Calculates calibration constants for the current limit for the specified output channel and range by comparing requested and measured outputs.
This node can only be called from an external calibration session.
Instrument calibration session obtained from the Initialize External Calibration node.
Output channel to which these calibration settings apply.
Only one channel at a time can be calibrated.
Number of elements in requested outputs and measured outputs.
Range to calibrate with these settings.
Array of the output values requested in the Configure Current Limit node.
Array of the output values measured by an external precision digital multimeter (DMM).
Error conditions that occur before this node runs.
The node responds to this input according to standard error behavior.
Standard Error Behavior
Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.
Default: No error
Handle used to identify the session in all subsequent NI-DCPower node calls.
Error information.
The node produces this output according to standard error behavior.
Standard Error Behavior
Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.
Refer to the calibration procedure for the device you are calibrating for detailed instructions on the appropriate use of this node.