Calculates the calibration constants for the residual voltage offsets for the specified output channel.
This node can be called only in an external calibration session. The node requires that the output be shorted prior to the node being invoked.
Instrument calibration session obtained from the Initialize External Calibration node.
Output channel to which these calibration settings apply.
Only one channel at a time can be calibrated.
Error conditions that occur before this node runs.
The node responds to this input according to standard error behavior.
Standard Error Behavior
Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.
Default: No error
Handle used to identify the session in all subsequent NI-DCPower node calls.
Error information.
The node produces this output according to standard error behavior.
Standard Error Behavior
Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.
Residual offsets account for minor offset effects on the device that lie outside of the self-calibration circuitry. These offsets can include multiplexer input offsets and leakage effects from internal switching.
Refer to the calibration procedure for the device you are calibrating for detailed instructions on the appropriate use of this node.
Where This Node Can Run:
Desktop OS: Windows
FPGA: Not supported
Web Server: Not supported in VIs that run in a web application
Hardware Support:
This node does not support the following hardware: