3, 2017

    Calculates the calibration constants for the output resistance of the specified channel by comparing requested and measured outputs.

    This node can only be called from an external calibration session.

    connector_pane_image
    datatype_icon

    session in

    Instrument calibration session obtained from the Initialize External Calibration node.

    datatype_icon

    channel name

    Output channel to which these calibration settings apply.

    Only one channel at a time can be calibrated.

    datatype_icon

    number of measurements

    Number of elements in requested outputs and measured outputs.

    datatype_icon

    error in

    Error conditions that occur before this node runs.

    The node responds to this input according to standard error behavior.

    Standard Error Behavior

    Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

    error in does not contain an error error in contains an error
    If no error occurred before the node runs, the node begins execution normally.

    If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

    If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

    Default: No error

    datatype_icon

    requested outputs

    Array of the output values requested in the Configure Current Limit node.

    datatype_icon

    measured outputs

    Array of the output values measured by an external precision digital multimeter (DMM).

    datatype_icon

    session out

    Handle used to identify the session in all subsequent NI-DCPower node calls.

    datatype_icon

    error out

    Error information.

    The node produces this output according to standard error behavior.

    Standard Error Behavior

    Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

    error in does not contain an error error in contains an error
    If no error occurred before the node runs, the node begins execution normally.

    If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

    If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

    Calibration Procedure

    Refer to the calibration procedure for the device you are calibrating for detailed instructions on the appropriate use of this node.

    Where This Node Can Run:

    Desktop OS: Windows

    FPGA: Not supported

    Hardware Support:

    This node supports the following devices:

    • PXIe-4154


    Recently Viewed Topics