Table Of Contents

Adjust Internal Reference Calibration (G Dataflow)

Version:
Last Modified: January 29, 2018

Programs the adjusted reference value to the device.

This node can only be called from an external calibration session.

connector_pane_image
datatype_icon

session in

Instrument calibration session obtained from the Initialize External Calibration node.

datatype_icon

error in

Error conditions that occur before this node runs.

The node responds to this input according to standard error behavior.

Standard Error Behavior

Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

error in does not contain an error error in contains an error
If no error occurred before the node runs, the node begins execution normally.

If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

Default: No error

datatype_icon

internal reference

Value that specifies the internal reference to be adjusted.

Name Value Description
5K Reference 1054 Calibration pin connected to 5 V internal reference.
100kOhm Reference 1055 Calibration pin connected to 100 kΩ internal reference.
datatype_icon

adjusted internal reference value

Updated value of the internal reference that will be programmed to the device.

datatype_icon

session out

Handle used to identify the session in all subsequent NI-DCPower node calls.

datatype_icon

error out

Error information.

The node produces this output according to standard error behavior.

Standard Error Behavior

Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

error in does not contain an error error in contains an error
If no error occurred before the node runs, the node begins execution normally.

If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

Calibration Procedure

Refer to the calibration procedure for the device you are calibrating for detailed instructions on the appropriate use of this node.

Where This Node Can Run:

Desktop OS: Windows

FPGA: Not supported

Web Server: Not supported in VIs that run in a web application

Hardware Support:

This node supports the following hardware:

  • PXIe-4162/4163


Recently Viewed Topics